Respiratory Monitoring Based on Tracheal Sounds: Continuous Time-Frequency Processing of the Phonospirogram Combined with Phonocardiogram-Derived Respiration
<p>Illustration of the microphone setup.</p> "> Figure 2
<p>Position of the microphones: two microphones are inserted into the support that is fixed at suprasternal notch with two strips of adhesive tape.</p> "> Figure 3
<p>Recording procedure: 0–60 s normal respiration (NR); 60–70 s apnea (holding breathing spontaneously, AP); 70–100 s normal respiration + speech (RS); 100–120 s normal respiration + environmental noise (RN).</p> "> Figure 4
<p>Diagram of the detection algorithm.</p> "> Figure 5
<p>The spectrogram for one recording.</p> "> Figure 6
<p>Temporal and frequency detection result of the same period from the same recording. A positive (resp. negative) pneumotachograph reference signal (purple) indicates inspiration (resp. expiration). The adaptive (pink) and minimum thresholds (gray) are marked in the figure. The envelope signal (temporal or PSD) is in blue, and the final detection result is in red with the peak “middle-timing moment”.</p> "> Figure 6 Cont.
<p>Temporal and frequency detection result of the same period from the same recording. A positive (resp. negative) pneumotachograph reference signal (purple) indicates inspiration (resp. expiration). The adaptive (pink) and minimum thresholds (gray) are marked in the figure. The envelope signal (temporal or PSD) is in blue, and the final detection result is in red with the peak “middle-timing moment”.</p> "> Figure 7
<p>Cardiac S peak detection. There is a small delay between the electrocardiogram (ECG) (<b>yellow</b>) and the extracted cardiac signal (<b>blue</b>). Predicting area PA(ncb) (purple) defined by <math display="inline"><semantics> <mrow> <mo>±</mo> <mn>20</mn> <mo>%</mo> </mrow> </semantics></math> of last heart beat-to-beat interval <math display="inline"><semantics> <mrow> <mi>b</mi> <mo>_</mo> <mi>b</mi> <mo>(</mo> <mi>n</mi> <mi>c</mi> <mi>b</mi> <mo>−</mo> <mn>1</mn> <mo>)</mo> </mrow> </semantics></math> and the adaptive threshold. The cardiac envelogram is in red, and the detected cardiac peak (S1/S2) is marked by a green star.</p> "> Figure 8
<p>PDR. The extracted cardiac signal is shown in green, and the processed envelogram is in red. Only one cardiac peak (S1/S2) detected (<b>pink star</b>) for each cardiac cycle. Positive (Negative) of the PDR detection result (<b>yellow</b>) indicates the inspiration (expiration) as reference signal.</p> "> Figure 9
<p>Final decision-making for apnea detection. Detection stat: 2—Speech; 1.5—Respiration; 0—Pauses; −1—apnea. The reference signal: pneumotachograph is in blue, and the apnea detection signal is in red.</p> "> Figure 10
<p>Final decision-making for breathing activities assessment. (<b>Top</b>): results from temporal detection. (<b>Middle</b>): results from frequency detection. (<b>Bottom</b>): results from cardiac (PDR) detection. The final detected events are marked in purple dashed lines.</p> "> Figure 11
<p>Example of breathing activity counting. Respiratory envelope (in blue), detection signal (in red). (<b>a</b>) If two continuous breathing activities are recognized as a single event, or the pause/apnea is detected as one breathing activity, a false positive detection will be reported; (<b>b</b>) if one breathing activity is recognized as two events, or one breathing activity is not detected, a false negative detection will be reported.</p> "> Figure 12
<p>Box plot for final respiratory detection result. The line and the cross inside the box correspond to the median and the mean value, respectively. The box-limits represent the first and the third quartiles, the points highlight outliers, and the upper and lower bars represent the maximum and minimum values, respectively. Specificity: blue box. Sensitivity: orange box. Accuracy: gray box.</p> "> Figure 13
<p>Influence of speech and noise on respiratory detection. Box plot (as box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for specificity (green and light gray boxes), sensitivity (orange and yellow boxes) and accuracy (dark gray and blue boxes) of temporal (T), frequency (F), and cardiac (C) detection domains, separated in NR+AP phases (in blue square) and RS + RN phases (in green square).</p> "> Figure 14
<p>Box plot (like box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for specificity (green rectangle), sensitivity (purple rectangle), and accuracy (red rectangle) in sitting (orange boxes) and lying (blue boxes) positions.</p> "> Figure 15
<p>Box plot (as box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for specificity (blue), sensitivity (green), and accuracy (orange) of temporal (T), frequency (F), and cardiac (C) detection domains, as well as global result (G).</p> "> Figure 16
<p>PDR (green) processed from tracheal sounds and ECG-Derived Respiration (EDR) (red) processed from ECG and the thoracic strap signal (blue). Their positive (negative) slopes indicate inspiration (expiration).</p> "> Figure 17
<p>(<b>Top</b>): the EDR signal and its local maxima (peaks) <math display="inline"><semantics> <mrow> <mi>i</mi> <mi>E</mi> <mi>D</mi> <mi>R</mi> </mrow> </semantics></math>. (<b>Middle</b>): the PDR signal and its local maxima <math display="inline"><semantics> <mrow> <mi>i</mi> <mi>P</mi> <mi>D</mi> <mi>R</mi> </mrow> </semantics></math>. (<b>Bottom</b>): the linear regression fit line between iEDR and iPDR (<math display="inline"><semantics> <mrow> <mi>a</mi> <mo>=</mo> <mn>0.98</mn> </mrow> </semantics></math>).</p> "> Figure 18
<p>(<b>Top</b>): the EDR signal and its local maxima (peaks) <math display="inline"><semantics> <mrow> <mi>i</mi> <mi>E</mi> <mi>D</mi> <mi>R</mi> </mrow> </semantics></math>. (<b>Middle</b>): the PDR signal and its local maxima <math display="inline"><semantics> <mrow> <mi>i</mi> <mi>P</mi> <mi>D</mi> <mi>R</mi> </mrow> </semantics></math>. (<b>Bottom</b>): the theory fit line(y = x) in red and the linear regression fit line between iEDR and iPDR (<math display="inline"><semantics> <mrow> <mi>a</mi> <mo>=</mo> <mn>0.68</mn> </mrow> </semantics></math>).</p> "> Figure 19
<p>Box plot (like box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for the evaluation of the results of PDR according to EDR, separated into NR+AP (blue), RS+SN (orange), and global results (gray).</p> "> Figure 20
<p>Box plot (like box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for the evaluation of the results of PDR according to EDR, separated in sitting (<b>left</b>) and lying (<b>right</b>) positions.</p> "> Figure 21
<p>The left, middle, and right figures show the linear regressions for NR + AP, RS + RN, and the whole recording. The blue points are from the group of R-R vs. S1-S1 or S2-S2 (group 1, at the middle group of each figure), the orange points are from the group of R-R vs. S2-S1 (group 2, at the top of each figure), and the yellow points are from the group of R-R vs. S1-S2 (group 3, at the bottom of each figure). The slopes of each fit line are presented the in figure title (<math display="inline"><semantics> <mrow> <mi>a</mi> <mn>1</mn> </mrow> </semantics></math>, <math display="inline"><semantics> <mrow> <mi>a</mi> <mn>2</mn> </mrow> </semantics></math>, and <math display="inline"><semantics> <mrow> <mi>a</mi> <mn>3</mn> </mrow> </semantics></math> are slopes for the group 1, group 2, and group 3, respectively).</p> "> Figure 22
<p>Box plot (like the box plot of <a href="#sensors-21-00099-f012" class="html-fig">Figure 12</a>) for the distribution of fit line slopes for each group in each part and the global signal (in different rectangle).</p> ">
Abstract
:1. Introduction
1.1. Apnea Detection
1.1.1. Detection in the Temporal Domain
1.1.2. Detection in the Frequency Domain
1.2. Phonocardiogram (PCG)-Derived Respiration
2. Material and Method
2.1. Material
2.2. Protocol
2.3. Algorithm
2.3.1. Speech Detection
2.3.2. Frequency Band Separation
2.3.3. Temporal Envelope Detection
- The minimum threshold:
- The adaptive threshold:
2.3.4. Frequency Detection
- The minimum threshold:
- The adaptive threshold:
2.3.5. PDR Detection
2.3.6. Decision-Making
Apnea Detection
Breathing Activities Assessment
Statistics
3. Results
3.1. Evaluation Method
- the specificity of respiratory detection (breathing/apnea) from TS must be >82.9%.
- the accuracy of respiratory detection (breathing/apnea) from TS must be >90.5%.
3.2. Apnea Detection
3.3. Breathing Activities Assessment
3.3.1. Influence of Speech and Noise
3.3.2. Influence of Position
3.3.3. Advantage of Combining Several Detection Domains
3.4. Cardiac Sound Processing
3.4.1. Correlation between PDR and EDR
3.4.2. Heart Beat-to-Beat Interval
4. Discussion
5. Conclusions
Author Contributions
Funding
Ethical Statement
Data Availability Statement
Conflicts of Interest
References
- Similowski, T.; Derenne, J.P. Stimulation phrénique implantée. Méd. Thér. 2001, 7, 457–469. [Google Scholar]
- Weese-Mayer, D.E.; Silvestri, J.M.; Kenny, A.S.; Ilbawi, M.N.; Hauptman, S.A.; Lipton, J.W.; Talonen, P.P.; Garcia, H.G.; Watt, J.W.; Exner, G.; et al. Diaphragm pacing with a quadripolar phrenic nerve electrode: An international study. PACE Pacing Clin. Electrophysiol. 1996, 19, 1311–1319. [Google Scholar] [CrossRef]
- Glenn, W.W.; Holcomb, W.G.; Gee, J.B.; Rath, R. Central hypoventilation; long-term ventilatory assistance by radiofrequency electrophrenic respiration. Ann. Surg. 1970, 172, 755–773. [Google Scholar] [CrossRef] [PubMed]
- Le Pimpec-Barthes, F.; Gonzalez-Bermejo, J.; Hubsch, J.P.; Duguet, A.; Morélot-Panzini, C.; Riquet, M.; Similowski, T. Intrathoracic phrenic pacing: A 10-year experience in France. J. Thorac. Cardiovasc. Surg. 2011, 142, 378–383. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Iber, C.; Ancoli-Israel, S.; Chesson, A.L.; Quan, S.F. The AASM Manual for the Scoring of Sleep and Associated Events: Rules, Terminology and Technical Specifications; American Academy of Sleep Medicine Westchester: Darien, IL, USA, 2007. [Google Scholar]
- DiMarco, A.F. Diaphragm Pacing. Clin. Chest Med. 2018, 39, 459–471. [Google Scholar] [CrossRef]
- Que, C.l.; Kolmaga, C.; Durand, L.G.; Kelly, S.M.; Macklem, P.T. Phonospirometry for noninvasive measurement of ventilation: Methodology and preliminary results. J. Appl. Physiol. 2002, 93, 1515–1526. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Corbishley, P.; Rodríguez-Villegas, E. Breathing detection: Towards a miniaturized, wearable, battery-operated monitoring system. IEEE Trans. Biomed. Eng. 2008, 55, 196–204. [Google Scholar] [CrossRef]
- Yadollahi, A.; Giannouli, E.; Moussavi, Z. Sleep apnea monitoring and diagnosis based on pulse oximetery and tracheal sound signals. Med. Biol. Eng. Comput. 2010, 48, 1087–1097. [Google Scholar] [CrossRef] [PubMed]
- Kalkbrenner, C.; Eichenlaub, M.; Rüdiger, S.; Kropf-Sanchen, C.; Rottbauer, W.; Brucher, R. Apnea and heart rate detection from tracheal body sounds for the diagnosis of sleep-related breathing disorders. Med. Biol. Eng. Comput. 2017, 56, 671–681. [Google Scholar] [CrossRef] [PubMed]
- Chuah, J.S.; Moussavi, Z. Automated detection of respiratory phases by acoustical means. In Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol.20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286), Hong Kong, China, 1 November 2001. [Google Scholar]
- Hult, P.; Fjallbrant, T.; Wranne, B.; Engdahl, O.; Ask, P. An improved bioacoustic method for monitoring of respiration. Technol. Health Care 2004, 12, 323–332. [Google Scholar] [CrossRef] [PubMed]
- Hult, P.; Fjallbrant, T.; Dahle, S.; Danielsson, P.; Ask, P. A method for respiration monitoring by the use of a bioacoustic signal. In Proceedings of the First International Conference on Advances in Medical Signal and Information Processing, Bristol, UK, 4–6 September 2000; IET: London, UK, 2000; Volume 2000, pp. 22–25. [Google Scholar] [CrossRef]
- Kalkbrenner, C.; Stark, P.; Kouemou, G.; Algorri, M.E.; Brucher, R. Sleep monitoring using body sounds and motion tracking. In Proceedings of the 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, IL, USA, 26–30 August 2014; Volume 6000, pp. 6941–6944. [Google Scholar] [CrossRef]
- Kulkas, A.; Huupponen, E.; Virkkala, J.; Tenhunen, M.; Saastamoinen, A.; Rauhala, E.; Himanen, S.L. New tracheal sound feature for apnoea analysis. Med. Biol. Eng. Comput. 2009, 47, 405–412. [Google Scholar] [CrossRef] [PubMed]
- Lei, B.; Rahman, S.A.; Song, I. Content-based classification of breath sound with enhanced features. Neurocomputing 2014, 141, 139–147. [Google Scholar] [CrossRef]
- Muthusamy, P.D.; Sundaraj, K.; Abd Manap, N. Computerized acoustical techniques for respiratory flow-sound analysis: A systematic review. Artif. Intell. Rev. 2019, 53, 501–3574. [Google Scholar] [CrossRef]
- Schmidt, M.; Schumann, A.; Müller, J.; Bär, K.J.; Rose, G. ECG derived respiration: Comparison of time-domain approaches and application to altered breathing patterns of patients with schizophrenia. Physiol. Meas. 2017, 38, 601–615. [Google Scholar] [CrossRef] [PubMed]
- Martin, A.; Voix, J. In-Ear Audio Wearable: Measurement of Heart and Breathing Rates for Health and Safety Monitoring. IEEE Trans. Biomed. Eng. 2018, 65, 1256–1263. [Google Scholar] [CrossRef] [PubMed]
- White Noise Soundscapes. City Traffic White Noise |10 Hours| For Sleeping, Studying or to Block Out Noise. Available online: https://www.youtube.com/watch?v=8s5H76F3SIs (accessed on 25 December 2020).
- Lu, X.; Guiraud, D.; Renaux, S.; Similowski, T.; Azevedo, C. Breathing detection from tracheal sounds in both temporal and frequency domains in the context of phrenic nerve stimulation. In Proceedings of the 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany, 23–27 July 2019; pp. 5473–5476. [Google Scholar] [CrossRef]
- Lu, X.; Guiraud, D.; Renaux, S.; Similowski, T.; Azevedo Coste, C.; Azevedo, C. Monitoring phrenic nerve stimulation-induced breathing via tracheal sounds. In Proceedings of the 2019 XXIII Annual Conference of the International Functional Electrical Stimulation Society (IFESS), Toronto, ON, Canada, 24–28 June 2019. [Google Scholar]
- Penzel, T.; Sabil, A.K. The use of tracheal sounds for the diagnosis of sleep apnoea. Breathe 2017, 13, e37–e45. [Google Scholar] [CrossRef] [PubMed]
- Liang, H.; Lukkarinen, S.; Hartimo, I. Heart sound segmentation algorithm based on heart sound envelogram. In Proceedings of the Computers in Cardiology, Lund, Sweden, 7–10 September 1997; Volume 24, pp. 105–108. [Google Scholar] [CrossRef]
- Plapous, C.; Marro, C.; Scalart, P. Improved signal-to-noise ratio estimation for speech enhancement. IEEE Trans. Audio Speech Lang. Process. 2006, 14, 2098–2108. [Google Scholar] [CrossRef] [Green Version]
- Ozkan, I.; Member, S.; Yilmaz, A.; Celebi, G. Improved Segmentation with Dynamic Threshold Adjustment for Phonocardiography Recordings. In Proceedings of the 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany, 23–27 July 2019; pp. 6681–6684. [Google Scholar]
State of | Explication | Conditions |
---|---|---|
0 | Pause/apnea | |
1 | Inspiration/Expiration | |
2 | Speech/Snoring |
State of | Explication | Conditions |
---|---|---|
0 | Pause/apnea | |
1 | Inspiration/Expiration | |
2 | Speech/Snoring |
State of | Explication | Conditions |
---|---|---|
−1 | expiration | |
1 | Inspiration | |
0 | Pause/apnea | s s |
2 | Speech/Snoring |
Detection Domain | Median Comparison | p-Value |
---|---|---|
Temporal | Specificity | p = 0.0966 |
sensitivity | p = 0.8829 | |
Accuracy | p = 0.3638 | |
Frequency | Specificity | p = 0.0088 * |
sensitivity | p = 0.2032 | |
Accuracy | p = 0.1219 | |
Cardiac (PDR) | Specificity | p = 0.016 * |
sensitivity | p = 0.7369 | |
Accuracy | p = 0.0851 |
Median Comparison | p-Value |
---|---|
Specificity | p = 0.2079 |
sensitivity | p = 0.7354 |
Accuracy | p = 1 |
Median Comparison | Detection Domain | p-Value |
---|---|---|
Specificity | Temporal | p = 0.00889 * |
Frequency | p = 0.01489 * | |
Cardiac | p = 9.896 × 10 * | |
Sensitivity | Temporal | p = 0.09933 |
Frequency | p = 0.6348 | |
Cardiac | p = 0.001801 * | |
Accuracy | Temporal | p = 0.8417 |
Frequency | p = 0.8028 | |
Cardiac | p = 0.2079 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Lu, X.; Azevedo Coste, C.; Nierat, M.-C.; Renaux, S.; Similowski, T.; Guiraud, D. Respiratory Monitoring Based on Tracheal Sounds: Continuous Time-Frequency Processing of the Phonospirogram Combined with Phonocardiogram-Derived Respiration. Sensors 2021, 21, 99. https://doi.org/10.3390/s21010099
Lu X, Azevedo Coste C, Nierat M-C, Renaux S, Similowski T, Guiraud D. Respiratory Monitoring Based on Tracheal Sounds: Continuous Time-Frequency Processing of the Phonospirogram Combined with Phonocardiogram-Derived Respiration. Sensors. 2021; 21(1):99. https://doi.org/10.3390/s21010099
Chicago/Turabian StyleLu, Xinyue, Christine Azevedo Coste, Marie-Cécile Nierat, Serge Renaux, Thomas Similowski, and David Guiraud. 2021. "Respiratory Monitoring Based on Tracheal Sounds: Continuous Time-Frequency Processing of the Phonospirogram Combined with Phonocardiogram-Derived Respiration" Sensors 21, no. 1: 99. https://doi.org/10.3390/s21010099
APA StyleLu, X., Azevedo Coste, C., Nierat, M.-C., Renaux, S., Similowski, T., & Guiraud, D. (2021). Respiratory Monitoring Based on Tracheal Sounds: Continuous Time-Frequency Processing of the Phonospirogram Combined with Phonocardiogram-Derived Respiration. Sensors, 21(1), 99. https://doi.org/10.3390/s21010099