[go: up one dir, main page]
More Web Proxy on the site http://driver.im/

Dual-Modality Volume Measurement Integrated on a Ventricular Assist Device

IEEE Trans Biomed Eng. 2022 Mar;69(3):1151-1161. doi: 10.1109/TBME.2021.3115019. Epub 2022 Feb 18.

Abstract

Objective: Ventricular assist devices (VADs) are implanted in patients suffering from end-stage heart failure to sustain the blood circulation. Real-time volume measurement could be a valuable tool to monitor patients and enable physiological control strategies to provide individualized therapy. However, volume measurement using one sensor modality requires re-calibration in the critical time post VAD implantation.

Methods: To overcome this limitation, we have integrated ultrasound and impedance volume measurement techniques into a cannula of an apical VAD. We tested both modalities across a volume range from 140-420 mL using two differently sized and shaped biventricular silicon heart phantoms, which were subjected to physiological pressures in an in-vitro test bench. We compared results from standard calibrated measurements with calculations found by a quadratic optimization for the single modality and their combination (dual-modality) and validated the results using twofold cross-validation.

Results: The dual-modality approach resulted in most favorable limits of agreement (LOA) of -0.83 ± 1.54% compared to -13.88 ± 5.90% for ultrasound and -43.45 ± 10.28% for electric impedance, separately.

Conclusion: The results of the dual-modality approach were as accurate as the standard calibrated measurement and valid over a large range of volumes (140-420 mL). In this in-vitro study, we show how a dual-modality ventricular volume measurement of ultrasound and electric impedance increases the robustness and renders calibration obsolete.

Significance: Ventricular volumes could be measured accurately in the critical period post VAD implantation despite ventricular remodeling.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Heart
  • Heart Failure*
  • Heart-Assist Devices*
  • Humans