US20030111142A1 - Bulk metallic glass medical instruments, implants, and methods of using same - Google Patents
Bulk metallic glass medical instruments, implants, and methods of using same Download PDFInfo
- Publication number
- US20030111142A1 US20030111142A1 US10/256,751 US25675102A US2003111142A1 US 20030111142 A1 US20030111142 A1 US 20030111142A1 US 25675102 A US25675102 A US 25675102A US 2003111142 A1 US2003111142 A1 US 2003111142A1
- Authority
- US
- United States
- Prior art keywords
- mri
- metallic glass
- bulk metallic
- procedure
- bmg
- Prior art date
- Legal status (The legal status is an assumption and is not a legal conclusion. Google has not performed a legal analysis and makes no representation as to the accuracy of the status listed.)
- Abandoned
Links
Images
Classifications
-
- C—CHEMISTRY; METALLURGY
- C22—METALLURGY; FERROUS OR NON-FERROUS ALLOYS; TREATMENT OF ALLOYS OR NON-FERROUS METALS
- C22C—ALLOYS
- C22C14/00—Alloys based on titanium
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B17/00—Surgical instruments, devices or methods, e.g. tourniquets
- A61B17/56—Surgical instruments or methods for treatment of bones or joints; Devices specially adapted therefor
- A61B17/58—Surgical instruments or methods for treatment of bones or joints; Devices specially adapted therefor for osteosynthesis, e.g. bone plates, screws, setting implements or the like
- A61B17/68—Internal fixation devices, including fasteners and spinal fixators, even if a part thereof projects from the skin
- A61B17/84—Fasteners therefor or fasteners being internal fixation devices
- A61B17/86—Pins or screws or threaded wires; nuts therefor
- A61B17/866—Material or manufacture
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61L—METHODS OR APPARATUS FOR STERILISING MATERIALS OR OBJECTS IN GENERAL; DISINFECTION, STERILISATION OR DEODORISATION OF AIR; CHEMICAL ASPECTS OF BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES; MATERIALS FOR BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES
- A61L31/00—Materials for other surgical articles, e.g. stents, stent-grafts, shunts, surgical drapes, guide wires, materials for adhesion prevention, occluding devices, surgical gloves, tissue fixation devices
- A61L31/02—Inorganic materials
- A61L31/026—Ceramic or ceramic-like structures, e.g. glasses
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61L—METHODS OR APPARATUS FOR STERILISING MATERIALS OR OBJECTS IN GENERAL; DISINFECTION, STERILISATION OR DEODORISATION OF AIR; CHEMICAL ASPECTS OF BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES; MATERIALS FOR BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES
- A61L31/00—Materials for other surgical articles, e.g. stents, stent-grafts, shunts, surgical drapes, guide wires, materials for adhesion prevention, occluding devices, surgical gloves, tissue fixation devices
- A61L31/14—Materials characterised by their function or physical properties, e.g. injectable or lubricating compositions, shape-memory materials, surface modified materials
- A61L31/18—Materials at least partially X-ray or laser opaque
-
- C—CHEMISTRY; METALLURGY
- C22—METALLURGY; FERROUS OR NON-FERROUS ALLOYS; TREATMENT OF ALLOYS OR NON-FERROUS METALS
- C22C—ALLOYS
- C22C45/00—Amorphous alloys
- C22C45/10—Amorphous alloys with molybdenum, tungsten, niobium, tantalum, titanium, or zirconium or Hf as the major constituent
-
- C—CHEMISTRY; METALLURGY
- C22—METALLURGY; FERROUS OR NON-FERROUS ALLOYS; TREATMENT OF ALLOYS OR NON-FERROUS METALS
- C22C—ALLOYS
- C22C9/00—Alloys based on copper
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B10/00—Other methods or instruments for diagnosis, e.g. instruments for taking a cell sample, for biopsy, for vaccination diagnosis; Sex determination; Ovulation-period determination; Throat striking implements
- A61B10/02—Instruments for taking cell samples or for biopsy
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B17/00—Surgical instruments, devices or methods, e.g. tourniquets
- A61B17/32—Surgical cutting instruments
- A61B17/3209—Incision instruments
- A61B17/3211—Surgical scalpels, knives; Accessories therefor
Definitions
- the present invention relates to medical, surgical, and dental hardware, especially medical instruments and biomedical appliances, and particularly to those instruments and appliances at least partially constructed of a bulk metallic glass (BMG), and to methods of using the same.
- BMG bulk metallic glass
- MRI magnetic resonance imaging
- the concept of magnetic susceptibility is central to many current research and development activities in magnetic resonance imaging (MRI).
- MRI magnetic resonance imaging
- the development of MR-guided surgery has created a need for surgical instruments and other devices with susceptibility tailored to the MR environment; susceptibility effects can lead to position errors of up to several millimeters in MR-guided stereotactic surgery; and the variation of magnetic susceptibility on a microscopic scale within tissues contributes to MR contrast and is the basis of functional MRI.
- Magnetic aspects of MR compatibility are discussed in terms of two levels of acceptability: Materials with the first kind of magnetic field compatibility are such that magnetic forces and torques do not interfere significantly when the materials are used within the magnetic field of the scanner; materials with the second kind of magnetic field compatibility meet the more demanding requirement that they produce only negligible artifacts within the MR image and their effect on the positional accuracy of features within the image is negligible or can readily be corrected.
- materials exhibiting magnetic field compatibility of the second kind have been studied and a group of materials that produce essentially no image distortion, even when located directly within the imaging field of view, is identified.
- a new implant material has long been needed which has a low MRI signature.
- a desirable material would have a lower elastic modulus and an extremely high elastic limit of about 2% compared to that of a typical metal, namely about 0.2%.
- Bone has an elastic limit of about 1%.
- Such material would be unique in its ability to flex elastically with the natural bending of the bones and so distribute stresses more uniformly. Faster healing rates would result from reduced stress shielding effects while minimizing stress concentrators. Because of these unique mechanical properties, screws could have a thinner shank and deeper threads yielding greater holding power.
- Applications where such material is desirable would include such as fracture fixation screws, rods, pins, knee and hip joint wear surfaces and shafts, and aneurysm clips. A large variety of other applications, changes and modifications would be obvious to those skilled in the art.
- a bulk metallic glass is defined for purposes herein as an amorphous metallic alloy that is cast in bulk form.
- a BMG is known to be inclusive of amorphous thin-film materials such as those that are typically deposited on surfaces.
- a medical instrument is defined for purposes herein as any device used by medical and/or dental personnel in any surgical and/or dental procedure.
- a biomedical appliance is defined for purposes herein as any medically functional device that is configured for disposition on or inside a living body, including surgically implanted orthopedic devices, dental implants, and the like.
- objects of the present invention include at least the following:
- an article at least a portion of which includes a bulk metallic glass having magnetic properties suitable for producing an MRI image, the article having a shape suitable for producing an MRI image and being configured for use as a medical instrument and/or a biomedical appliance.
- a method of carrying out a medical procedure includes the steps of: providing a medical instrument or a biomedical appliance, at least a portion of which includes a bulk metallic glass; and using the medical instrument or biomedical appliance to carry out a medical procedure.
- a method of carrying out an MRI-guided procedure includes the steps of: providing an article, at least a portion of which includes a bulk metallic glass; and using the article to carry out an MRI-guided procedure.
- FIG. 1 is a full-scale MRI image of a 7 mm diameter rod of a conventional copper alloy (Cu-4 Cri-2 Nb) in salt water.
- FIG. 2 is a full-scale MRI image of a 7 mm diameter rod of BAM-11 in salt water in accordance with the present invention.
- FIG. 3 is a full-scale MRI image of a 7 mm diameter rod of Ni-free BMG in salt water in accordance with the present invention.
- FIG. 4 is a full-scale MRI image of a bullet nosed 6 mm diameter rod of Ti-6 Al-4.
- FIG. 5 is a full-scale MRI image of a bullet nosed 6 mm diameter rod of BAM-11 in salt water in accordance with the present invention.
- FIG. 6 is a view of a typical conventional orthopedic bone screw.
- FIG. 7 is a view of an orthopedic bone screw of an improved design made possible by the use of BMG in accordance with the present invention.
- BMG bulk metallic glass
- the invention described and claimed herein involves both apparatus and methods for the use of medical instruments and surgically implantable devices made of bulk metallic glasses during surgical and dental procedures in an intervention MRI in which the MRI guides the surgery in real time.
- the bulk metallic glass has an excellent MRI signature due to its amorphous structure.
- An accurate signature of the metallic implant or instrument is of paramount importance for accurate position information of the instrument or implant device.
- this material Because of the unique atomic structure in an amorphous metallic glass, this material possesses unique magnetic properties that allow the excellent MRI signature.
- the composition tested contains substantial amounts of a ferromagnetic element, Ni.
- Ni a ferromagnetic element
- All of the other elements in this alloy, namely Zr, Ni, Ti, and Al all have susceptibilities substantially higher than that of human tissues thereby producing a positional error in the MRI image.
- Only copper has susceptibility similar to human or animal tissues. Comparison images show that the image of the BMG is better even than a copper alloy. All bulk metallic glasses that are not expressly designed for good soft or hard ferromagnetic properties are expected to have this good MRI signature.
- BMG alloys have a lower modulus and an extremely high elastic limit of about 2% as compared to that of a typical metal, namely about 0.2%.
- Bone has an elastic limit of about 1%.
- BMGs are unique in their ability to flex elastically with the natural bending of bones and so distribute stresses more uniformly. Faster healing rates result from reduced stress shielding effects while minimizing stress concentrators. Because of the unique mechanical properties of the BMGs, screws can have a thinner shank and deeper threads yielding greater holding power. Compared to the lowest modulus developmental titanium alloy, for a given load, the BMG will require 1/4 the cross section to carry the load and will undergo twice the deflection.
- the area will be 1/3 to carry the load and the BMG will have 5 times the deflection.
- Potential applications include fracture fixation screws, rods, pins, hip joint wear surfaces and shafts, aneurysm clips, endodontic files and orthodontic arch wires as well as components of devices such as pacemakers, neurostimulators, medicine-metering pumps, and equipment for remotely-viewed microsurgery.
- FIGS. 1 and 2 show comparison MRI images of a copper alloy and a BMG alloy (Zr -17.9 Cu -14.6 Ni -5.0 Ti -10.0 Al, referred to here as BAM-11 (all compositions are in atomic %) in a flask of salt water showing much less bloom with the BMG alloy.
- the unique magnetic properties of BMG materials yield the ability to fabricate MRI-friendly implant devices as well as a new class of medical instruments for use within the interventional MRI environment.
- Minimally invasive procedures not possible via conventional surgical and dental techniques can be successfully performed through interventional MRI, indicating a critical need for instruments to facilitate these procedures.
- alloy BAM-11 has a yield strength of 1900 MPa, an elastic limit of 2 to 2.2%, Young's Modulus of 90 GPa, Vickers hardness of 590 kg/mm2, and a toughness of 55 to 60 MPa ⁇ square root ⁇ M.
- Table 1 lists mechanical properties of BMGs and compares them to that of the three most commonly used implant materials and to one of the new low-modulus titanium alloys under development.
- Table 2 lists some potential applications for BMG alloys.
- FIG. 3 shows an MRI image of this new alloy compared to a Cu alloy (FIG. 1) and the BAM-11 alloy (FIG. 2). It is evident that without Ni an even better MRI image is obtained. This is attributed to both removal of Nickel and the high percentage of amorphous material.
- Magnetic properties of BMGs are also of interest.
- One of the first commercial uses of rapidly solidified amorphous metals was the Fe—B based alloy for read-write heads and now transformer cores.
- the BMGs also have been found to have unusual magnetic properties including a soft magnetic alloy with zero magnetostriction, a hard magnetic alloy with only 30% Fe that does not saturate at 15 Tesla and has the highest ever measured coercivity, 8.4T, although at liquid helium temperature.
- the magnetic susceptibility is related to density of electronic states at the Fermi energy. Knowledge of the composition dependence of the susceptibility would provide important input in refining models of the amorphous state through first principles calculations of the electronic structure and hence the Fermi energy density of states.
- FIG. 6 shows of a typical conventional orthopedic bone screw design
- FIG. 7 shows an orthopedic bone screw of an improved design made possible by the use of BMG in accordance with the present invention.
- BMG is uniquely suited for MRI compatible medical instruments and surgically implantable devices because of minimal generation of image distortion. Information readily available from MRI images is critically helpful for many types of surgical procedures. MRI images not only function to guide the surgeon's tools to the location of the procedure along the least damaging path, but also to differentiate and define tissue types for facilitation of more efficient and complete procedures. BMG constructed surgical tools allow for a host of procedures to be performed in the MRI environment. BMG offers the best MRI-compatible cutting instruments available. As such, procedures involving the cutting of bone could be MRI-guided. Examples of such procedures include craniotomy procedures involving tumors, embolisms or strokes and orthopedic procedures involving femoral avascular necrosis and vertebral fusions.
- a broad range of other non-medical, even non-biological applications is available to the skilled artisan.
- Remote viewing and control of diagnostic apparatus and apparatus used to maintain or repair any structure which is viewable using MRI technology is enabled by using instruments and implants constructed of BMG, with minimal invasion of the structure, minimal disturbance of the system the structure functions within, and minimal disruption of processes.
- Applications might include diagnosis, maintenance, and repair of such structures as electronic structures, composite structures used in aerospace, marine, and other endeavors, or remotely disarming of hazardous structures such as bombs, mines, and other explosives.
- ferromagnetic metallic glasses are not suitable for MRI imaging and are specifically excluded from the scope of the present invention. See, for example, U.S. Pat. No. 5,976,274 issued on Nov. 2, 1999 to Inoue, et al. entitled “Soft Magnetic Amorphous Alloy and High Hardness Amorphous Alloy and High Hardness Tool Using the Same” and U.S. Pat. No. 4,653,500 issued on Mar. 31, 1987 to Osada, et al. entitled “Electrocardiographic amorphous alloy electrode”.
Landscapes
- Health & Medical Sciences (AREA)
- Chemical & Material Sciences (AREA)
- Engineering & Computer Science (AREA)
- Life Sciences & Earth Sciences (AREA)
- Surgery (AREA)
- Public Health (AREA)
- Animal Behavior & Ethology (AREA)
- Veterinary Medicine (AREA)
- Orthopedic Medicine & Surgery (AREA)
- Materials Engineering (AREA)
- Metallurgy (AREA)
- Mechanical Engineering (AREA)
- General Health & Medical Sciences (AREA)
- Heart & Thoracic Surgery (AREA)
- Organic Chemistry (AREA)
- Vascular Medicine (AREA)
- Epidemiology (AREA)
- Molecular Biology (AREA)
- Medical Informatics (AREA)
- Biomedical Technology (AREA)
- Nuclear Medicine, Radiotherapy & Molecular Imaging (AREA)
- Neurology (AREA)
- Physics & Mathematics (AREA)
- Optics & Photonics (AREA)
- Ceramic Engineering (AREA)
- Inorganic Chemistry (AREA)
- Materials For Medical Uses (AREA)
Abstract
MRI-compatible medical instruments and appliances are made using bulk metallic glass alloys. MRI-guided methods include the use of articles that include bulk metallic glass alloys.
Description
- This application is a divisional application of U.S. patent application Ser. No. 09/799,445, filed on Mar. 5, 2001, the entirety of which is incorporated herein by reference.
- [0002] This invention was made with Government support under Contract No. DE-AC05-00OR22725 awarded by the United States Department of Energy. The Government has certain rights in the invention.
- The present invention relates to medical, surgical, and dental hardware, especially medical instruments and biomedical appliances, and particularly to those instruments and appliances at least partially constructed of a bulk metallic glass (BMG), and to methods of using the same.
- The concept of magnetic susceptibility is central to many current research and development activities in magnetic resonance imaging (MRI). For example, the development of MR-guided surgery has created a need for surgical instruments and other devices with susceptibility tailored to the MR environment; susceptibility effects can lead to position errors of up to several millimeters in MR-guided stereotactic surgery; and the variation of magnetic susceptibility on a microscopic scale within tissues contributes to MR contrast and is the basis of functional MRI. The magnetic aspects of MR compatibility are discussed in terms of two levels of acceptability: Materials with the first kind of magnetic field compatibility are such that magnetic forces and torques do not interfere significantly when the materials are used within the magnetic field of the scanner; materials with the second kind of magnetic field compatibility meet the more demanding requirement that they produce only negligible artifacts within the MR image and their effect on the positional accuracy of features within the image is negligible or can readily be corrected. Several materials exhibiting magnetic field compatibility of the second kind have been studied and a group of materials that produce essentially no image distortion, even when located directly within the imaging field of view, is identified. Because of demagnetizing effects, the shape and orientation, as well as the susceptibility, of articles within and adjacent to the imaging region is important in MRI, but the use of literature values for the susceptibility of materials is often difficult because of inconsistent traditions in the definitions and units used for magnetic parameters—particularly susceptibility.
- Thus, methods and apparatus have long been sought to permit surgical procedures involving surgical instruments and/or surgically implanted appliances to be guided or monitored by MRI.
- For this to be possible, a new implant material has long been needed which has a low MRI signature. In addition, it would be desirable for such materials to also have high hardness, tensile strength, and toughness. A desirable material would have a lower elastic modulus and an extremely high elastic limit of about 2% compared to that of a typical metal, namely about 0.2%. Bone has an elastic limit of about 1%. Such material would be unique in its ability to flex elastically with the natural bending of the bones and so distribute stresses more uniformly. Faster healing rates would result from reduced stress shielding effects while minimizing stress concentrators. Because of these unique mechanical properties, screws could have a thinner shank and deeper threads yielding greater holding power. Applications where such material is desirable would include such as fracture fixation screws, rods, pins, knee and hip joint wear surfaces and shafts, and aneurysm clips. A large variety of other applications, changes and modifications would be obvious to those skilled in the art.
- Current implant materials produce a distortion or blooming (enlargement) in the MRI image. Larger implants are even internally heated during an MRI. This is especially important for aneurysm clips where later imaging is often needed and where no movement of the clip as a result of an MRI is essential.
- Definitions:
- A bulk metallic glass (BMG) is defined for purposes herein as an amorphous metallic alloy that is cast in bulk form. A BMG is known to be inclusive of amorphous thin-film materials such as those that are typically deposited on surfaces.
- A medical instrument is defined for purposes herein as any device used by medical and/or dental personnel in any surgical and/or dental procedure.
- A biomedical appliance is defined for purposes herein as any medically functional device that is configured for disposition on or inside a living body, including surgically implanted orthopedic devices, dental implants, and the like.
- Accordingly, objects of the present invention include at least the following:
- provision of new and improved medical instruments and biomedical appliances having at least one of: desirably high elastic limit, hardness, strength, toughness, and ability to hold a cutting edge;
- provision of new and improved MRI-compatible medical instruments and surgically implantable orthopedic appliances that allow surgeries to be guided in real time by MRI imaging;
- provision of new and improved surgically implantable appliances (e.g., orthopedic, endodontic, etc.) with mechanical properties compatible with those of bone, low corrosion rate, and good biocompatibility (lack of rejection by human or animal tissue);
- provision of new and improved articles configured for use as tools, instruments, and parts used for maintaining, inspecting, modifying, operating, or exploring both man-made and naturally-occurring structures which are internally or externally viewable by MRI; and
- provision of new and improved methods of performing medical and dental procedures utilizing the aforementioned instruments and appliances.
- Further and other objects of the present invention will become apparent from the description contained herein.
- In accordance with one aspect of the present invention, the foregoing and other objects are achieved by an article, at least a portion of which includes a bulk metallic glass having magnetic properties suitable for producing an MRI image, the article having a shape suitable for producing an MRI image and being configured for use as a medical instrument and/or a biomedical appliance.
- In accordance with another aspect of the present invention, a method of carrying out a medical procedure includes the steps of: providing a medical instrument or a biomedical appliance, at least a portion of which includes a bulk metallic glass; and using the medical instrument or biomedical appliance to carry out a medical procedure.
- In accordance with a further aspect of the present invention, a method of carrying out an MRI-guided procedure includes the steps of: providing an article, at least a portion of which includes a bulk metallic glass; and using the article to carry out an MRI-guided procedure.
- In the drawings:
- FIG. 1 is a full-scale MRI image of a 7 mm diameter rod of a conventional copper alloy (Cu-4 Cri-2 Nb) in salt water.
- FIG. 2 is a full-scale MRI image of a 7 mm diameter rod of BAM-11 in salt water in accordance with the present invention.
- FIG. 3 is a full-scale MRI image of a 7 mm diameter rod of Ni-free BMG in salt water in accordance with the present invention.
- FIG. 4 is a full-scale MRI image of a bullet nosed 6 mm diameter rod of Ti-6 Al-4.
- FIG. 5 is a full-scale MRI image of a bullet nosed 6 mm diameter rod of BAM-11 in salt water in accordance with the present invention.
- FIG. 6 is a view of a typical conventional orthopedic bone screw.
- FIG. 7 is a view of an orthopedic bone screw of an improved design made possible by the use of BMG in accordance with the present invention.
- For a better understanding of the present invention, together with other and further articles, advantages and capabilities thereof, reference is made to the following disclosure and appended claims in connection with the above-described drawings.
- It has been discovered that certain bulk metallic glass (BMG) alloys have a very low MRI signature. BMG materials have high hardness, tensile strength, and toughness. The present invention is based on the discovery that the unique properties of BMG alloys make them especially suitable for biomedical implant applications as well as for medical instruments.
- The invention described and claimed herein involves both apparatus and methods for the use of medical instruments and surgically implantable devices made of bulk metallic glasses during surgical and dental procedures in an intervention MRI in which the MRI guides the surgery in real time. The bulk metallic glass has an excellent MRI signature due to its amorphous structure. An accurate signature of the metallic implant or instrument is of paramount importance for accurate position information of the instrument or implant device.
- It is recognized that a large range of specific compositions of BMG are known to and may be used by the skilled artisan in this invention. . For example, see U.S. Pat. No. 5,735,975 issued on Apr. 7, 1998 to Lin, et al. entitled “Quinary Metallic Glass Alloys” and U.S. Pat. No. 5,803,996 issued on Sep. 8, 1998 to Inoue, et al. entitled “Rod-Shaped or Tubular Amorphous Zr Alloy Made by Die Casting and Method for Manufacturing said Amorphous Zr Alloy”. The patents of both Lin and Inoue are incorporated herein by reference. The specific composition of the alloy used in examples described herein is Zr -17.9 Cu -14.6 Ni -5 Ti -10 Al (at %).
- It is further recognized that attempts have been made in the past to provide medical instruments which cause reduced or enhanced artifact on diagnostic images such as MRI images. One such example is U.S. Pat. No. 5,895,401, issued Apr. 20, 1999, “Controlled-Artifact Magnetic Resonance Instruments” by Daum et al. However, Daum teaches a crystalline alloy comprising at least 85% titanium and does not teach an alloy which is amorphous such as the bulk metallic glass taught herein. Conversely, the bulk metallic glass as taught herein contains no more than 12 at. % titanium. Thus the present invention and Daum teach grossly different materials.
- Because of the unique atomic structure in an amorphous metallic glass, this material possesses unique magnetic properties that allow the excellent MRI signature.
- The composition tested contains substantial amounts of a ferromagnetic element, Ni. In a crystalline structure, the presence of Ni produces substantial blooming in an MRI image. All of the other elements in this alloy, namely Zr, Ni, Ti, and Al all have susceptibilities substantially higher than that of human tissues thereby producing a positional error in the MRI image. Only copper has susceptibility similar to human or animal tissues. Comparison images show that the image of the BMG is better even than a copper alloy. All bulk metallic glasses that are not expressly designed for good soft or hard ferromagnetic properties are expected to have this good MRI signature.
- BMG alloys have a lower modulus and an extremely high elastic limit of about 2% as compared to that of a typical metal, namely about 0.2%. Bone has an elastic limit of about 1%. BMGs are unique in their ability to flex elastically with the natural bending of bones and so distribute stresses more uniformly. Faster healing rates result from reduced stress shielding effects while minimizing stress concentrators. Because of the unique mechanical properties of the BMGs, screws can have a thinner shank and deeper threads yielding greater holding power. Compared to the lowest modulus developmental titanium alloy, for a given load, the BMG will require 1/4 the cross section to carry the load and will undergo twice the deflection. Compared to stainless steel, the area will be 1/3 to carry the load and the BMG will have 5 times the deflection. Potential applications include fracture fixation screws, rods, pins, hip joint wear surfaces and shafts, aneurysm clips, endodontic files and orthodontic arch wires as well as components of devices such as pacemakers, neurostimulators, medicine-metering pumps, and equipment for remotely-viewed microsurgery.
- FIGS. 1 and 2 show comparison MRI images of a copper alloy and a BMG alloy (Zr -17.9 Cu -14.6 Ni -5.0 Ti -10.0 Al, referred to here as BAM-11 (all compositions are in atomic %) in a flask of salt water showing much less bloom with the BMG alloy. This is a surprising result, since one would expect the susceptibility to be close to that of zirconium as the major constituent and the nickel in the alloy should make it higher. While copper is the metal with a susceptibility closest to that of living human or animal tissue, it is too soft for many uses, especially medical instruments and implants. The toxicity of beryllium all but prohibits the use of the harder Be—Cu alloys. For these reasons, the unique magnetic properties of BMG materials yield the ability to fabricate MRI-friendly implant devices as well as a new class of medical instruments for use within the interventional MRI environment. Minimally invasive procedures not possible via conventional surgical and dental techniques can be successfully performed through interventional MRI, indicating a critical need for instruments to facilitate these procedures.
- Prior to the recent development of these bulk metallic glasses, rapid solidification such as melt spinning or gas atomization producing thin ribbons or powders (<150 mm) was required to achieve the sufficiently high cooling rates necessary for glass formation. Due to their unique amorphous microstructure with a number of different elements present, the BMGs exhibit a number of exceptional properties. For example, alloy BAM-11 has a yield strength of 1900 MPa, an elastic limit of 2 to 2.2%, Young's Modulus of 90 GPa, Vickers hardness of 590 kg/mm2, and a toughness of 55 to 60 MPa{square root}M. Table 1 lists mechanical properties of BMGs and compares them to that of the three most commonly used implant materials and to one of the new low-modulus titanium alloys under development. Table 2 lists some potential applications for BMG alloys.
- (Table 1 begins next page)
TABLE 1 Mechanical properties of bulk metallic glass compared to the three leading implant materials Experimen- tal Ti- BMG Ti-6Al- 35Nb-5 Property (BAM 11) Co-Cr 4V 316L-CW Bone Ta-7 Zr Tensile Yield 1900 450 830 690 Compressive 547 Strength, MPa 130-150 (cortical) Elastic Strain 2-2.2 0.18 0.67 0.34 1% 0.9 Limit, % Plastic Strain 1 8 10 12 — 19 to failure, % Young's Modulus, 90 248 124 200 17* 55 GPa Hardness, Vickers, 590 350-390 320 365 — — Kg/mm2 Toughness, Mpa m1/2 55-60 — 57 100 — — Fatigue load for failure — 310 520 240 — 265 at 107 cycles, MPa Density, g/cc 5.9 8.5 4.4 8 — — Biocompatibility Good, Good/ Good Good/ — good initial question- question- evaluation able able Magnetic Very Probably Ti is Austenitic is Human tissue Probably Susceptibility compat- ferro- 182 × 3-6 × is −11 to similar to ible magnetic 10−6 10−3 −7 × Ti CW is 10−6 ferro- magnetic -
TABLE 2 Summary of potential health field application for bulk metallic glasses Critical Property Application Unique BMG Property Measurement Comments Fracture Fixation and Fusion Lower modulus, high Fatigue, crack initiation at Some plates require plastic Plates strength drill holes, corrosion, deformation to adjust fit at biocompatability time of insertion which BMG could not accommodate. Screws Toughness, high strength Fatigue, corrosion, Less wear debris will lead to biocompatability less immune system response. Hip joints-wear surface Low coefficient of friction, Wear, corrosion, Less wear debris will lead to hardness biocompatability less immune system response. Hip joints-shaft Low modulus Corrosion, biocompatability Lower modulus distributes load better Cutting tools, scalpel, bone Toughness, high elastic Edge holding, susceptibility May have longer life than biopsy osteotome, limit, hardness SS that quickly dull. endodontic files Aneurysm clips High elastic limit, good MRI Creep, corrosion, Can do emergency MRI signature biocompatability later without possible fatal results Orthodontist Arch Wires High elastic limit, low Processing to preform or Need to slide in mounts coefficient of friction cast to proper shape (hard enough to be compatible with ceramic brackets), maintain force over large elastic deformations MRI interventional surgical Good MRI signature, Edge holding, susceptibility A unique material for the instruments toughness, hardness next generation of surgical care in an interventional MRI - Initial screening tests on BAM-11 performed at the Biomaterials and Orthopedic Research Department at the University of Mississippi Medical Center have shown biocompatibility comparable to current implant materials. For initial biocompatibility screening, two cell lines were selected: microphage and fibroblast. Because these cell types are key in inflammation and encapsulation processes they are generally predictive of soft tissue biocompatibility. Four analyses of biocompatibility were conducted for each cell type: 1) cellular viability, 2) catalase activity, 3) TNF beta cytokine concentration and 4) lactate dehydrogenase concentration. The results are presented in Table 3.
TABLE 3 Initial Biocompatability Tests of BAM-11 compared to two control specimens* BAM-11 Ti, commercially pure Polyethylene Macrophage Viability, 85 90 91 % Macrophage/Catalase 18 22 12 Activity/standardized activity/protein quan- tity Macrophage/Lactate 6 8 6 Dehydrogenase, stan- dardized activity/pro- tein quantity Macrophage/Cytokine, 19 18 11 picograms/protein quantity Fibroblast Viability, % 95 90 92 survival Fibroblasts/Catalase 9 5 5.5 Activity standardized activity/protein quantity Fibroblast/Lactate 6.5 4 5 Dehydrogenase, stan- dardized activity/ protein quantity - Screening tests on BAM-11 specimens performed at the Biomaterials and Orthopedic Research Department at the University of Mississippi Medical Center have shown corrosion resistance comparable to current implant materials. All specimens were wet ground with SiC paper, 80, 240, 320, 600, and 1500 grit followed by ultrasonic cleaning in distilled water for 5 min. The titanium was additionally passivated in 40% HNO3 for 30 min according to an ASTM standard. Cyclic polarization tests were conducted on triplicate samples of the alloys in Ringer's solution (9.0 g/L-NaCl, 0.42 g/L-KCl, 0.25 g/L-CaCl2). Specimens were allowed to reach an open-circuit potential (Ecorr) for a period of one hour. A potential scan increasing at a rate of 0.1667 mV/s (ASTM G5) was then initiated at 100 mV below Ecorr and continued until a current threshold of 1×10-2A/cm2 was reached. At this point the scan was reversed and decreased in the same rate until Ecorr was reached. The results are presented in Table 4.
TABLE 4 Initial corrosion tests of BAM-11 Alloy Ecorr (mV) Rbr (mV) Icorr (na/cm2) Titanium −51 ± 61.5 None recorded 8.2 ± 3.4 316L SS −72.7 ± 20 323 ± 66.4 14.1 ± 6.7 BAM11 −228.3 ± 38.6 −65.3 ± 53.0 56.1 ± 32.8 - While currently used surgically implantable orthopedic appliances contain nickel, we have developed compositions that eliminate nickel due to long-term biocompatibility concerns.
- Alloys with composition of Zr-32.5 Cu-5 Ti-10 Al (at. %) were arc cast into a water-cooled copper mold and as cast were 99% amorphous. FIG. 3 shows an MRI image of this new alloy compared to a Cu alloy (FIG. 1) and the BAM-11 alloy (FIG. 2). It is evident that without Ni an even better MRI image is obtained. This is attributed to both removal of Nickel and the high percentage of amorphous material.
- It is believed that removal of the nickel is not necessary for use of the BMGs as instruments and temporary fixation devices such as immobilizing screws. Instruments such as bone biopsy tools, scalpel blades, and the like have been fabricated by well-known, conventional techniques such as casting, machining, laser welding, grinding, and polishing.
- Magnetic properties of BMGs are also of interest. One of the first commercial uses of rapidly solidified amorphous metals was the Fe—B based alloy for read-write heads and now transformer cores. The BMGs also have been found to have unusual magnetic properties including a soft magnetic alloy with zero magnetostriction, a hard magnetic alloy with only 30% Fe that does not saturate at 15 Tesla and has the highest ever measured coercivity, 8.4T, although at liquid helium temperature. In general, the magnetic susceptibility is related to density of electronic states at the Fermi energy. Knowledge of the composition dependence of the susceptibility would provide important input in refining models of the amorphous state through first principles calculations of the electronic structure and hence the Fermi energy density of states. The preliminary result presented here concerning a good MRI image was surprising considering the composition and especially the presence of nickel in the alloy. Actual direct measurements of the magnetic susceptibility by a squid magnetometer yield 109×10-6 for BAM-11 versus 182×10-6 for Ti-6 Al-4 V agreeing with the MRI results.
- Mechanical properties of the BMG described herein are superior for many surgically implantable appliances and medical instruments than currently used materials. These properties include higher yield strength, higher elastic strain limit, lower Young's modulus (better for reducing stress shielding), higher hardness, and comparable toughness. For example, Because of these unique mechanical properties, screws could have a thinner shank and deeper threads yielding greater holding power. For example, FIG. 6 shows of a typical conventional orthopedic bone screw design, while FIG. 7 shows an orthopedic bone screw of an improved design made possible by the use of BMG in accordance with the present invention.
- BMG is uniquely suited for MRI compatible medical instruments and surgically implantable devices because of minimal generation of image distortion. Information readily available from MRI images is critically helpful for many types of surgical procedures. MRI images not only function to guide the surgeon's tools to the location of the procedure along the least damaging path, but also to differentiate and define tissue types for facilitation of more efficient and complete procedures. BMG constructed surgical tools allow for a host of procedures to be performed in the MRI environment. BMG offers the best MRI-compatible cutting instruments available. As such, procedures involving the cutting of bone could be MRI-guided. Examples of such procedures include craniotomy procedures involving tumors, embolisms or strokes and orthopedic procedures involving femoral avascular necrosis and vertebral fusions.
- A broad range of other non-medical, even non-biological applications is available to the skilled artisan. Remote viewing and control of diagnostic apparatus and apparatus used to maintain or repair any structure which is viewable using MRI technology is enabled by using instruments and implants constructed of BMG, with minimal invasion of the structure, minimal disturbance of the system the structure functions within, and minimal disruption of processes. Applications might include diagnosis, maintenance, and repair of such structures as electronic structures, composite structures used in aerospace, marine, and other endeavors, or remotely disarming of hazardous structures such as bombs, mines, and other explosives.
- It should be noted that ferromagnetic metallic glasses are not suitable for MRI imaging and are specifically excluded from the scope of the present invention. See, for example, U.S. Pat. No. 5,976,274 issued on Nov. 2, 1999 to Inoue, et al. entitled “Soft Magnetic Amorphous Alloy and High Hardness Amorphous Alloy and High Hardness Tool Using the Same” and U.S. Pat. No. 4,653,500 issued on Mar. 31, 1987 to Osada, et al. entitled “Electrocardiographic amorphous alloy electrode”.
- While there has been shown and described what are at present considered the preferred embodiments of the invention, it will be obvious to those skilled in the art that various changes and modifications can be made therein without departing from the scope of the inventions defined by the appended claims.
Claims (7)
1. A method of carrying out a medical procedure comprising the steps of:
a. providing an article selected from the group consisting of medical instruments and biomedical appliances, at least a portion of said article comprising a bulk metallic glass; and
b. using said article to carry out a medical procedure wherein an MRI image of at least part of said article is made.
2. A method in accordance with claim 1 wherein said medical procedure is an orthopedic procedure.
3. A method in accordance with claim 1 wherein said medical procedure is an endodontic procedure.
4. A method in accordance with claim 1 wherein said medical procedure comprises an MRI-guided procedure.
5. A method of carrying out an MRI-guided procedure comprising the steps of:
a. providing an article, at least a portion of said article comprising a bulk metallic glass; and
b. using said article to carry out an MRI-guided procedure.
6. A method in accordance with claim 5 wherein said MRI-guided procedure further comprises a medical procedure.
7. A method in accordance with claim 5 wherein said MRI-guided procedure further comprises a non-medical procedure.
Priority Applications (1)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
US10/256,751 US20030111142A1 (en) | 2001-03-05 | 2002-09-27 | Bulk metallic glass medical instruments, implants, and methods of using same |
Applications Claiming Priority (2)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
US09/799,445 US20020162605A1 (en) | 2001-03-05 | 2001-03-05 | Bulk metallic glass medical instruments, implants, and methods of using same |
US10/256,751 US20030111142A1 (en) | 2001-03-05 | 2002-09-27 | Bulk metallic glass medical instruments, implants, and methods of using same |
Related Parent Applications (1)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
US09/799,445 Division US20020162605A1 (en) | 2001-03-05 | 2001-03-05 | Bulk metallic glass medical instruments, implants, and methods of using same |
Publications (1)
Publication Number | Publication Date |
---|---|
US20030111142A1 true US20030111142A1 (en) | 2003-06-19 |
Family
ID=25175927
Family Applications (2)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
US09/799,445 Abandoned US20020162605A1 (en) | 2001-03-05 | 2001-03-05 | Bulk metallic glass medical instruments, implants, and methods of using same |
US10/256,751 Abandoned US20030111142A1 (en) | 2001-03-05 | 2002-09-27 | Bulk metallic glass medical instruments, implants, and methods of using same |
Family Applications Before (1)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
US09/799,445 Abandoned US20020162605A1 (en) | 2001-03-05 | 2001-03-05 | Bulk metallic glass medical instruments, implants, and methods of using same |
Country Status (1)
Country | Link |
---|---|
US (2) | US20020162605A1 (en) |
Cited By (19)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
EP1632584A1 (en) * | 2004-09-06 | 2006-03-08 | Eidgenössische Technische Hochschule Zürich | Amorphous alloys on the base of Zr and their use |
US20060210880A1 (en) * | 1993-11-19 | 2006-09-21 | Medtronic, Inc. | Current collector |
US20070248881A1 (en) * | 2006-03-24 | 2007-10-25 | Medtronic, Inc. | Implantable medical device and lithium battery |
US20080103543A1 (en) * | 2006-10-31 | 2008-05-01 | Medtronic, Inc. | Implantable medical device with titanium alloy housing |
US20080125848A1 (en) * | 2006-06-30 | 2008-05-29 | Kusleika Richard S | Medical devices with amorphous metals, and methods therefor |
US20090139612A1 (en) * | 2007-11-30 | 2009-06-04 | Kun Lu | Zr-based amorphous alloy and a preparing method thereof |
US20090288741A1 (en) * | 2008-03-21 | 2009-11-26 | Faliang Zhang | Amorphous Alloy and A Preparation Method Thereof |
US8014867B2 (en) | 2004-12-17 | 2011-09-06 | Cardiac Pacemakers, Inc. | MRI operation modes for implantable medical devices |
US8032228B2 (en) | 2007-12-06 | 2011-10-04 | Cardiac Pacemakers, Inc. | Method and apparatus for disconnecting the tip electrode during MRI |
US8086321B2 (en) | 2007-12-06 | 2011-12-27 | Cardiac Pacemakers, Inc. | Selectively connecting the tip electrode during therapy for MRI shielding |
US8160717B2 (en) | 2008-02-19 | 2012-04-17 | Cardiac Pacemakers, Inc. | Model reference identification and cancellation of magnetically-induced voltages in a gradient magnetic field |
US8311637B2 (en) | 2008-02-11 | 2012-11-13 | Cardiac Pacemakers, Inc. | Magnetic core flux canceling of ferrites in MRI |
US8527046B2 (en) | 2000-04-20 | 2013-09-03 | Medtronic, Inc. | MRI-compatible implantable device |
US8565874B2 (en) | 2009-12-08 | 2013-10-22 | Cardiac Pacemakers, Inc. | Implantable medical device with automatic tachycardia detection and control in MRI environments |
US8571661B2 (en) | 2008-10-02 | 2013-10-29 | Cardiac Pacemakers, Inc. | Implantable medical device responsive to MRI induced capture threshold changes |
US8639331B2 (en) | 2009-02-19 | 2014-01-28 | Cardiac Pacemakers, Inc. | Systems and methods for providing arrhythmia therapy in MRI environments |
US9357996B2 (en) * | 2010-09-08 | 2016-06-07 | DePuy Synthes Products, Inc. | Fixation device with magnesium core |
US10694950B2 (en) | 2016-10-13 | 2020-06-30 | International Business Machines Corporation | Probe structure for physiological measurements using surface enhanced Raman spectroscopy |
US10694951B2 (en) | 2016-10-13 | 2020-06-30 | International Business Machines Corporation | Probe structure for physiological measurements using surface enhanced Raman spectroscopy |
Families Citing this family (24)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
US20040267349A1 (en) * | 2003-06-27 | 2004-12-30 | Kobi Richter | Amorphous metal alloy medical devices |
US8382821B2 (en) | 1998-12-03 | 2013-02-26 | Medinol Ltd. | Helical hybrid stent |
AU2003254319A1 (en) * | 2002-08-05 | 2004-02-23 | Liquidmetal Technologies | Metallic dental prostheses made of bulk-solidifying amorphous alloys and method of making such articles |
WO2004016197A1 (en) * | 2002-08-19 | 2004-02-26 | Liquidmetal Technologies, Inc. | Medical implants |
AU2003287682A1 (en) * | 2002-11-18 | 2004-06-15 | Liquidmetal Technologies | Amorphous alloy stents |
WO2004112862A1 (en) * | 2003-06-26 | 2004-12-29 | Eidgenössische Technische Hochschule Zürich | Prosthesis and method for the production thereof |
US9039755B2 (en) | 2003-06-27 | 2015-05-26 | Medinol Ltd. | Helical hybrid stent |
US9155639B2 (en) | 2009-04-22 | 2015-10-13 | Medinol Ltd. | Helical hybrid stent |
US7901447B2 (en) | 2004-12-29 | 2011-03-08 | Boston Scientific Scimed, Inc. | Medical devices including a metallic film and at least one filament |
US8992592B2 (en) | 2004-12-29 | 2015-03-31 | Boston Scientific Scimed, Inc. | Medical devices including metallic films |
US8632580B2 (en) | 2004-12-29 | 2014-01-21 | Boston Scientific Scimed, Inc. | Flexible medical devices including metallic films |
US8998973B2 (en) | 2004-03-02 | 2015-04-07 | Boston Scientific Scimed, Inc. | Medical devices including metallic films |
US8591568B2 (en) | 2004-03-02 | 2013-11-26 | Boston Scientific Scimed, Inc. | Medical devices including metallic films and methods for making same |
US7854760B2 (en) | 2005-05-16 | 2010-12-21 | Boston Scientific Scimed, Inc. | Medical devices including metallic films |
US20070166349A1 (en) * | 2006-01-13 | 2007-07-19 | Dawn White | Orthopaedic implants fabricated from amorphous or partially amorphous calcium-based metal alloys |
US20070178383A1 (en) * | 2006-01-31 | 2007-08-02 | Viavattine Joseph J | Current collector |
US8267918B2 (en) * | 2006-03-14 | 2012-09-18 | Kci Licensing, Inc. | System and method for percutaneously administering reduced pressure treatment using a flowable manifold |
US20070217163A1 (en) * | 2006-03-15 | 2007-09-20 | Wilson Greatbatch | Implantable medical electronic device with amorphous metallic alloy enclosure |
WO2008113192A1 (en) * | 2007-03-16 | 2008-09-25 | Bien-Air Holding S.A. | Handheld part for dental or surgical use |
JP4783934B2 (en) * | 2009-06-10 | 2011-09-28 | 株式会社丸ヱム製作所 | Metal glass fastening screw |
TWI554391B (en) * | 2015-05-15 | 2016-10-21 | 國立臺灣科技大學 | Thin film metallic glass |
US11219744B2 (en) * | 2017-04-21 | 2022-01-11 | Medtronic Vascular, Inc. | Push wire for endoluminal medical device |
US20200282222A1 (en) * | 2019-02-06 | 2020-09-10 | Liquidmetal Technologies, Inc. | Implantable medical device with bulk metallic glass enclosure |
CN113528883B (en) * | 2021-07-13 | 2022-05-20 | 上海理工大学 | Copper-niobium alloy for medical biopsy puncture needle |
Citations (37)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
US3940293A (en) * | 1972-12-20 | 1976-02-24 | Allied Chemical Corporation | Method of producing amorphous cutting blades |
US4116682A (en) * | 1976-12-27 | 1978-09-26 | Polk Donald E | Amorphous metal alloys and products thereof |
US4318738A (en) * | 1978-02-03 | 1982-03-09 | Shin-Gijutsu Kaihatsu Jigyodan | Amorphous carbon alloys and articles manufactured from said alloys |
US4557766A (en) * | 1984-03-05 | 1985-12-10 | Standard Oil Company | Bulk amorphous metal alloy objects and process for making the same |
US4653500A (en) * | 1985-06-18 | 1987-03-31 | Fukuda Denshi Co., Ltd. | Electrocardiographic amorphous alloy electrode |
US4762677A (en) * | 1987-11-03 | 1988-08-09 | Allied-Signal Inc. | Method of preparing a bulk amorphous metal article |
US4762678A (en) * | 1987-11-03 | 1988-08-09 | Allied-Signal Inc. | Method of preparing a bulk amorphous metal article |
US4827931A (en) * | 1986-01-13 | 1989-05-09 | Longmore Donald B | Surgical catheters with suturing device and NMR opaque material |
US4880482A (en) * | 1987-04-28 | 1989-11-14 | Mitsui Engineering & Shipbuilding Co., Ltd. | Highly corrosion-resistant amorphous alloy |
US5169597A (en) * | 1989-12-21 | 1992-12-08 | Davidson James A | Biocompatible low modulus titanium alloy for medical implants |
US5190546A (en) * | 1983-10-14 | 1993-03-02 | Raychem Corporation | Medical devices incorporating SIM alloy elements |
US5290266A (en) * | 1992-08-14 | 1994-03-01 | General Electric Company | Flexible coating for magnetic resonance imaging compatible invasive devices |
US5324368A (en) * | 1991-05-31 | 1994-06-28 | Tsuyoshi Masumoto | Forming process of amorphous alloy material |
US5372660A (en) * | 1993-08-26 | 1994-12-13 | Smith & Nephew Richards, Inc. | Surface and near surface hardened medical implants |
US5380375A (en) * | 1992-04-07 | 1995-01-10 | Koji Hashimoto | Amorphous alloys resistant against hot corrosion |
US5460663A (en) * | 1991-10-16 | 1995-10-24 | Ykk Corporation | High corrosion resistant amorphous alloys |
US5482580A (en) * | 1994-06-13 | 1996-01-09 | Amorphous Alloys Corp. | Joining of metals using a bulk amorphous intermediate layer |
US5549797A (en) * | 1991-05-15 | 1996-08-27 | Koji Hashimoto | Highly corrosion-resistant amorphous alloys |
US5578359A (en) * | 1994-11-29 | 1996-11-26 | Hewlett Packard Company | Magnetic shielding garment for electro-biologic measurements |
US5618359A (en) * | 1995-02-08 | 1997-04-08 | California Institute Of Technology | Metallic glass alloys of Zr, Ti, Cu and Ni |
US5647361A (en) * | 1992-09-28 | 1997-07-15 | Fonar Corporation | Magnetic resonance imaging method and apparatus for guiding invasive therapy |
US5706814A (en) * | 1992-12-16 | 1998-01-13 | Kabushiki Kaisha Egawa | Method of determining a position of a probe relative to a tooth using MRI |
US5711363A (en) * | 1996-02-16 | 1998-01-27 | Amorphous Technologies International | Die casting of bulk-solidifying amorphous alloys |
US5735975A (en) * | 1996-02-21 | 1998-04-07 | California Institute Of Technology | Quinary metallic glass alloys |
US5769861A (en) * | 1995-09-28 | 1998-06-23 | Brainlab Med. Computersysteme Gmbh | Method and devices for localizing an instrument |
US5772803A (en) * | 1996-08-26 | 1998-06-30 | Amorphous Technologies International | Torsionally reacting spring made of a bulk-solidifying amorphous metallic alloy |
US5797443A (en) * | 1996-09-30 | 1998-08-25 | Amorphous Technologies International | Method of casting articles of a bulk-solidifying amorphous alloy |
US5803996A (en) * | 1995-01-25 | 1998-09-08 | Research Development Corporation Of Japan | Rod-shaped or tubular amorphous Zr alloy made by die casting and method for manufacturing said amorphous Zr alloy |
US5823778A (en) * | 1996-06-14 | 1998-10-20 | The United States Of America As Represented By The Secretary Of The Air Force | Imaging method for fabricating dental devices |
US5842858A (en) * | 1995-05-11 | 1998-12-01 | Artma Biomedical, Inc. | Method of imaging a person's jaw and a model therefor |
US5895401A (en) * | 1995-08-24 | 1999-04-20 | Daum Gmbh | Controlled-artifact magnetic resonance instruments |
US5976274A (en) * | 1997-01-23 | 1999-11-02 | Akihisa Inoue | Soft magnetic amorphous alloy and high hardness amorphous alloy and high hardness tool using the same |
US5980652A (en) * | 1996-05-21 | 1999-11-09 | Research Developement Corporation Of Japan | Rod-shaped or tubular amorphous Zr alloy made by die casting and method for manufacturing said amorphous Zr alloy |
US6077367A (en) * | 1997-02-19 | 2000-06-20 | Alps Electric Co., Ltd. | Method of production glassy alloy |
US6306228B1 (en) * | 1998-07-08 | 2001-10-23 | Japan Science And Technology Corporation | Method of producing amorphous alloy excellent in flexural strength and impact strength |
US6675033B1 (en) * | 1999-04-15 | 2004-01-06 | Johns Hopkins University School Of Medicine | Magnetic resonance imaging guidewire probe |
US6714809B2 (en) * | 2000-11-20 | 2004-03-30 | Surgi-Vision, Inc. | Connector and guidewire connectable thereto |
-
2001
- 2001-03-05 US US09/799,445 patent/US20020162605A1/en not_active Abandoned
-
2002
- 2002-09-27 US US10/256,751 patent/US20030111142A1/en not_active Abandoned
Patent Citations (38)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
US3940293A (en) * | 1972-12-20 | 1976-02-24 | Allied Chemical Corporation | Method of producing amorphous cutting blades |
US4116682A (en) * | 1976-12-27 | 1978-09-26 | Polk Donald E | Amorphous metal alloys and products thereof |
US4318738A (en) * | 1978-02-03 | 1982-03-09 | Shin-Gijutsu Kaihatsu Jigyodan | Amorphous carbon alloys and articles manufactured from said alloys |
US5190546A (en) * | 1983-10-14 | 1993-03-02 | Raychem Corporation | Medical devices incorporating SIM alloy elements |
US4557766A (en) * | 1984-03-05 | 1985-12-10 | Standard Oil Company | Bulk amorphous metal alloy objects and process for making the same |
US4653500A (en) * | 1985-06-18 | 1987-03-31 | Fukuda Denshi Co., Ltd. | Electrocardiographic amorphous alloy electrode |
US4827931A (en) * | 1986-01-13 | 1989-05-09 | Longmore Donald B | Surgical catheters with suturing device and NMR opaque material |
US4880482A (en) * | 1987-04-28 | 1989-11-14 | Mitsui Engineering & Shipbuilding Co., Ltd. | Highly corrosion-resistant amorphous alloy |
US4762677A (en) * | 1987-11-03 | 1988-08-09 | Allied-Signal Inc. | Method of preparing a bulk amorphous metal article |
US4762678A (en) * | 1987-11-03 | 1988-08-09 | Allied-Signal Inc. | Method of preparing a bulk amorphous metal article |
US5169597A (en) * | 1989-12-21 | 1992-12-08 | Davidson James A | Biocompatible low modulus titanium alloy for medical implants |
US5549797A (en) * | 1991-05-15 | 1996-08-27 | Koji Hashimoto | Highly corrosion-resistant amorphous alloys |
US5324368A (en) * | 1991-05-31 | 1994-06-28 | Tsuyoshi Masumoto | Forming process of amorphous alloy material |
US6027586A (en) * | 1991-05-31 | 2000-02-22 | Tsuyoshi Masumoto | Forming process of amorphous alloy material |
US5460663A (en) * | 1991-10-16 | 1995-10-24 | Ykk Corporation | High corrosion resistant amorphous alloys |
US5380375A (en) * | 1992-04-07 | 1995-01-10 | Koji Hashimoto | Amorphous alloys resistant against hot corrosion |
US5290266A (en) * | 1992-08-14 | 1994-03-01 | General Electric Company | Flexible coating for magnetic resonance imaging compatible invasive devices |
US5647361A (en) * | 1992-09-28 | 1997-07-15 | Fonar Corporation | Magnetic resonance imaging method and apparatus for guiding invasive therapy |
US5706814A (en) * | 1992-12-16 | 1998-01-13 | Kabushiki Kaisha Egawa | Method of determining a position of a probe relative to a tooth using MRI |
US5372660A (en) * | 1993-08-26 | 1994-12-13 | Smith & Nephew Richards, Inc. | Surface and near surface hardened medical implants |
US5482580A (en) * | 1994-06-13 | 1996-01-09 | Amorphous Alloys Corp. | Joining of metals using a bulk amorphous intermediate layer |
US5578359A (en) * | 1994-11-29 | 1996-11-26 | Hewlett Packard Company | Magnetic shielding garment for electro-biologic measurements |
US5803996A (en) * | 1995-01-25 | 1998-09-08 | Research Development Corporation Of Japan | Rod-shaped or tubular amorphous Zr alloy made by die casting and method for manufacturing said amorphous Zr alloy |
US5618359A (en) * | 1995-02-08 | 1997-04-08 | California Institute Of Technology | Metallic glass alloys of Zr, Ti, Cu and Ni |
US5842858A (en) * | 1995-05-11 | 1998-12-01 | Artma Biomedical, Inc. | Method of imaging a person's jaw and a model therefor |
US5895401A (en) * | 1995-08-24 | 1999-04-20 | Daum Gmbh | Controlled-artifact magnetic resonance instruments |
US5769861A (en) * | 1995-09-28 | 1998-06-23 | Brainlab Med. Computersysteme Gmbh | Method and devices for localizing an instrument |
US5711363A (en) * | 1996-02-16 | 1998-01-27 | Amorphous Technologies International | Die casting of bulk-solidifying amorphous alloys |
US5735975A (en) * | 1996-02-21 | 1998-04-07 | California Institute Of Technology | Quinary metallic glass alloys |
US5980652A (en) * | 1996-05-21 | 1999-11-09 | Research Developement Corporation Of Japan | Rod-shaped or tubular amorphous Zr alloy made by die casting and method for manufacturing said amorphous Zr alloy |
US5823778A (en) * | 1996-06-14 | 1998-10-20 | The United States Of America As Represented By The Secretary Of The Air Force | Imaging method for fabricating dental devices |
US5772803A (en) * | 1996-08-26 | 1998-06-30 | Amorphous Technologies International | Torsionally reacting spring made of a bulk-solidifying amorphous metallic alloy |
US5797443A (en) * | 1996-09-30 | 1998-08-25 | Amorphous Technologies International | Method of casting articles of a bulk-solidifying amorphous alloy |
US5976274A (en) * | 1997-01-23 | 1999-11-02 | Akihisa Inoue | Soft magnetic amorphous alloy and high hardness amorphous alloy and high hardness tool using the same |
US6077367A (en) * | 1997-02-19 | 2000-06-20 | Alps Electric Co., Ltd. | Method of production glassy alloy |
US6306228B1 (en) * | 1998-07-08 | 2001-10-23 | Japan Science And Technology Corporation | Method of producing amorphous alloy excellent in flexural strength and impact strength |
US6675033B1 (en) * | 1999-04-15 | 2004-01-06 | Johns Hopkins University School Of Medicine | Magnetic resonance imaging guidewire probe |
US6714809B2 (en) * | 2000-11-20 | 2004-03-30 | Surgi-Vision, Inc. | Connector and guidewire connectable thereto |
Cited By (37)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
US20060210880A1 (en) * | 1993-11-19 | 2006-09-21 | Medtronic, Inc. | Current collector |
US8527046B2 (en) | 2000-04-20 | 2013-09-03 | Medtronic, Inc. | MRI-compatible implantable device |
EP1632584A1 (en) * | 2004-09-06 | 2006-03-08 | Eidgenössische Technische Hochschule Zürich | Amorphous alloys on the base of Zr and their use |
WO2006026882A1 (en) * | 2004-09-06 | 2006-03-16 | Eidgenössische Technische Hochschule Zürich | Amorphous alloys on the base of zr and their use |
US20080190521A1 (en) * | 2004-09-06 | 2008-08-14 | Eidgenossische Technische Hochschule Zurich | Amorphous Alloys on the Base of Zr and their Use |
US8886317B2 (en) | 2004-12-17 | 2014-11-11 | Cardiac Pacemakers, Inc. | MRI operation modes for implantable medical devices |
US8543207B2 (en) | 2004-12-17 | 2013-09-24 | Cardiac Pacemakers, Inc. | MRI operation modes for implantable medical devices |
US8014867B2 (en) | 2004-12-17 | 2011-09-06 | Cardiac Pacemakers, Inc. | MRI operation modes for implantable medical devices |
US7684860B2 (en) | 2006-03-24 | 2010-03-23 | Medtronic, Inc. | Components for reducing image distortion |
US20100145183A1 (en) * | 2006-03-24 | 2010-06-10 | Medtronic, Inc. | Implantable medical device |
US7890165B2 (en) | 2006-03-24 | 2011-02-15 | Medtronic, Inc. | Implantable medical device with reduced MRI image distortion |
US7927737B2 (en) | 2006-03-24 | 2011-04-19 | Medtronic, Inc. | Implantable medical device and lithium battery |
US9393408B2 (en) | 2006-03-24 | 2016-07-19 | Medtronic, Inc. | Implantable medical device |
US8923969B2 (en) | 2006-03-24 | 2014-12-30 | Medtronic, Inc. | Implantable medical device |
US8131368B2 (en) | 2006-03-24 | 2012-03-06 | Medtronic, Inc. | Implantable medical device with material for reducing MRI image distortion |
US20070248881A1 (en) * | 2006-03-24 | 2007-10-25 | Medtronic, Inc. | Implantable medical device and lithium battery |
US8548591B2 (en) | 2006-03-24 | 2013-10-01 | Medtronic Inc. | Implantable medical device |
US8057530B2 (en) | 2006-06-30 | 2011-11-15 | Tyco Healthcare Group Lp | Medical devices with amorphous metals, and methods therefor |
US20080125848A1 (en) * | 2006-06-30 | 2008-05-29 | Kusleika Richard S | Medical devices with amorphous metals, and methods therefor |
US20080103543A1 (en) * | 2006-10-31 | 2008-05-01 | Medtronic, Inc. | Implantable medical device with titanium alloy housing |
US20090139612A1 (en) * | 2007-11-30 | 2009-06-04 | Kun Lu | Zr-based amorphous alloy and a preparing method thereof |
US8554335B2 (en) | 2007-12-06 | 2013-10-08 | Cardiac Pacemakers, Inc. | Method and apparatus for disconnecting the tip electrode during MRI |
US8897875B2 (en) | 2007-12-06 | 2014-11-25 | Cardiac Pacemakers, Inc. | Selectively connecting the tip electrode during therapy for MRI shielding |
US8032228B2 (en) | 2007-12-06 | 2011-10-04 | Cardiac Pacemakers, Inc. | Method and apparatus for disconnecting the tip electrode during MRI |
US8086321B2 (en) | 2007-12-06 | 2011-12-27 | Cardiac Pacemakers, Inc. | Selectively connecting the tip electrode during therapy for MRI shielding |
US8311637B2 (en) | 2008-02-11 | 2012-11-13 | Cardiac Pacemakers, Inc. | Magnetic core flux canceling of ferrites in MRI |
US8160717B2 (en) | 2008-02-19 | 2012-04-17 | Cardiac Pacemakers, Inc. | Model reference identification and cancellation of magnetically-induced voltages in a gradient magnetic field |
US20090288741A1 (en) * | 2008-03-21 | 2009-11-26 | Faliang Zhang | Amorphous Alloy and A Preparation Method Thereof |
US8571661B2 (en) | 2008-10-02 | 2013-10-29 | Cardiac Pacemakers, Inc. | Implantable medical device responsive to MRI induced capture threshold changes |
US9561378B2 (en) | 2008-10-02 | 2017-02-07 | Cardiac Pacemakers, Inc. | Implantable medical device responsive to MRI induced capture threshold changes |
US8639331B2 (en) | 2009-02-19 | 2014-01-28 | Cardiac Pacemakers, Inc. | Systems and methods for providing arrhythmia therapy in MRI environments |
US8977356B2 (en) | 2009-02-19 | 2015-03-10 | Cardiac Pacemakers, Inc. | Systems and methods for providing arrhythmia therapy in MRI environments |
US9381371B2 (en) | 2009-12-08 | 2016-07-05 | Cardiac Pacemakers, Inc. | Implantable medical device with automatic tachycardia detection and control in MRI environments |
US8565874B2 (en) | 2009-12-08 | 2013-10-22 | Cardiac Pacemakers, Inc. | Implantable medical device with automatic tachycardia detection and control in MRI environments |
US9357996B2 (en) * | 2010-09-08 | 2016-06-07 | DePuy Synthes Products, Inc. | Fixation device with magnesium core |
US10694950B2 (en) | 2016-10-13 | 2020-06-30 | International Business Machines Corporation | Probe structure for physiological measurements using surface enhanced Raman spectroscopy |
US10694951B2 (en) | 2016-10-13 | 2020-06-30 | International Business Machines Corporation | Probe structure for physiological measurements using surface enhanced Raman spectroscopy |
Also Published As
Publication number | Publication date |
---|---|
US20020162605A1 (en) | 2002-11-07 |
Similar Documents
Publication | Publication Date | Title |
---|---|---|
US20030111142A1 (en) | Bulk metallic glass medical instruments, implants, and methods of using same | |
Mehjabeen et al. | Zirconium alloys for orthopaedic and dental applications | |
Li et al. | Recent advances in bulk metallic glasses for biomedical applications | |
Hayes et al. | The use of titanium and stainless steel in fracture fixation | |
Davidson et al. | New surface-hardened, low-modulus, corrosion-resistant Ti-13Nb-13Zr alloy for total hip arthroplasty | |
US6409852B1 (en) | Biocompatible low modulus titanium alloy for medical implant | |
Mears | Metals in medicine and surgery | |
Andersen | 1.1 Metals for use in medicine | |
Kuo et al. | The use of nickel-titanium alloy in orthopedic surgery in China | |
Suzuki et al. | Appraising the potential of Zr-based biomedical alloys to reduce magnetic resonance imaging artifacts | |
JP2004089580A (en) | Biomaterial member | |
Rodríguez-González | Biomaterials in orthopaedic surgery | |
Manaka et al. | Galvanic Corrosion among Ti–6Al–4V ELI Alloy, Co–Cr–Mo Alloy, 316L-Type Stainless Steel, and Zr–1Mo Alloy for Orthopedic Implants | |
Horton et al. | Biomedical potential of a zirconium-based bulk metallic glass | |
WO2018047611A1 (en) | Biocompatible alloy and medical product | |
Bolmaro et al. | Investigation of cast and annealed Ti25Nb10Zr alloy as material for orthopedic devices | |
Cigada et al. | In vivo behavior of a high performance duplex stainless steel | |
Sridhar et al. | Biomaterials corrosion | |
Ayers et al. | Spine rod straightening as a possible cause for revision | |
Dimitriadis et al. | Microstructure and Mechanical Properties of Co-Cr Alloy Fabricated by Selective Laser Melting Technology for Removable Partial Denture Frameworks | |
JP6041045B2 (en) | Zirconium alloy for living body and bone fixing device using the same | |
JP2012021198A (en) | Amorphous alloy and biomaterial | |
Mehjabeen | Design and development of low elastic modulus Ti-Nb-Zr alloys for biomedical applications | |
Cho et al. | Zirconia-polyurethane aneurysm clip | |
Sumita et al. | Durability of metallic implant materials |
Legal Events
Date | Code | Title | Description |
---|---|---|---|
STCB | Information on status: application discontinuation |
Free format text: EXPRESSLY ABANDONED -- DURING EXAMINATION |