Bot et al., 2015 - Google Patents
Analysis of stereotactic accuracy in patients undergoing deep brain stimulation using Nexframe and the Leksell frameBot et al., 2015
- Document ID
- 6107070676050048826
- Author
- Bot M
- Van Den Munckhof P
- Bakay R
- Sierens D
- Stebbins G
- Verhagen Metman L
- Publication year
- Publication venue
- Stereotactic and functional neurosurgery
External Links
Snippet
Objective: To determine and compare the accuracy of Nexframe and the Leksell stereotactic frame in deep brain stimulation (DBS) procedures. Background: The 'frameless' Nexframe uses bone fiducials rather than a head-mounted frame, which offers potential benefits for …
- 210000004556 Brain 0 title abstract description 17
Classifications
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B5/00—Detecting, measuring or recording for diagnostic purposes; Identification of persons
- A61B5/05—Detecting, measuring or recording for diagnosis by means of electric currents or magnetic fields; Measuring using microwaves or radiowaves
- A61B5/055—Detecting, measuring or recording for diagnosis by means of electric currents or magnetic fields; Measuring using microwaves or radiowaves involving electronic or nuclear magnetic resonance, e.g. magnetic resonance imaging
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B17/00—Surgical instruments, devices or methods, e.g. tourniquets
- A61B17/34—Trocars; Puncturing needles
- A61B17/3403—Needle locating or guiding means
- A61B2017/3405—Needle locating or guiding means using mechanical guide means
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B5/00—Detecting, measuring or recording for diagnostic purposes; Identification of persons
- A61B5/48—Other medical applications
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
-
- G—PHYSICS
- G06—COMPUTING; CALCULATING; COUNTING
- G06F—ELECTRICAL DIGITAL DATA PROCESSING
- G06F19/00—Digital computing or data processing equipment or methods, specially adapted for specific applications
- G06F19/30—Medical informatics, i.e. computer-based analysis or dissemination of patient or disease data
- G06F19/32—Medical data management, e.g. systems or protocols for archival or communication of medical images, computerised patient records or computerised general medical references
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Bot et al. | Analysis of stereotactic accuracy in patients undergoing deep brain stimulation using Nexframe and the Leksell frame | |
Germano et al. | Clinical use of the optical digitizer for intracranial neuronavigation | |
Vakharia et al. | Accuracy of intracranial electrode placement for stereoelectroencephalography: A systematic review and meta‐analysis | |
Zonenshayn et al. | Comparison of anatomic and neurophysiological methods for subthalamic nucleus targeting | |
Maciunas et al. | The application accuracy of stereotactic frames | |
Ganslandt et al. | Magnetic source imaging combined with image-guided frameless stereotaxy: a new method in surgery around the motor strip | |
Giller et al. | Stereotactic pallidotomy and thalamotomy using individual variations of anatomic landmarks for localization | |
Liu et al. | Frameless ROSA® robot-assisted lead implantation for deep brain stimulation: technique and accuracy | |
Hamid et al. | Targeting the subthalamic nucleus for deep brain stimulation: technical approach and fusion of pre-and postoperative MR images to define accuracy of lead placement | |
Shahlaie et al. | Intraoperative computed tomography for deep brain stimulation surgery: technique and accuracy assessment | |
Larson et al. | An optimized system for interventional magnetic resonance imaging-guided stereotactic surgery: preliminary evaluation of targeting accuracy | |
Thani et al. | Accuracy of postoperative computed tomography and magnetic resonance image fusion for assessing deep brain stimulation electrodes | |
Holloway et al. | A quantitative assessment of the accuracy and reliability of O-arm images for deep brain stimulation surgery | |
Randazzo et al. | Three-dimensional localization of cortical electrodes in deep brain stimulation surgery from intraoperative fluoroscopy | |
Pfisterer et al. | Fiducial versus nonfiducial neuronavigation registration assessment and considerations of accuracy | |
Sharma et al. | Accuracy and precision of targeting using frameless stereotactic system in deep brain stimulator implantation surgery | |
Bednarz et al. | Evaluation of the spatial accuracy of magnetic resonance imaging-based stereotactic target localization for gamma knife radiosurgery of functional disorders | |
Holtzheimer III et al. | Magnetic resonance imaging versus computed tomography for target localization in functional stereotactic neurosurgery | |
Shamir et al. | Target and trajectory clinical application accuracy in neuronavigation | |
Balanescu et al. | A personalized stereotactic fixture for implantation of depth electrodes in stereoelectroencephalography | |
Lee et al. | Interventional MRI–guided deep brain stimulation lead implantation | |
Carlson et al. | Stereotactic accuracy and surgical utility of the O-arm in deep brain stimulation surgery | |
Pezeshkian et al. | Accuracy of frame-based stereotactic magnetic resonance imaging vs frame-based stereotactic head computed tomography fused with recent magnetic resonance imaging for postimplantation deep brain stimulator lead localization | |
Ringel et al. | VarioGuide: a new frameless image-guided stereotactic system—accuracy study and clinical assessment | |
Budke et al. | Implantation of depth electrodes in children using VarioGuide® frameless navigation system |