The first clinical treatment with kilovoltage intrafraction monitoring (KIM): a real‐time image guidance methodKeall et al., 2015
View PDF- Document ID
- 4735223087302007456
- Author
- Keall P
- Aun Ng J
- O'Brien R
- Colvill E
- Huang C
- Rugaard Poulsen P
- Fledelius W
- Juneja P
- Simpson E
- Bell L
- Alfieri F
- Eade T
- Kneebone A
- Booth J
- Publication year
- Publication venue
- Medical physics
External Links
Snippet
Purpose: Kilovoltage intrafraction monitoring (KIM) is a real‐time image guidance method that uses widely available radiotherapy technology, ie, a gantry‐mounted x‐ray imager. The authors report on the geometric and dosimetric results of the first patient treatment using KIM …
- 201000011510 cancer 0 abstract description 30
Classifications
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1064—Monitoring, verifying, controlling systems and methods for adjusting radiation treatment in response to monitoring
- A61N5/1065—Beam adjustment
- A61N5/1067—Beam adjustment in real time, i.e. during treatment
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1049—Monitoring, verifying, controlling systems and methods for verifying the position of the patient with respect to the radiation beam
- A61N2005/1061—Monitoring, verifying, controlling systems and methods for verifying the position of the patient with respect to the radiation beam using an x-ray imaging system having a separate imaging source
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1075—Monitoring, verifying, controlling systems and methods for testing, calibrating, or quality assurance of the radiation treatment apparatus
- A61N2005/1076—Monitoring, verifying, controlling systems and methods for testing, calibrating, or quality assurance of the radiation treatment apparatus using a dummy object placed in the radiation field, e.g. phantom
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/103—Treatment planning systems
- A61N5/1037—Treatment planning systems taking into account the movement of the target, e.g. 4D-image based planning
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N2005/1085—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy characterised by the type of particles applied to the patient
- A61N2005/1087—Ions; Protons
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N2005/1092—Details
- A61N2005/1096—Elements inserted into the radiation path placed on the patient, e.g. bags, bolus, compensators
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/103—Treatment planning systems
- A61N5/1031—Treatment planning systems using a specific method of dose optimization
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1077—Beam delivery systems
- A61N5/1081—Rotating beam systems with a specific mechanical construction, e.g. gantries
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1042—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy with spatial modulation of the radiation beam within the treatment head
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/04—Positioning of patients; Tiltable beds or the like
- A61B6/0407—Tables or beds
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/48—Diagnostic techniques
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/54—Control of devices for radiation diagnosis
-
- G—PHYSICS
- G06—COMPUTING; CALCULATING; COUNTING
- G06F—ELECTRICAL DIGITAL DATA PROCESSING
- G06F19/00—Digital computing or data processing equipment or methods, specially adapted for specific applications
- G06F19/30—Medical informatics, i.e. computer-based analysis or dissemination of patient or disease data
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Keall et al. | The first clinical treatment with kilovoltage intrafraction monitoring (KIM): a real‐time image guidance method | |
Schnarr et al. | Feasibility of real‐time motion management with helical tomotherapy | |
Ge et al. | Toward the development of intrafraction tumor deformation tracking using a dynamic multi‐leaf collimator | |
Potters et al. | American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for image-guided radiation therapy (IGRT) | |
Mao et al. | Fast internal marker tracking algorithm for onboard MV and kV imaging systems | |
Li et al. | Evaluation of the geometric accuracy of surrogate‐based gated VMAT using intrafraction kilovoltage x‐ray images | |
Nguyen et al. | The first clinical implementation of a real-time six degree of freedom target tracking system during radiation therapy based on Kilovoltage Intrafraction Monitoring (KIM) | |
Mukumoto et al. | Accuracy verification of infrared marker‐based dynamic tumor‐tracking irradiation using the gimbaled x‐ray head of the Vero4DRT (MHI‐TM2000) a | |
Shiinoki et al. | Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study | |
Zeng et al. | Intrafraction tumor motion during deep inspiration breath hold pancreatic cancer treatment | |
Jin et al. | Time delay measurement for linac based treatment delivery in synchronized respiratory gating radiotherapy | |
Cai et al. | 3D delivered dose assessment using a 4DCT‐based motion model | |
Ng et al. | Quality assurance for the clinical implementation of kilovoltage intrafraction monitoring for prostate cancer VMAT | |
Legge et al. | Real‐time intrafraction prostate motion during linac based stereotactic radiotherapy with rectal displacement | |
Bryant et al. | Registration of clinical volumes to beams‐eye‐view images for real‐time tracking | |
Wang et al. | Target repositional accuracy and PTV margin verification using three‐dimensional cone‐beam computed tomography (CBCT) in stereotactic body radiotherapy (SBRT) of lung cancers | |
Teske et al. | Real‐time markerless lung tumor tracking in fluoroscopic video: Handling overlapping of projected structures | |
Mueller et al. | The markerless lung target tracking AAPM Grand Challenge (MATCH) results | |
Hewson et al. | The accuracy and precision of the KIM motion monitoring system used in the multi‐institutional TROG 15.01 Stereotactic Prostate Ablative Radiotherapy with KIM (SPARK) trial | |
Nakamura et al. | Evaluation of dynamic tumor-tracking intensity-modulated radiotherapy for locally advanced pancreatic cancer | |
Inoue et al. | Development of system using beam's eye view images to measure respiratory motion tracking errors in image‐guided robotic radiosurgery system | |
Koike et al. | Dosimetric impact of intra-fraction prostate motion under a tumour-tracking system in hypofractionated robotic radiosurgery | |
Sengupta et al. | The dosimetric error due to uncorrected tumor rotation during real‐time adaptive prostate stereotactic body radiation therapy | |
Chung et al. | Reconstruction of implanted marker trajectories from cone‐beam CT projection images using interdimensional correlation modeling | |
Floriano et al. | Retrospective evaluation of CTV to PTV margins using CyberKnife in patients with thoracic tumors |