Abstract
Objective
The aim of this study was to evaluate images quality and radiation doses of Cone Beam Computed Tomography (CBCT) for dental and maxillofacial imaging testing five different acquisition protocols.
Methods
Dose measurements of different acquisition protocols were calculated for Pax Zenith three-dimensional (3D) Cone Beam (Vatech, Korea) and for conventional orthopantomography (OPT) and cephalometric skull imaging Ortophos (Sirona Dental Systems, Bernsheim, Germany). The absorbed organ doses were measured using an anthropomorphic phantom loaded with thermoluminescent dosimeters at 58 sites related to sensitive organs. Five different CBCT protocols were evaluated for image quality and radiation doses. They differed in FOV, image resolution, kVp, mA, acquisition time in seconds and radiation dose. Measurements were then carried out with the orthopantomograph. Equivalent and effective doses were calculated.
Results
The reference protocol with large FOV, high resolution quality images, 95 kVp, 5 mA and acquisition time of 24 s resulted in a DAP value of 1556 mGy cm2 instead the protocol with reduced kVp from 95 to 80 kVp translated into a value of DAP inferior to 35% (from 1556 to 1013 mGy cm2). Going from a high resolution to a normal resolution, there was a reduction of the acquisition time to 15 s which allowed further dose reduction of approximately 40% (628 mGy cm2); this protocol resulted in a value of effective dose of 35 microSievert (μSv). Moreover, the effect of changing FOV has been evaluated, considering two scans with a reduced FOV (160 × 140 and 120 × 90 mm, respectively).
Conclusions
CBCT low-dose protocol with large FOV, normal resolution quality images, 80 kVp, 5 mA and acquisition time of 15 s resulted in a value of effective dose of 35 microSievert (μSv). This protocol allows the study of maxillofacial region with high quality of images and a very low radiation dose and, therefore, could be proposed in selected case where a complete assessment of dental and maxillofacial region is useful for treatment planning.
Similar content being viewed by others
References
Cohnen M, Kemper J, Mobes O, Pawelzik J, Modder U (2002) Radiation dose in dental radiology. Eur Radiol 12:634–637
Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K (1999) Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol 28:245–248 (PubMed Citation)
Mah JK, Danforth RA, Bumann A, Hatcher D (2003) Radiation absorbed in maxillofacial imaging with a new dental computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 96:508–513
Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: newtom cone beam CT and orthophos plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234
Yammamoto K, Ueno K, Seo K, Shinohara D (2003) Development of dento-maxillofacial cone beam X-ray computed tomography system. Orthod Craniofac Res 6(Suppl 1):160–162
Nakajima A, Sameshima GT, Arai Y, Homme Y, Shimizu N, Dougherty H (2005) Two- and three-dimensional orthodontic imaging using limited cone beam-computed tomography. Angle Orthod 75:895–903
Honda K, Larheim TA, Matsumoto K, Iwai K (2006) Osseous abnormalities of the mandibular condyle: diagnostic reliability of cone beam computed tomography compared with helical computed tomography based on an autopsy material. Dentomaxillofac Radiol 35:152–157
Tsilakis K, Syriopoulos K, Stamatakis HC (2004) Radiographic examination of the temporomandibular joint using cone beam computed tomography. Dentomaxillofac Radiol 33:196–201
Maki K, Inou N, Takanishi A, Miller AJ (2003) Computer-assisted simulations in orthodontic diagnosis and the application of a new cone beam X-ray computed tomography. Orthod Craniofac Res 6(Suppl. 1):95–101
Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:106–114
Rampado O, Bianchi SD, PeruzzoCornetto A, Rossetti V, Ropolo R (2014) Radiochromic films for dental CT dosimetry: a feasibility study. Phys Med 30(1):18–24. doi:10.1016/j.ejmp.2012.06.002
Tapiovaara M, Lakkisto M, Servomaa A (1997) PCXMC-A PC-based Monte Carlo program for calculating patient doses in medical X-ray examinations. Finnish Centre for Radiation and Nuclear Safety (STUK), Helsinki
Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Gröndahl K (2008) Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol 37(2):72–79
ICRP Pubblication 103 (2008) The 2007 recommendations of the International Commission on Radiological Protection. Ann ICRP 37:2–4
European Commission (2012) Radiaton protection no. 172: cone beam CT for dental and maxillofacial Radiology. Evidence based guidelines. http://www.sedentexct.eu/files/radiation_protection_172.pdf
Bianchi SD, Rampado O, Luberto L, Genovesio AF, Bianchi CC, Ropolo R (2005) Image quality analysis and low dose dental CT. Int Congr Ser 1281:1177–1181
Kapila S, Conley RS, Harrell WE (2011) The current status of cone beam computed tomography imaging in orthodontics. Dentomaxillofac Radiol 40:24–34
International Commission on Radiation Protection (1991) 1990 Recommendations of the International Commission on Radiological Protection, ICRP Publication 60. Ann ICRP 21:1–201
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Funding
There are any source of funding or financial interest.
Conflict of interest
The authors declare no conflicts of interest.
Ethical standards
This article does not contain any studies with human participants or animals performed by any of the authors.
Rights and permissions
About this article
Cite this article
Feragalli, B., Rampado, O., Abate, C. et al. Cone beam computed tomography for dental and maxillofacial imaging: technique improvement and low-dose protocols. Radiol med 122, 581–588 (2017). https://doi.org/10.1007/s11547-017-0758-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11547-017-0758-2