[go: up one dir, main page]
More Web Proxy on the site http://driver.im/ skip to main content
research-article

A survey on Motion Artifact Correction in Magnetic Resonance Imaging for Improved Diagnostics

Published: 22 February 2024 Publication History

Abstract

Motion artifacts occur in magnetic resonance imaging (MRI) due to the motion or movement of the object being scanned. Motion artifacts can have various origins such as voluntary or involuntary patient movement, faulty components, improper software configuration, etc. Blurry MRI scans are generated due to the presence of motion artifacts. In some cases motion artifact induced MRI scans can tamper with crucial information and consequently leads to a faulty diagnosis. We attempt to provide a review of current suggested technologies (such as, deep learning, and encoding) used to remove motion artifacts from the MRI scans. The different approaches are summarized in brief, with their advantages and disadvantages. We expect the readers to be better equipped with the knowledge and tools available for magnetic resonance image artifact removal. We have also proposed many simple cases and their corresponding solutions in the discussion section of this manuscript. It was found that deep learning with CNN can reach maximum accuracy reported in motion artifact correction is about 97%. With the advancements in the field of artificial intelligence, and more and more sample data available (due to non-linear expansion of big data and cloud storage capabilities), it is expected that motion artifacts could soon be eliminated quite accurately.

References

[1]
1234.info: motion artifact correction with motionSCOUT (under development). 2019. https://www.robinmedical.com/motion_artifact.html.
[2]
Al-Masni MA, Lee S, Al-Shamiri AK, Gho SM, Choi YH, and Kim DH A knowledge interaction learning for multi-echo MRI motion artifact correction towards better enhancement of SWI Comput Biol Medicine 2023 153
[3]
Al-Masni MA, Lee S, Yi J, Kim S, Gho SM, Choi YH, and Kim DH Stacked U-Nets with self-assisted priors towards robust correction of rigid motion artifact in brain MRI NeuroImage 2022 259
[4]
Atkinson D, Hill DL, Stoyle PN, Summers PE, Clare S, Bowtell R, and Keevil SF Automatic compensation of motion artifacts in MRI Magn Reson Med 1999 41 1 163-170
[5]
Atkinson D, Hill DL, Stoyle PN, Summers PE, and Keevil SF Automatic correction of motion artifacts in magnetic resonance images using an entropy focus criterion IEEE Trans Med Imaging 1997 16 6 903-910
[6]
Bailes D, Gilderdale D, Bydder G, Collins A, and Firmin D Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging J Comput Assist Tomogr 1985 9 4 835-838
[7]
Bernstein MA, Huston J III, and Ward HA Imaging artifacts at 30 T J Magn Reson Imaging 2006 24 4 735-746
[8]
Blumenthal JD, Zijdenbos A, Molloy E, and Giedd JN Motion artifact in magnetic resonance imaging: implications for automated analysis Neuroimage 2002 16 1 89-92
[9]
Brau AC and Brittain JH Generalized self-navigated motion detection technique: preliminary investigation in abdominal imaging Magn Reson Med 2006 55 2 263-270
[10]
Castellanos NP and Makarov VA Recovering EEG brain signals: artifact suppression with wavelet enhanced independent component analysis J Neurosci Methods 2006 158 2 300-312
[11]
Chung H, Kim J, Yoon JH, Lee JM, Ye JC. Simultaneous super-resolution and motion artifact removal in diffusion-weighted MRI using unsupervised deep learning. 2021. arXiv preprint arXiv:2105.00240.
[12]
Ciric R, Rosen AF, Erus G, Cieslak M, Adebimpe A, Cook PA, Bassett DS, Davatzikos C, Wolf DH, and Satterthwaite TD Mitigating head motion artifact in functional connectivity MRI Nat Protoc 2018 13 12 2801-2826
[13]
Cui L, Song Y, Wang Y, Wang R, Wu D, Xie H, Li J, and Yang G Motion artifact reduction for magnetic resonance imaging with deep learning and k-space analysis PLoS One 2023 18 1
[14]
Dietrich TJ, Ulbrich EJ, Zanetti M, Fucentese SF, and Pfirrmann CW Propeller technique to improve image quality of MRI of the shoulder Am J Roentgenol 2011 197 6 W1093-W1100
[15]
Duffy BA, Zhang W, Tang H, Zhao L, Law M, Toga AW, Kim H. Retrospective correction of motion artifact affected structural MRI images using deep learning of simulated motion. 2018.
[16]
Duffy BA, Zhang W, Tang H, Zhao L, Law M, Toga AW, Kim H. Retrospective correction of motion artifact affected structural MRI images using deep learning of simulated motion. In: Medical imaging with deep learning. 2022.
[17]
Duffy BA, Zhao L, Sepehrband F, Min J, Wang DJ, Shi Y, Toga AW, Kim H, Initiative ADN, et al. Retrospective motion artifact correction of structural MRI images using deep learning improves the quality of cortical surface reconstructions NeuroImage 2021 230
[18]
Duong S, Phung SL, Bouzerdoum A, Taylor HB, Puckett A, and Schira MM Susceptibility artifact correction for sub-millimeter fMRI using inverse phase encoding registration and T1 weighted regularization J Neurosci Methods 2020 336
[19]
Ehman RL and Felmlee JP Adaptive technique for high-definition MR imaging of moving structures Radiology 1989 173 1 255-263
[20]
[21]
Fantini I, Yasuda C, Bento M, Rittner L, Cendes F, and Lotufo R Automatic MR image quality evaluation using a deep CNN: a reference-free method to rate motion artifacts in neuroimaging Comput Med Imaging Graph 2021 90
[22]
Goto M, Abe O, Miyati T, Yamasue H, Gomi T, Takeda T. Head motion and correction methods in resting-state functional MRI. Magn Reson Med Sci. 2015;rev–2015.
[23]
Graves MJ and Mitchell DG Body MRI artifacts in clinical practice: a physicist’s and radiologist’s perspective J Magn Reson Imaging 2013 38 2 269-287
[24]
Hedley M, Yan H, and Rosenfeld D Motion artifact correction in MRI using generalized projections IEEE Trans Medical Imaging 1991 10 1 40-46
[25]
Herrick RC, Hayman LA, Taber KH, Diaz-Marchan PJ, and Kuo MD Artifacts and pitfalls in MR imaging of the orbit: a clinical review Radiographics 1997 17 3 707-724
[26]
de Hollander G, Keuken MC, van der Zwaag W, Forstmann BU, and Trampel R Comparing functional MRI protocols for small, iron-rich basal ganglia nuclei such as the subthalamic nucleus at 7 T and 3 T Hum Brain Map 2017 38 6 3226-3248
[27]
Kaur P, Kumaran SS, Tripathi R, Khushu S, and Kaushik S Protocol error artifacts in MRI: sources and remedies revisited Radiography 2007 13 4 291-306
[28]
Kim SY, Park SH, Wu EH, Wang ZJ, Hope TA, Chang WC, and Yeh BM Transient respiratory motion artifact during arterial phase MRI with gadoxetate disodium: risk factor analyses Am J Roentgenol 2015 204 6 1220-1227
[29]
Kochunov P, Lancaster JL, Glahn DC, Purdy D, Laird AR, Gao F, and Fox P Retrospective motion correction protocol for high-resolution anatomical MRI Hum Brain Map 2006 27 12 957-962
[30]
Krupa K and Bekiesińska-Figatowska M Artifacts in magnetic resonance imaging Pol J Radiol 2015 80 93
[31]
Kubicek J, Strycek M, Cerny M, Penhaker M, Prokop O, and Vilimek D Quantitative and comparative analysis of effectivity and robustness for enhanced and optimized non-local mean filter combining pixel and patch information on mr images of musculoskeletal system Sensors 2021 21 12 4161
[32]
Küstner T, Liebgott A, Mauch L, Martirosian P, Bamberg F, Nikolaou K, Yang B, Schick F, and Gatidis S Automated reference-free detection of motion artifacts in magnetic resonance images Magn Reson Mater Phys Biol Med 2018 31 2 243-256
[33]
Kuzmina E, Razumov A, Rogov OY, Adalsteinsson E, White J, Dylov DV. Autofocusing+: noise-resilient motion correction in magnetic resonance imaging. In: International conference on medical image computing and computer-assisted intervention. Springer; 2022. pp. 365–375.
[34]
Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, and Wong-You-Cheong JJ Comparison of sagittal T2-weighted blade and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection Am J Roentgenol 2011 197 2 W307-W313
[35]
Liau J, Lee J, Schroeder ME, Sirlin CB, and Bydder M Cardiac motion in diffusion-weighted MRI of the liver: artifact and a method of correction J Magn Reson Imaging 2012 35 2 318-327
[36]
Lim A, Lo J, Wagner MW, Ertl-Wagner B, and Sussman D Motion artifact correction in fetal MRI based on a generative adversarial network method Biomed Signal Process Control 2023 81
[37]
Liu J, Kocak M, Supanich M, and Deng J Motion artifacts reduction in brain MRI by means of a deep residual network with densely connected multi-resolution blocks (DRN-DCMB) Magn Reson Imaging 2020 71 69-79
[38]
Lorch B, Vaillant G, Baumgartner C, Bai W, Rueckert D, and Maier A Automated detection of motion artefacts in MR imaging using decision forests J Med Eng 2017 2017 4501647
[39]
Lyu Q, Shan H, Xie Y, Kwan AC, Otaki Y, Kuronuma K, Li D, and Wang G Cine cardiac MRI motion artifact reduction using a recurrent neural network IEEE Trans Med Imaging 2021 40 2170-2181
[40]
Maclaren J, Herbst M, Speck O, and Zaitsev M Prospective motion correction in brain imaging: a review Magn Reson Med 2013 69 3 621-636
[41]
Mirowitz SA Motion artifact as a pitfall in diagnosis of meniscal tear on gradient reoriented MRI of the knee J Comput Assist Tomogr 1994 18 2 279-282
[42]
Mohaldin AH, et al. MRI image motion blur correction. 2018.
[43]
Morelli JN, Runge VM, Ai F, Attenberger U, Vu L, Schmeets SH, Nitz WR, and Kirsch JE An image-based approach to understanding the physics of MR artifacts Radiographics 2011 31 3 849-866
[44]
Mucelli RP, Amezich W, Longo R, Ricci C. Artifacts in MRI: a proposal for classification based upon the encoding process. In: Frontiers in European radiology. Springer; 1993. pp. 187–211.
[45]
Odille F, Cîndea N, Mandry D, Pasquier C, Vuissoz PA, and Felblinger J Generalized MRI reconstruction including elastic physiological motion and coil sensitivity encoding Magn Reson Med 2008 59 6 1401-1411
[46]
Oh G, Lee JE, Ye JC. Unsupervised MR motion artifact deep learning using outlier-rejecting bootstrap aggregation. 2020. arXiv preprint arXiv:2011.06337.
[47]
Oksuz I Brain MRI artefact detection and correction using convolutional neural networks Comput Methods Progr Biomed 2021 199
[48]
Paling MR and Brookeman JR Respiration artifacts in MR imaging: reduction by breath holding J Comput Assist Tomogr 1986 10 6 1080-1082
[49]
Pattany PM, Phillips JJ, Chiu LC, Lipcamon JD, Duerk JL, McNally JM, and Mohapatra SN Motion artifact suppression technique (MAST) for MR imaging J Comput Assist Tomogr 1987 11 3 369-377
[50]
Peh WC and Chan JH Artifacts in musculoskeletal magnetic resonance imaging: identification and correction Skelet Radiol 2001 30 4 179-191
[51]
Pipe JG Motion correction with propeller MRI: application to head motion and free-breathing cardiac imaging Magn Reson Med 1999 42 5 963-969
[52]
Power JD, Barnes KA, Snyder AZ, Schlaggar BL, and Petersen SE Steps toward optimizing motion artifact removal in functional connectivity MRI; a reply to carp Neuroimage 2013 76 439-41
[53]
Rohlfing T, Maurer CR Jr, O’Dell WG, and Zhong J Modeling liver motion and deformation during the respiratory cycle using intensity-based nonrigid registration of gated MR images Med Phys 2004 31 3 427-432
[54]
Rotman M, Brada R, Beniaminy I, Ahn S, Hardy CJ, Wolf L. Correcting motion artifacts in MRI scans using a deep neural network with automatic motion timing detection. In: Medical imaging 2021: physics of medical imaging, vol. 11595. International Society for Optics and Photonics; 2021. p. 1159514.
[55]
Runge V, Clanton J, Partain C, and James A Jr Respiratory gating in magnetic resonance imaging at 0.5 Tesla Radiology 1984 151 2 521-523
[56]
Safari M, Fatemi A, Archambault L. MAUDGAN: motion artifact unsupervised disentanglement generative adversarial network of multicenter MRI data with different brain tumors. medRxiv. 2023;2023–03.
[57]
Sanchez-Panchuelo RM, Francis S, Bowtell R, and Schluppeck D Mapping human somatosensory cortex in individual subjects with 7 T functional MRI J Neurophysiol 2010 103 5 2544-2556
[58]
Sánchez-Panchuelo RM, Francis ST, Schluppeck D, and Bowtell RW Correspondence of human visual areas identified using functional and anatomical MRI in vivo at 7 T J Magn Reson Imaging 2012 35 2 287-299
[59]
Satterthwaite TD, Elliott MA, Gerraty RT, Ruparel K, Loughead J, Calkins ME, Eickhoff SB, Hakonarson H, Gur RC, Gur RE, et al. An improved framework for confound regression and filtering for control of motion artifact in the preprocessing of resting-state functional connectivity data Neuroimage 2013 64 240-256
[60]
Scarff CJ, Reynolds A, Goodyear BG, Ponton CW, Dort JC, and Eggermont JJ Simultaneous 3-T fMRI and high-density recording of human auditory evoked potentials Neuroimage 2004 23 3 1129-1142
[61]
Sciarra A, Chatterjee S, Dünnwald M, Placidi G, Nürnberger A, Speck O, Oeltze-Jafra S. Reference-less SSIM regression for detection and quantification of motion artefacts in brain MRIS. In: Medical imaging with deep learning. 2022.
[62]
Shaw R, Sudre C, Ourselin S, Cardoso MJ. MRI k-space motion artefact augmentation: model robustness and task-specific uncertainty. 2018.
[63]
Singh KR, Sharma A, and Singh GK Attention-guided residual W-Net for supervised cardiac magnetic resonance imaging segmentation Biomed Signal Process Control 2023 86
[64]
Smith TB MRI artifacts and correction strategies Imaging Med 2010 2 4 445
[65]
Softways, M.I.G.: MRI artifacts—flow artifact. 2018. https://www.mr-tip.com/serv1.php?type=art &sub=Flow+Artifact.
[66]
Sommer K, Saalbach A, Brosch T, Hall C, Cross N, and Andre J Correction of motion artifacts using a multiscale fully convolutional neural network Am J Neuroradiol 2020 41 3 416-423
[67]
Spieker V, Eichhorn H, Hammernik K, Rueckert D, Preibisch C, Karampinos DC, Schnabel JA: Deep learning for retrospective motion correction in MRI: a comprehensive review. 2023. arXiv preprint arXiv:2305.06739.
[68]
Stadler A, Schima W, Ba-Ssalamah A, Kettenbach J, and Eisenhuber E Artifacts in body MR imaging: their appearance and how to eliminate them Eur Radiol 2007 17 5 1242-1255
[69]
Tamada D, Kromrey ML, Ichikawa S, Onishi H, and Motosugi U Motion artifact reduction using a convolutional neural network for dynamic contrast enhanced MR imaging of the liver Magn Reson Med Sci 2020 19 1 64
[70]
Tan J, Mougenot C, Pichardo S, Drake JM, and Waspe AC Motion compensation using principal component analysis and projection onto dipole fields for abdominal magnetic resonance thermometry Magn Reson Med 2019 81 1 195-207
[71]
Teitelbaum G, Bradley W Jr, and Klein B MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils Radiology 1988 166 3 657-664
[72]
[73]
Tripathi V, Tibdewal M, Mishra R. Comparison of different denoising networks on motion artifacted MRI scans. In: Machine intelligence techniques for data analysis and signal processing: proceedings of the 4th international conference MISP 2022, vol. 1. Springer; 2023. pp. 301–314.
[74]
Tripathi V, Tibdewal MN, and Mishra R Modified restricted Boltzmann machine (MRBM) for denoising of motion artifact-induced MRI scans Res Biomed Eng 2023 39 1 73-83
[75]
Vakli P, Weiss B, Szalma J, Barsi P, Gyuricza I, Kemenczky P, Somogyi E, Nárai Á, Gál V, Hermann P, et al. Automatic brain MRI motion artifact detection based on end-to-end deep learning is similarly effective as traditional machine learning trained on image quality metrics Med Image Anal 2023 88
[76]
Vargas M, Delavelle J, Kohler R, Becker C, and Lovblad K Brain and spine MRI artifacts at 3 Tesla J Neuroradiol 2009 36 2 74-81
[77]
White M, Hawkes D, Melbourne A, Collins D, Coolens C, Hawkins M, Leach M, and Atkinson D Motion artifact correction in free-breathing abdominal MRI using overlapping partial samples to recover image deformations Magn Reson Med 2009 62 2 440-449
[78]
Wood ML and Henkelman RM Suppression of respiratory motion artifacts in magnetic resonance imaging Med Phys 1986 13 6 794-805
[79]
Yang RK, Roth CG, Ward RJ, deJesus JO, and Mitchell DG Optimizing abdominal MR imaging: approaches to common problems Radiographics 2010 30 1 185-199
[80]
Zanin M, Aitya NA, Basilio J, Baumbach J, Benis A, Behera CK, Bucholc M, Castiglione F, Chouvarda I, Comte B, et al. An early stage researcher’s primer on systems medicine terminology Netw Syst Med 2021 4 1 2-50
[81]
Zhou SK, Le HN, Luu K, Nguyen HV, and Ayache N Deep reinforcement learning in medical imaging: a literature review Medi Image Anal 2021 73
[82]
Zhuo J and Gullapalli RP MR artifacts, safety, and quality control Radiographics 2006 26 1 275-297

Index Terms

  1. A survey on Motion Artifact Correction in Magnetic Resonance Imaging for Improved Diagnostics
        Index terms have been assigned to the content through auto-classification.

        Recommendations

        Comments

        Please enable JavaScript to view thecomments powered by Disqus.

        Information & Contributors

        Information

        Published In

        cover image SN Computer Science
        SN Computer Science  Volume 5, Issue 3
        Mar 2024
        750 pages

        Publisher

        Springer-Verlag

        Berlin, Heidelberg

        Publication History

        Published: 22 February 2024
        Accepted: 28 December 2023
        Received: 13 July 2022

        Author Tags

        1. Misleading diagnosis
        2. Motion artifact
        3. Breathing motion
        4. Patient movement

        Qualifiers

        • Research-article

        Contributors

        Other Metrics

        Bibliometrics & Citations

        Bibliometrics

        Article Metrics

        • 0
          Total Citations
        • 0
          Total Downloads
        • Downloads (Last 12 months)0
        • Downloads (Last 6 weeks)0
        Reflects downloads up to 21 Dec 2024

        Other Metrics

        Citations

        View Options

        View options

        Media

        Figures

        Other

        Tables

        Share

        Share

        Share this Publication link

        Share on social media