Comparison of the Primary Stability of Porous Tantalum and Titanium Acetabular Revision Constructs
<p>Photograph of titanium Gription cup and augment (<b>left</b>) and tantalum Trabecular Metal augment and cup (<b>right</b>) after implant explantation, demontrating the differences in their hole positions and augment geometry.</p> "> Figure 2
<p>Schematic graph displaying the load applied for each sample over the 3000 test cycles.</p> "> Figure 3
<p>Graph displaying the average relative motion (µm) at the tantalum and titanium augment and cup interfaces at the three tested load levels (30%, 50% and 100% load).</p> "> Figure 4
<p>Graph shows the average relative motion (µm) between the tantalum and titanium augment and adjacent composite bone at the three tested load levels (30%, 50% and 100% load).</p> "> Figure 5
<p>Graph showing the average relative motion (µm) between the tantalum and titanium cup and composite bone at the three tested load levels (30%, 50% and 100% load).</p> ">
Abstract
:1. Introduction
2. Materials and Methods
3. Statistical Analysis
4. Results
5. Discussion
6. Conclusions
Author Contributions
Funding
Acknowledgments
Conflicts of Interest
References
- Kurtz, S.; Ong, K.; Lau, E.; Mowat, F.; Halpern, M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J. Bone Jt. Surg. Am. 2007, 89, 780–785. [Google Scholar] [CrossRef]
- Mahmoud, A.N.; Sundberg, M.; Flivik, G. Comparable Results With Porous Metal Augments in Combination With Either Cemented or Uncemented Cups in Revision Hip Arthroplasty: An Analysis of One Hundred Forty-Seven Revisions at a Mean of Five Years. J. Arthroplast. 2017, 32, 1612–1617. [Google Scholar] [CrossRef] [PubMed]
- Pollock, F.H.; Whiteside, L.A. The fate of massive allografts in total hip acetabular revision surgery. J. Arthroplast. 1992, 7, 271–276. [Google Scholar] [CrossRef]
- Beckmann, N.A.; Weiss, S.; Klotz, M.C.; Gondan, M.; Jaeger, S.; Bitsch, R.G. Loosening after acetabular revision: Comparison of trabecular metal and reinforcement rings. A systematic review. J. Arthroplast. 2014, 29, 229–235. [Google Scholar] [CrossRef] [PubMed]
- Jones, L.C.; Hungerford, D.S. Cement disease. Clin. Orthop. Relat. Res. 1987, 225, 192–206. [Google Scholar] [CrossRef]
- Hosny, H.A.H.; El-Bakoury, A.; Srinivasan, S.C.M.; Yarlagadda, R.; Keenan, J. Tritanium Acetabular Cup in Revision Hip Replacement: A Six to Ten Years of Follow-Up Study. J. Arthroplast. 2018, 33, 2566–2570. [Google Scholar] [CrossRef]
- Meneghini, R.M.; Meyer, C.; Buckley, C.A.; Hanssen, A.D.; Lewallen, D.G. Mechanical stability of novel highly porous metal acetabular components in revision total hip arthroplasty. J. Arthroplast. 2010, 25, 337–341. [Google Scholar] [CrossRef]
- Kärrholm, J.; Mohaddes, M.; Odin, D.; Vinblad, J.; Rogmark, C.; Rolfson, O. Swedish Hip Arthroplasty Register Annual Report 2017; Svenska Höftprotesregistret Registercentrum Västra Götaland SE-413 45: Göteborg, Sweden, 2018. [Google Scholar]
- Rossmann, M.; Ansorge, C.; Lausmann, C.; Suero, E.M.; Gehrke, T.; Citak, M. An alternative treatment option for Paprosky Type IIIb acetabular defect using multiple tantalum wedges—A case report. J. Clin. Orthop. Trauma 2020, 11, 70–72. [Google Scholar] [CrossRef]
- Houdek, M.T.; Abdel, M.P.; Perry, K.I.; Salduz, A.; Rose, P.S.; Sim, F.H.; Lewallen, D.G. Outcome of Patients Treated With Porous Tantalum Acetabular Implants for Neoplastic Periacetabular Lesions. J. Am. Acad. Orthop. Surg. 2020, 28, 256–262. [Google Scholar] [CrossRef]
- Pilliar, R.M.; Lee, J.M.; Maniatopoulos, C. Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin. Orthop. Relat. Res. 1986, 208, 108–113. [Google Scholar] [CrossRef]
- Aspenberg, P.; Goodman, S.; Toksvig-Larsen, S.; Ryd, L.; Albrektsson, T. Intermittent micromotion inhibits bone ingrowth. Titanium implants in rabbits. Acta Orthop. Scand. 1992, 63, 141–145. [Google Scholar] [CrossRef] [PubMed]
- Beckmann, N.A.; Bitsch, R.G.; Gondan, M.; Schonhoff, M.; Jaeger, S. Comparison of the stability of three fixation techniques between porous metal acetabular components and augments. Bone Jt. Res. 2018, 7, 282–288. [Google Scholar] [CrossRef] [PubMed]
- Paprosky, W.G.; Perona, P.G.; Lawrence, J.M. Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J. Arthroplast. 1994, 9, 33–44. [Google Scholar] [CrossRef]
- Bergmann, G.; Graichen, F.; Rohlmann, A.; Bender, A.; Heinlein, B.; Duda, G.N.; Heller, M.O.; Morlock, M.M. Realistic loads for testing hip implants. Biomed. Mater. Eng. 2010, 20, 65–75. [Google Scholar] [CrossRef] [PubMed]
- Bergmann, G.; Deuretzbacher, G.; Heller, M.; Graichen, F.; Rohlmann, A.; Strauss, J.; Duda, G.N. Hip contact forces and gait patterns from routine activities. J. Biomech. 2001, 34, 859–871. [Google Scholar] [CrossRef]
- Bergmann, G.; Bender, A.; Dymke, J.; Duda, G.; Damm, P. Standardized Loads Acting in Hip Implants. PLoS ONE 2016, 11, e0155612. [Google Scholar] [CrossRef]
- Ulrich, S.D.; Seyler, T.M.; Bennett, D.; Delanois, R.E.; Saleh, K.J.; Thongtrangan, I.; Kuskowski, M.; Cheng, E.Y.; Sharkey, P.F.; Parvizi, J.; et al. Total hip arthroplasties: What are the reasons for revision? Int. Orthop. 2008, 32, 597–604. [Google Scholar] [CrossRef] [Green Version]
- Davies, J.H.; Laflamme, G.Y.; Delisle, J.; Fernandes, J. Trabecular metal used for major bone loss in acetabular hip revision. J. Arthroplast. 2011, 26, 1245–1250. [Google Scholar] [CrossRef]
- Abolghasemian, M.; Tangsataporn, S.; Sternheim, A.; Backstein, D.; Safir, O.; Gross, A.E. Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss: A mid-term review. Bone Jt. J. 2013, 95, 166–172. [Google Scholar] [CrossRef]
- Jafari, S.M.; Bender, B.; Coyle, C.; Parvizi, J.; Sharkey, P.F.; Hozack, W.J. Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin. Orthop. Relat. Res. 2010, 468, 459–465. [Google Scholar] [CrossRef] [Green Version]
- Steno, B.; Kokavec, M.; Necas, L. Acetabular revision arthroplasty using trabecular titanium implants. Int. Orthop. 2015, 39, 389–395. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Delanois, R.E.; Gwam, C.U.; Mohamed, N.; Khlopas, A.; Chughtai, M.; Malkani, A.L.; Mont, M.A. Midterm Outcomes of Revision Total Hip Arthroplasty With the Use of a Multihole Highly-Porous Titanium Shell. J. Arthroplast. 2017, 32, 2806–2809. [Google Scholar] [CrossRef] [PubMed]
- Gallart, X.; Fernandez-Valencia, J.A.; Riba, J.; Bori, G.; Garcia, S.; Tornero, E.; Combalia, A. Trabecular TitaniumTM cups and augments in revision total hip arthroplasty: Clinical results, radiology and survival outcomes. HIP Int. 2016, 26, 486–491. [Google Scholar] [CrossRef] [PubMed]
- De Meo, F.; Cacciola, G.; Bellotti, V.; Bruschetta, A.; Cavaliere, P. Trabecular Titanium acetabular cups in hip revision surgery: Mid-term clinical and radiological outcomes. HIP Int. 2018, 28, 61–65. [Google Scholar] [CrossRef]
- Banerjee, S.; Issa, K.; Kapadia, B.H.; Pivec, R.; Khanuja, H.S.; Mont, M.A. Systematic review on outcomes of acetabular revisions with highly-porous metals. Int. Orthop. 2014, 38, 689–702. [Google Scholar] [CrossRef] [Green Version]
- Levine, B.; Della Valle, C.J.; Jacobs, J.J. Applications of porous tantalum in total hip arthroplasty. J. Am. Acad. Orthop. Surg. 2006, 14, 646–655. [Google Scholar] [CrossRef]
- Bruggemann, A.; Mallmin, H.; Bengtsson, M.; Hailer, N.P. Safety of Use of Tantalum in Total Hip Arthroplasty. J. Bone Jt. Surg. Am. 2020, 102, 368–374. [Google Scholar] [CrossRef]
- Bruggemann, A.; Fredlund, E.; Mallmin, H.; Hailer, N.P. Are porous tantalum cups superior to conventional reinforcement rings? Acta Orthop. 2017, 88, 35–40. [Google Scholar] [CrossRef] [Green Version]
- Pulido, L.; Rachala, S.R.; Cabanela, M.E. Cementless acetabular revision: Past, present, and future. Revision total hip arthroplasty: The acetabular side using cementless implants. Int. Orthop. 2011, 35, 289–298. [Google Scholar] [CrossRef] [Green Version]
- Goldman, A.H.; Armstrong, L.C.; Owen, J.R.; Wayne, J.S.; Jiranek, W.A. Does Increased Coefficient of Friction of Highly Porous Metal Increase Initial Stability at the Acetabular Interface? J. Arthroplast. 2016, 31, 721–726. [Google Scholar] [CrossRef]
- Engh, C.A.; O’Connor, D.; Jasty, M.; McGovern, T.F.; Bobyn, J.D.; Harris, W.H. Quantification of implant micromotion, strain shielding, and bone resorption with porous-coated anatomic medullary locking femoral prostheses. Clin. Orthop. Relat. Res. 1992, 285, 13–29. [Google Scholar] [CrossRef]
- MacKenzie, J.R.; Callaghan, J.J.; Pedersen, D.R.; Brown, T.D. Areas of contact and extent of gaps with implantation of oversized acetabular components in total hip arthroplasty. Clin. Orthop. Relat. Res. 1994, 298, 127–136. [Google Scholar] [CrossRef]
- Schwartz, J.T., Jr.; Engh, C.A.; Forte, M.R.; Kukita, Y.; Grandia, S.K. Evaluation of initial surface apposition in porous-coated acetabular components. Clin. Orthop. Relat. Res. 1993, 293, 174–187. [Google Scholar] [CrossRef]
- Ong, K.L.; Lehman, J.; Notz, W.I.; Santner, T.J.; Bartel, D.L. Acetabular cup geometry and bone-implant interference have more influence on initial periprosthetic joint space than joint loading and surgical cup insertion. J. Biomech. Eng. 2006, 128, 169–175. [Google Scholar] [CrossRef] [PubMed]
- Bondarenko, S.; Dedukh, N.; Filipenko, V.; Akonjom, M.; Badnaoui, A.A.; Schwarzkopf, R. Comparative analysis of osseointegration in various types of acetabular implant materials. HIP Int. 2018, 28, 622–628. [Google Scholar] [CrossRef]
- Lochel, J.; Janz, V.; Hipfl, C.; Perka, C.; Wassilew, G.I. Reconstruction of acetabular defects with porous tantalum shells and augments in revision total hip arthroplasty at ten-year follow-up. Bone Jt. J. 2019, 101, 311–316. [Google Scholar] [CrossRef]
- Fraile Suari, A.; Marques Lopez, F.; Cuenca Llavall, M.; Tey Pons, M.; Leon Garcia, A. Reconstruction for pelvic discontinuity and massive acetabular defects. Rev. Esp. Cir. Ortop. Traumatol. 2020, 64, 64–73. [Google Scholar] [CrossRef]
- Whitehouse, M.R.; Masri, B.A.; Duncan, C.P.; Garbuz, D.S. Continued good results with modular trabecular metal augments for acetabular defects in hip arthroplasty at 7 to 11 years. Clin. Orthop. Relat. Res. 2015, 473, 521–527. [Google Scholar] [CrossRef] [Green Version]
- Meneghini, R.M.; Hull, J.R.; Russo, G.S.; Lieberman, J.R.; Jiranek, W.A. Porous Tantalum Buttress Augments for Severe Acetabular Posterior Column Deficiency. Surg. Technol. Int. 2015, 27, 240–244. [Google Scholar]
- Konan, S.; Duncan, C.P.; Masri, B.A.; Garbuz, D.S. Porous tantalum uncemented acetabular components in revision total hip arthroplasty: A minimum ten-year clinical, radiological and quality of life outcome study. Bone Jt. J. 2016, 98, 767–771. [Google Scholar] [CrossRef]
- Chacko, V.; Agrawal, P.; Porter, M.L.; Board, T.N. Early results of a high friction surface coated uncemented socket in revision hip arthroplasty. HIP Int. 2019. [Google Scholar] [CrossRef] [PubMed]
- Naziri, Q.; Issa, K.; Pivec, R.; Harwin, S.F.; Delanois, R.E.; Mont, M.A. Excellent results of primary THA using a highly porous titanium cup. Orthopedics 2013, 36, e390–e394. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Interface | Augment/Cup | Bone/Augment | Bone/Cup | |||
---|---|---|---|---|---|---|
Implant Material | Titanium | Tantalum | Titanium | Tantalum | Titanium | Tantalum |
Load | Mean (SD) | Mean (SD) | Mean (SD) | Mean (SD) | Mean (SD) | Mean (SD) |
30% | 11.0 (1.9) | 22.5 (6.1) | 20.0 (7.3) | 29.7 (8.1) | 27.9 (8.0) | 46.3 (18.6) |
50% | 10.9 (2.1) | 24.7 (5.7) | 38.7 (17.8) | 31.7 (9.7) | 50.2 (18.6) | 51.4 (19.8) |
100% | 11.3 (4.2) | 23.7 (6.6) | 84.3 (40.2) | 39.4 (15.0) | 107.2 (44.0) | 61.6 (20.5) |
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Beckmann, N.A.; Bitsch, R.G.; Schonhoff, M.; Siebenrock, K.-A.; Schwarze, M.; Jaeger, S. Comparison of the Primary Stability of Porous Tantalum and Titanium Acetabular Revision Constructs. Materials 2020, 13, 1783. https://doi.org/10.3390/ma13071783
Beckmann NA, Bitsch RG, Schonhoff M, Siebenrock K-A, Schwarze M, Jaeger S. Comparison of the Primary Stability of Porous Tantalum and Titanium Acetabular Revision Constructs. Materials. 2020; 13(7):1783. https://doi.org/10.3390/ma13071783
Chicago/Turabian StyleBeckmann, Nicholas A., Rudi G. Bitsch, Mareike Schonhoff, Klaus-Arno Siebenrock, Martin Schwarze, and Sebastian Jaeger. 2020. "Comparison of the Primary Stability of Porous Tantalum and Titanium Acetabular Revision Constructs" Materials 13, no. 7: 1783. https://doi.org/10.3390/ma13071783
APA StyleBeckmann, N. A., Bitsch, R. G., Schonhoff, M., Siebenrock, K. -A., Schwarze, M., & Jaeger, S. (2020). Comparison of the Primary Stability of Porous Tantalum and Titanium Acetabular Revision Constructs. Materials, 13(7), 1783. https://doi.org/10.3390/ma13071783