Abstract
Purpose
To demonstrate key integrative advances in microscope-integrated intraoperative optical coherence tomography (iOCT) technology that will facilitate adoption and utilization during ophthalmic surgery.
Methods
We developed a second-generation prototype microscope-integrated iOCT system that interfaces directly with a standard ophthalmic surgical microscope. Novel features for improved design and functionality included improved profile and ergonomics, as well as a tunable lens system for optimized image quality and heads-up display (HUD) system for surgeon feedback. Novel material testing was performed for potential suitability for OCT-compatible instrumentation based on light scattering and transmission characteristics. Prototype surgical instruments were developed based on material testing and tested using the microscope-integrated iOCT system. Several surgical maneuvers were performed and imaged, and surgical motion visualization was evaluated with a unique scanning and image processing protocol.
Results
High-resolution images were successfully obtained with the microscope-integrated iOCT system with HUD feedback. Six semi-transparent materials were characterized to determine their attenuation coefficients and scatter density with an 830 nm OCT light source. Based on these optical properties, polycarbonate was selected as a material substrate for prototype instrument construction. A surgical pick, retinal forceps, and corneal needle were constructed with semi-transparent materials. Excellent visualization of both the underlying tissues and surgical instrument were achieved on OCT cross-section. Using model eyes, various surgical maneuvers were visualized, including membrane peeling, vessel manipulation, cannulation of the subretinal space, subretinal intraocular foreign body removal, and corneal penetration.
Citation: Ehlers JP, Srivastava SK, Feiler D, Noonan AI, Rollins AM, Tao YK (2014) Integrative Advances for OCT-Guided Ophthalmic Surgery and Intraoperative OCT: Microscope Integration, Surgical Instrumentation, and Heads-Up Display Surgeon Feedback. PLoS ONE 9(8): e105224. https://doi.org/10.1371/journal.pone.0105224
Editor: Alfred S. Lewin, University of Florida, United States of America
Received: April 20, 2014; Accepted: July 20, 2014; Published: August 20, 2014
Copyright: © 2014 Ehlers et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The authors confirm that all data underlying the findings are fully available without restriction. All relevant data are within the paper and its supporting information files.
Funding: This work was funded by the National Institutes of Health/National Eye Institute (http://www.nei.nih.gov/) K23-EY022947-01A1 (JPE); Ohio Department of Development TECH-13-059 (http://development.ohio.gov/bs_thirdfrontier/ipp.htm) (JPE, SKS, YKT); National Institutes of Health/Natonal Eye Institute (http://www.nei.nih.gov/) R01-EY023039-01 (JPE, SKS, YKT). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: Co-authors JP Ehlers, SK Srivastava, and YK Tao are co-inventors on intellectual property related to the intraoperative OCT technology described in this manuscript: Provisional Patents Filed: 1. Optimized systems and methods for surgical microscope integrated optical coherence tomography Tao, YK, Ehlers, JP, Srivastava, SK Assignee: The Cleveland Clinic Foundation Filing Date: 2013-08-22 US 61/868,645 Patent Applications: 1. Imaging and visualization systems, instruments, and methods using optical coherence tomography Izatt, JA, Toth, CA, Farsiu, S, Hahn, P, Tao, YK, Ehlers, JP, Migacz, JV, Chiu, SJ Assignee: Duke University Filing Date: 2012-01-19 US 20120184846 2. Surgical instruments for OCT assisted procedures Ehlers, JP, Srivastava, SK, Tao, YK Assignee: The Cleveland Clinic Foundation Filing Date: 2013-05-06 US 20130296694 Issued Patent: 1. System and methods for surgical microscope and optical coherence tomography (OCT) imaging Izatt, JA, Tao, YK, Toth, CA Assignee: Duke University Issue Date: 2013-02-05 US 8366271. Co-authors JP Ehlers, SK Srivastava, and YK Tao have licensed intellectual property to and JP Ehlers and SK Srivastava receive royalties from Bioptigen, Inc. not related to the intraoperative OCT technology described in this manuscript. Co-authors JP Ehlers, SK Srivastava, and YK Tao are also currently part of a joint development agreement with Synergetics USA, Inc. for the development of semitransparent OCT-compatible ophthalmic surgical instruments. None of the co-authors' Competing Interests alters the authors' adherence to PLOS ONE policies on data and materials sharing nor does it impact the scientific rigor of the work presented here.
Introduction
OCT has revolutionized the clinical practice of ophthalmology. OCT provides non-invasive high-resolution optical biopsy of tissue microstructure that facilitates disease diagnosis and guides therapeutic decision-making [1]. Seamless integration of this technology into ophthalmic surgery has the potential to be the foundation for a new paradigm in the surgical management of ophthalmic disease.
Over the last several years, intraoperative OCT (iOCT) has emerged as a viable adjunct to ophthalmic surgery. Anterior segment and vitreoretinal surgical procedures have been imaged with iOCT, and macular holes, vitreomacular traction, retinal detachment, and epiretinal membranes have all been visualized with iOCT in early foundational studies [2]–[9]. Intrasurgical feedback using iOCT during lamellar keratoplasty procedures has also been described [10]–[13].
Early applications of iOCT used portable OCT systems to provide intraoperative visualization either through a handheld, externally mounted, or microscope-mounted systems [2]–[4], [6]–[11]. Novel designs for handheld ophthalmic imaging systems have also been presented, which use state-of-the-art light-sources or multimodality imaging, and have potential applications in intraoperative imaging by improving imaging speed and ergonomics [14], [15]. Although these systems provide excellent visualization of the anatomic alterations that occur during surgery, they are limited by the “pause” required to bring the system into alignment, and “real-time” visualization of surgical maneuvers is not feasible with these systems.
In addition to microscope-mounted imaging systems, integrated approaches have also been presented that combine optics with prototype surgical instruments for intraoperative imaging and guidance. These include fiber probes [16], [17] for intraocular OCT imaging and integrated metrology and feedback systems for guiding instrument positions within the eye [18], [19]. Although these instruments may provide useful intraoperative feedback, their utility in guiding ophthalmic surgical maneuvers have not been demonstrated. Furthermore, fiber probe based instruments have traditionally been limited by inherent trade-offs in imaging field-of-view (FOV), sampling density, and resolution, and manual positioning and aiming of probes results in poor imaging stability, all of which needs to be addressed prior to clinical translation.
In order to address these limitations, microscope-integrated iOCT systems have been developed [5], [9], [20]–[23]. Preliminary integrated systems allowed for visualization of surgical instrumentation and instrument-tissue interactions [23]–[25]. However, the development of these landmark systems also revealed additional hurdles that must be overcome to achieve true seamless integration, including optimal ergonomic impact, heads-up display (HUD), improved focus control, and OCT-friendly surgical instrumentation [5], [20].
In this report, we describe several iterative advances in iOCT technology, including a second-generation microscope-integrated iOCT system, OCT-friendly surgical instrumentation, optimized dynamic iOCT imaging, and HUD visualization.
Methods
Microscope-Integrated iOCT System
The microscope-integrated iOCT system was implemented as an attachment to a Leica M844 (Wezlar, Germany) ophthalmic surgical microscope (Figure 1) [26]. The surgical microscope provides 3.5–21x motorized magnification for anterior segment viewing, and additional magnification for the posterior segment is set by the use of contact/non-contact ophthalmic lenses. An electrically tunable lens allowed for real-time focus control, which can compensate for ∼13.9 D of defocus for optimized imaging of both anterior and posterior segment and fine control of the OCT focus independent of the surgical view. Anterior segment iOCT visualization was achieved through the objective lens of the ophthalmic surgical microscope. Posterior segment iOCT visualization was obtained with either an Oculus BIOM3 (Wezlar, Germany) adapter for wide-field indirect ophthalmoscopy or contact lens for high magnification viewing (Volk, Mentor, OH). The relay optics in the custom-built scan-head focuses to a 24 µm spot during anterior segment imaging through a WD 200 mm objective lens or 1.4 mm collimated spot at the pupil for posterior segment imaging through a BIOM widefield ophthalmic lens (53565/WFE). All OCT images were acquired with 700 µW optical power at 830 nm center wavelength with 106 dB SNR at a line-rate of 50 kHz, a 5.2 µm axial resolution in air, and 1.14 mm 6 dB falloff. Pathlength changes for anterior and posterior segment imaging was adjusted by the imaging technician using a manually adjustable translation stage in the reference arm. Custom software (Bioptigen, Durham, NC) was utilized to perform real-time data acquisition, processing, archiving, and display.
(A) Color photograph of system (red circle) on Leica 844 microscope. (B) Schematic of iOCT system, including tunable lens and monolithic design. DM, dichroic mirror; f, collimating, objective, scan, tube, and electrically tunable lenses; G, galvanometer scanners; M, fold mirrors.
Utilizing the DI C800 (Leica Microsystems, Heerbrugg, Switzerland), a VGA display was also folded into the optical path of the microscope ocular, which provided overlaid OCT images for a customizable HUD for surgeon feedback. Various approaches to HUD were tested for surgeon feedback, including aiming targets and live display of OCT cross-sectional data.
OCT-Compatible Instrumentation
Six semitransparent materials were imaged using a Bioptigen SDOIS system (830±30 nm, 50 kHz line-rate). The materials were translated at 1 micron increments over a 1.5 mm range relative to a fixed focal plane, and B-scans were acquired at each position (Figure 2). Intensity changes resulting from fall-off and defocus were corrected in post-processing using experimentally measured values. Each B-scan was then averaged to remove speckle noise, and the resulting scattering profile was fit to a Beer's law model using least-squares minimization to calculate the attenuation coefficient [27]. The scattering density was calculated by measuring the percentage, by volume, of scatterers with intensities two standard deviations above the OCT noise floor.
(A) OCT set-up with single material in place for evaluation of attenuation coefficient and scattering density. (B) Sample set of material cylinders tested for optical properties.
Based on the material testing, OCT-compatible surgical implement prototypes were developed and tested. Surgical instruments tested included a corneal needle (specifically designed for deep anterior lamellar keratoplasty), surgical pick, and retinal forceps.
Intraoperative Imaging Protocol
Various scan methods were utilized to visualize instrument motion and surgical maneuvers. As previously described, a surgical maneuver scanning protocol captured 5 B-scans, each comprised of 256 A-scans, along the axis of the instrument over a 10.0×1.0 mm FOV. The scans were oriented such that the densely sampled, long scanning axis (i.e., 10.0 mm) was aligned parallel to the axis of the surgical instrument. The direction of motion associated with the surgical maneuver was also constrained to this axis [24], [28]. These subsampled volumetric datasets were acquired at 39 volumes per second at the instrument tip and displayed in real-time for surgical guidance. The HUD system was utilized with various targets to assess localization to the area of interest. Additional scan protocols were also performed with iOCT B-scans oriented perpendicular to the long axis of the instrument. Post-processing included image averaging, registration, and spatial compounding with ImageJ (freeware; National Institutes of Health; Bethesda, MD) and Matlab 2010a (Matlab, MathWorks, Natick, MA).
Model eye studies
Fresh cadaveric porcine eyes were obtained for imaging (J. H. Routh Packing, Sandusky, OH). To optimize corneal clarity and image quality, iOCT scanning was performed within 6 hours of harvesting the eyes. Despite efforts to acquire images immediately after enucleation, some image quality variation was observed in the surgical microscope view and on iOCT as a result of corneal defects and lens opacity in the cadaveric eye. Each eye was positioned under the microscope-integrated iOCT system using a suction platform to maintain fixation. A 23-gauge valved trocar cannula system was utilized for all posterior segment surgical maneuvers. Cannulas were placed in the standard location for 3-port vitrectomy, including infusion placement in the inferotemporal quadrant. A VersaVIT (Synergetics, St. Louis, MO) was used for infusion and vitrectomy as needed. A 23-gauge fiberoptic light pipe was used for illumination and surgical instruments were introduced through the final cannula. Initial focus was achieved by the surgeon using the BIOM or microscope pedals if a contact lens was used. Once gross focus was achieved, the tunable lens was used to for fine adjustment of the OCT image.
Volume scanning was performed prior to instrument manipulation with each instrument at multiple locations: above the cornea, within the cornea, within the anterior chamber, superficial to the retina in vitreous, and on the retinal surface. Static volumetric scans (10×10 mm) sampled at 500×500 pix. (A-scans × B-scans) were obtained for densely-sampled visualization of tissue-instrument interactions in post-processing. Dynamic imaging was performed during surgical manipulations with various instruments including a transretinal needle, diamond dusted membrane scraper (Synergetics, O'fallon, MO), vitreoretinal forceps. These maneuvers included simulated membrane peeling, vessel manipulation, cannulation of the subretinal space, subretinal intraocular foreign body removal, and corneal penetration. All simulated surgical maneuvers were performed by a trained vitreoretinal surgeon and the iOCT system was operated by an imaging technician adjacent to the surgical table.
Results
Intraoperative Imaging and Heads-up Visualization
The microscope-integrated iOCT system provided high-resolution images of the anterior chamber and retina in cadaveric porcine eyes. Posterior segment images were obtained with both direct and indirect lens systems, and the tunable lens provided significant flexibility and versatility for optimizing image quality. Instruments were successfully visualized in numerous anatomic locations without difficulty, including outside the eye, within the cornea, in the anterior chamber, in the vitreous cavity, and on the retinal surface.
The HUD system was successfully provided two separate feedback mechanisms to the surgeon. The first was a targeting box to allow the surgeon to identify the coordinates of the iOCT scan to help co-localize the instrument-tissue interaction with the iOCT FOV (Figure 3A, Video S1). This provided improved efficiency and coordination between the surgeon and technician for visualizing surgical maneuvers.
(A) Ocular feedback with target region for surgeon to area of iOCT FOV. (B) Second ocular feedback system with OCT data displayed adjacent to surgeon's view of the surgical field with OCT display both on axis and perpendicular to instrument. Video S1. (Note: contrast non-uniformity in the iOCT HUD images are a result of capturing video snapshots through the surgical microscope ocular). Scale bar: 200 µm.
The second HUD feedback mechanism provided real-time iOCT overlays, which allow the surgeon to simultaneously view the surgical field and iOCT B-scans at the area of interest (Figure 3B). This provided important feedback to the surgeon regarding depth information (e.g., corneal needle penetration) and relative location of the instrument to the tissue of interest. Software control of the HUD overlays allowed live display of iOCT B-scans during aiming and data acquisition, real-time scaling of iOCT display sizes relative to the surgical field, and review of sequential B-scans in a volumetric dataset.
Material Testing and Surgical Instrument Prototyping
The measured optical properties of the six semitransparent materials (polycarbonate, PVC, Radel, Ultem, polysofonate, and PETG) showed comparable attenuation coefficients with values ranging from 2.16–3.31 mm−1. These values were consistent with the semitransparent nature of the materials within the OCT wavelength range tested. The index of refraction for each material was also measured at 830 nm, and these values ranged between n = 1.51–1.65. The materials showed dramatically different scattering densities, ranging from 0.13–23.2% by volume, which can be clearly distinguished on OCT cross-sections (Figure 4).
The measured attenuation coefficients (yellow) ranged between 2.16–3.31 mm−1 and scattering densities ranged from 0.13–23.2% (blue). Index of refraction for each material at 830 nm were also measured. Representative OCT cross-sections of each material are shown.
Prototype instruments were successfully created, including retinal forceps, a surgical pick (Figures 5 and 6; Video S2), and an OCT-compatible needle (Figure 7; Video S3). The prototype instruments in this study were machined to accommodate a 23-gauge valved trocar cannula for simulated vitreoretinal surgery. Specifically, all instruments included a 640 µm diameter metallic shaft and 450 µm diameter semitransparent tip. The needle had an inner diameter of ∼180 µm. OCT B-scans revealed excellent visualization of the underlying tissues in addition to the tip of the instruments. This confirmed optimal light scattering and transmission properties of the substrate materials.
(A) B-scan of prototype surgical pick (orange arrow) in the porcine anterior chamber between the endothelium and the anterior lens capsule. (B) B-scan of prototype surgical pick exerting significant downward pressure on the porcine retinal surface. In both frames, excellent visualization of the instrument profile and underlying tissues is achieved. Images show averages of 5 co-registered adjacent B-scans. Scale bar: 200 µm.
(B) Summed voxel projection showing perpendicular OCT scans (white lines) at needle tip. (C) Intraoperative OCT B-scan parallel to needle from (B) revealing excellent visualization of the needle while maintaining optimal clarity of the underlying tissue. The bore of the needle (white asterisk) is visible on the OCT scan. (D) Intraoperative OCT B-scan perpendicular to needle tip from (B) revealing tissue disruption at the needle tip (i.e., retinal hole) with excellent visualization of underlying and surrounding tissues. Video S2. Scale bar: 50 µ.
(B) Summed voxel projection showing perpendicular OCT scans (white lines) at tip of surgical pick. (C, D) Intraoperative OCT B-scan parallel (C) and perpendicular (D) to pick from (B) revealing excellent visualization of the pick tip (white asterisk) while maintaining optimal clarity of the underlying tissue. Video S3. Scale bar: 500 µm.
Intraoperative Maneuver Visualization and Motion-Capture
Successful imaging of conventional surgical instruments was obtained using the microscope-integrated iOCT system. Corneal needle penetration for simulated deep anterior lamellar keratoplasty was captured with dynamic cross-sectional and static volumetric imaging (Figure 8). Vitreoretinal procedures performed using the diamond dusted membrane scraper (Synergetics, O'fallon, MO) was imaged using spatially compounded cross-sections and densely sampled three-dimensional volumes for visualization of dynamic and static datasets in post-processing. The iOCT images were similar to those previously presented, showing excellent transmission through areas of bare silicone with increased shadowing at the diamond dusted tip [24], [25]. Utilizing the HUD system, images of the membrane scraper could be rapidly obtained, and dynamic imaging was used to demonstrate intrasurgical instrument motion on the retinal surface (Figure 9, Video S1). Additional simulated vitreoretinal surgical maneuvers captured with dynamic imaging included membrane peeling, vascular manipulation, subretinal intraocular foreign body removal, and subretinal space cannulation with injection of triamcinolone (Figure 10, Video S5). Spatial compounding enhanced surgical instrument visualization and minimized underlying shadowing.
(B) Summed voxel projection showing perpendicular OCT scans (white lines) at tip of needle. (C, D) Intraoperative OCT B-scan parallel (C) and perpendicular (D) to needle from (B) revealing visualization of the penetration depth of the needle tip (white asterisk) relative to the corneal endothelium. Video S4. Scale bar: 500 µm.
Each frame shows progressive OCT-motion (below) corresponding to the surgical view with HUD of OCT target (above). Video S1. Scale bar: 200 µm.
Cross-sectional images show (A) semitransparent needle tip entering the subretinal space, (B) initial injection volume displacing surrounding tissue, (C) injection into the subretinal space with elevation of the retina. (D) Immediately post-injection, excess fluid and triamcinolone is observed leaking from the injection site (arrow). Video S5. Scale bar: 500 µm.
Discussion
In this study, we demonstrated several iterative advances in integrative technology for iOCT. Application of a HUD system represents a major advance in integrative technology. The ability to visualize iOCT data without the use of an external monitor significantly enhanced ergonomics by providing immediate feedback to surgeons through the microscope oculars without needing to look away from the surgical field. In addition to the iOCT system, we characterized the optical properties of several materials and demonstrated the feasibility of OCT-compatible instrumentation, which improves visualization of underlying tissue structures and instrument-tissue interactions.
Dynamic imaging with spatial compounding provided outstanding visualization of numerous ophthalmic surgical maneuvers. However, spatial averaging over adjacent B-scans, acquired across the width of a surgical instrument, limits high-speed visualization of surgical dynamics and may introduce averaging artifacts. Recent development of novel high-speed OCT light sources [29], [30] and real-time volumetric OCT processing and visualization algorithms [31], [32] may potentially allow for real-time acquisition and visualization of volumetric datasets for imaging intraoperative maneuvers. However, visualization of three-dimensional data conventionally requires real-time rotation of rendered datasets and serial fly-throughs of cross-sectional images, which may prove cumbersome and distracting for surgical guidance. Potential visualization methods may include automated surgical instrument tracking for iOCT or HUD overlays of en face instrument positions calculated using segmentation of retinal layers and instrument surfaces. Future studies are required to explore the utility of different methods for intraoperative guidance.
Our preliminary testing indicates that materials with minimal attenuation and refractive index are optimal candidates for OCT-compatible instruments. Minimal attenuation results in reduced signal loss and shadowing of tissue structures underneath the instrument, and low refractive index results in smaller axial image jump as a result of optical pathlength differences between the instrument material and air/vitreous. Initial prototype instruments were machined from polycarbonate, but PETG may also be an appropriate candidate material because of similarities in their optical properties at the OCT wavelength range tested. Scatterer density may also be taken into account because a combination of low attenuation and high scattering may provide enhanced OCT contrast. However, machining methods, such as surface roughing, or adding scattering dopants into the substrate material may allow for more flexibility in designing specific scattering properties into OCT-compatible instruments.
Vitreoretinal surgery has made remarkable progress over the last few decades. Advances in surgical microscopes, small-gauge instrumentation, illumination and vital dyes have resulted in significant improvements in the ophthalmic surgeon's toolbox and patient outcomes [33]–[36]. As OCT has transformed the landscape of clinical ophthalmology, true seamless integration of iOCT into the operating room may be the next paradigm shift in ophthalmic surgery.
Preliminary research suggests that iOCT may provide critical information regarding the impact of surgical maneuvers, intrasurgical architectural dynamics, and pathophysiology of surgical diseases and may impact surgical decision-making [2]–[6], [8], [37]. Using iOCT, studies have examined multiple vitreoretinal diseases including vitreomacular traction syndrome, macular holes, retinal detachment, epiretinal membrane, optic pit-related maculopathy, and ROP [2], [3], [5], [8], [23], [37], [38]. The dynamic nature of surgery provides a unique opportunity to study the pathophysiology of underlying diseases and architectural changes that occur during surgical manipulation. These underlying dynamics have been demonstrated with iOCT for optic pit-related maculopathy, retinal detachment, and macular holes [2], [4], [37].
Utilizing information obtained from iOCT, surgeon decision-making may be altered and patient outcomes may be impacted. The impact of iOCT on surgeon decision-making has been demonstrated in vitreomacular traction and epiretinal membrane surgery (Ehlers et al, IOVS 2013. 54: ARVO E-abstract 3308) [3]. In these reports, iOCT demonstrated potential impact on patient outcomes by aiding surgical decision-making (e.g., additional membrane peeling, gas tamponade). Additionally, feedback during deep anterior lamellar keratoplasty regarding dissection depth may be helpful in optimizing the procedure [13].
In order for iOCT to become a key component of ophthalmic surgery, significant improvements are needed in integration to minimize disruption to work-flow while maximizing surgical feedback. Previously described microscope-integrated OCT systems were landmark advances for integrative solutions [5], [20]. However, significant shortcomings still existed in the system-surgeon interaction platform. Namely, HUD systems, OCT-compatible instrumentation, automated aiming, customizable focus control, and software analysis platforms were all lacking.
In this report, we demonstrate significant iterative progress for many of these hurdles, particularly related to focus control, HUD systems, and OCT-compatible instrumentation. We have previously demonstrated novel algorithms for pathology segmentation that may be applied to identify the rapid alterations that occur during surgical intervention [4], [39]. Automated segmentation of retinal layers has been previously described [40]. The application of similar algorithms could potentially be utilized for real-time surgical feedback on instrument-tissue proximity. Automated layer segmentation could potentially be utilized to track intra-tissue depth and open doors to new surgical procedures (e.g., targeted drug delivery).
The systems described in this report represent significant but preliminary advances in integrative technology. Additional progress is needed in automated aiming and tracking of surgical instrumentation. Tailored image-feedback is also needed to optimize the HUD system for optimal performance and minimizing information overload. The OCT-compatible instruments represent proofs-of-concept and still need additional iterations to maximize functionality and expand the surgical tool repertoire. Specifically, material properties will need to be rigorously examined to identify differences between the semi-transparent plastics identified in this manuscript versus conventional metallic instruments. These material properties, including hardness and tensile strength, will be critical in identifying substrates that may be easily manufactured and machined to tolerances acceptable for surgical implementation, specifically in instrument precision and durability. Our simulated corneal surgery using an OCT-compatible needle identified potential differences in the overall feel compared to conventional metallic needles, which will need to be addressed in future studies.
From femtosecond cataract surgery to ultra-high speed vitrectomy, intrasurgical technology has experienced significant growth in the past several years. iOCT is an emerging technology, and interest in the application of OCT to surgical care continues to gain momentum. Currently, two microscope-integrated systems have approval in the EU for human use: the Haag-Streit iOCT and Carl Zeiss Meditec RESCAN 700. The Haag-Streit iOCT system utilizes a camera port add-on OCT system from OPMedT [9]. Carl Zeiss Meditec recently developed a second-generation iOCT prototype, the RESCAN 700, that is built on the Lumera 700 platform [9], [23]. The Haag-Streit system uses an external monitor, whereas the RESCAN 700 provides a HUD system for surgical use. At the time of this report, neither of these systems is FDA-approved in the United States.
iOCT represents an exciting opportunity for expansion of our knowledge base of surgical ophthalmic conditions and a potential paradigm shift in ophthalmic surgical management. This study demonstrates subsequent integrative advances that may be critical to the widespread assimilation of this technology into every day surgical practice, including an optimized form-factor, OCT-compatible instrumentation, and a HUD display feedback. Further research and development is needed to develop software platforms, automated instrument tracking, and surgical instruments in order to continue to optimally integrate this technology seamlessly into the operating room theater.
Supporting Information
Video S1.
Heads-up display feedback for intraoperative OCT (iOCT) system. Microscope ocular feedback with target region for surgeon to area of iOCT field of view (top). Simultaneous iOCT video of instrument-tissue interaction (bottom).
https://doi.org/10.1371/journal.pone.0105224.s001
(MOV)
Video S2.
Three dimensional reconstruction of OCT-compatible needle in porcine eye (top left). Summed voxel projection showing perpendicular OCT scans at needle tip (top right). Intraoperative OCT B-scan, parallel to needle, reveals excellent visualization of the needle while maintaining optimal clarity of the underlying tissue (bottom left). Intraoperative OCT B-scan, perpendicular to needle tip, reveals tissue disruption at the needle tip (i.e., retinal hole) with excellent visualization of underlying and surrounding tissues (bottom right).
https://doi.org/10.1371/journal.pone.0105224.s002
(MOV)
Video S3.
Three dimensional reconstruction of OCT-compatible surgical pick in porcine eye (top left). Summed voxel projection showing perpendicular OCT scans at tip of surgical pick (top right). Intraoperative OCT B-scan parallel (bottom left) and perpendicular (bottom right) to surgical pick revealing excellent visualization of the pick tip while maintaining optimal clarity of the underlying tissue.
https://doi.org/10.1371/journal.pone.0105224.s003
(MOV)
Video S4.
Three dimensional reconstruction of OCT-compatible needle in porcine cornea (top left). Summed voxel projection showing perpendicular OCT scans at tip of needle (top right). Intraoperative OCT B-scan parallel (bottom left) and perpendicular (bottom right) to needle captures the penetration depth of the needle tip relative to the corneal endothelium.
https://doi.org/10.1371/journal.pone.0105224.s004
(MOV)
Video S5.
Subretinal cannulation and injection in porcine eye. Cross-sectional images show semitransparent needle tip entering the subretinal space, initial injection volume displacing surrounding tissue, injection into the subretinal space with elevation of the retina. Immediately post-injection, excess fluid and triamcinolone is observed leaking from the injection site.
https://doi.org/10.1371/journal.pone.0105224.s005
(MOV)
Author Contributions
Conceived and designed the experiments: JPE SKS AMR YKT. Performed the experiments: JPE SKS DF AIN YKT. Analyzed the data: JPE SKS DF AIN AMR YKT. Contributed reagents/materials/analysis tools: JPE SKS AMR YKT. Contributed to the writing of the manuscript: JPE SKS DF AIN AMR YKT.
References
- 1. Chen TC, Cense B, Pierce MC, Nassif N, Park BH, et al. (2005) Spectral domain optical coherence tomography: ultra-high speed, ultra-high resolution ophthalmic imaging. Arch Ophthalmol 123: 1715–1720.
- 2. Ehlers JP, Ohr MP, Kaiser PK, Srivastava SK (2013) Novel microarchitectural dynamics in rhegmatogenous retinal detachments identified with intraoperative optical coherence tomography. Retina 33: 1428–1434.
- 3.
Ehlers JP, Tam T, Kaiser PK, Martin DF, Smith G, et al.. (2014) Utility of intraoperative optical coherence tomography during vitrectomy surgery for vitreomacular traction syndrome. Retina: In press.
- 4.
Ehlers JP, Xu D, Kaiser PK, Singh RP, Srivastava SK (2013) INTRASURGICAL DYNAMICS OF MACULAR HOLE SURGERY: An Assessment of Surgery-Induced Ultrastructural Alterations with Intraoperative Optical Coherence Tomography. Retina.
- 5. Binder S, Falkner-Radler CI, Hauger C, Matz H, Glittenberg C (2011) Feasibility of intrasurgical spectral-domain optical coherence tomography. Retina 31: 1332–1336.
- 6. Dayani PN, Maldonado R, Farsiu S, Toth CA (2009) Intraoperative use of handheld spectral domain optical coherence tomography imaging in macular surgery. Retina 29: 1457–1468.
- 7. Lee LB, Srivastava SK (2011) Intraoperative spectral-domain optical coherence tomography during complex retinal detachment repair. Ophthalmic Surg Lasers Imaging 42 Online e71–74.
- 8. Ray R, Baranano DE, Fortun JA, Schwent BJ, Cribbs BE, et al. (2011) Intraoperative microscope-mounted spectral domain optical coherence tomography for evaluation of retinal anatomy during macular surgery. Ophthalmology 118: 2212–2217.
- 9. Ehlers JP, Tao YK, Srivastava SK (2014) The Value of Intraoperative OCT Imaging in Vitreoretinal Surgery. Current Opinions in Ophthalmology In press
- 10. De Benito-Llopis L, Mehta JS, Angunawela RI, Ang M, Tan DT (2014) Intraoperative anterior segment optical coherence tomography: a novel assessment tool during deep anterior lamellar keratoplasty. Am J Ophthalmol 157: 334–341 e333.
- 11. Knecht PB, Kaufmann C, Menke MN, Watson SL, Bosch MM (2010) Use of intraoperative fourier-domain anterior segment optical coherence tomography during descemet stripping endothelial keratoplasty. Am J Ophthalmol 150: 360–365 e362.
- 12. Steven P, Le Blanc C, Velten K, Lankenau E, Krug M, et al. (2013) Optimizing descemet membrane endothelial keratoplasty using intraoperative optical coherence tomography. JAMA Ophthalmol 131: 1135–1142.
- 13. Scorcia V, Busin M, Lucisano A, Beltz J, Carta A, et al. (2013) Anterior segment optical coherence tomography-guided big-bubble technique. Ophthalmology 120: 471–476.
- 14. Lu CD, Kraus MF, Potsaid B, Liu JJ, Choi W, et al. (2013) Handheld ultrahigh speed swept source optical coherence tomography instrument using a MEMS scanning mirror. Biomed Opt Express 5: 293–311.
- 15. Larocca F, Nankivil D, Farsiu S, Izatt JA (2013) Handheld simultaneous scanning laser ophthalmoscopy and optical coherence tomography system. Biomed Opt Express 4: 2307–2321.
- 16. Joos KM, Shen JH (2013) Miniature real-time intraoperative forward-imaging optical coherence tomography probe. Biomed Opt Express 4: 1342–1350.
- 17. Han S, Sarunic MV, Wu J, Humayun M, Yang C (2008) Handheld forward-imaging needle endoscope for ophthalmic optical coherence tomography inspection. Journal of Biomedical Optics 13: 020505–020505-020503.
- 18. Song C, Park DY, Gehlbach PL, Park SJ, Kang JU (2013) Fiber-optic OCT sensor guided “SMART” micro-forceps for microsurgery. Biomed Opt Express 4: 1045–1050.
- 19. Balicki M, Han JH, Iordachita I, Gehlbach P, Handa J, et al. (2009) Single fiber optical coherence tomography microsurgical instruments for computer and robot-assisted retinal surgery. Med Image Comput Comput Assist Interv 12: 108–115.
- 20. Ehlers JP, Tao YK, Farsiu S, Maldonado R, Izatt JA, et al. (2011) Integration of a spectral domain optical coherence tomography system into a surgical microscope for intraoperative imaging. Invest Ophthalmol Vis Sci 52: 3153–3159.
- 21. Tao Y, Ehlers J, Toth C, Izatt J (2010) Intraoperative spectral domain optical coherence tomography for vitreoretinal surgery. Opt Lett 35: 3315–3317.
- 22. Hahn P, Migacz J, O'Donnell R, Day S, Lee A, et al. (2013) Preclinical evaluation and intraoperative human retinal imaging with a high-resolution microscope-integrated spectral domain optical coherence tomography device. Retina 33: 1328–1337.
- 23. Ehlers JP, Kaiser PK, Srivastava SK (2014) Intraoperative Optical Coherence Tomography Utilizing the RESCAN 700: Preliminary Results from the DISCOVER study. Br J Ophthalmol In Press
- 24. Ehlers JP, Tao YK, Farsiu S, Maldonado R, Izatt JA, et al. (2013) Visualization of real-time intraoperative maneuvers with a microscope-mounted spectral domain optical coherence tomography system. Retina 33: 232–236.
- 25. Hahn P, Migacz J, O'Connell R, Izatt JA, Toth CA (2013) Unprocessed real-time imaging of vitreoretinal surgical maneuvers using a microscope-integrated spectral-domain optical coherence tomography system. Graefes Arch Clin Exp Ophthalmol 251: 213–220.
- 26. Tao YK, Srivastava SK, Ehlers JP (2014) Microscope-integrated intraoperative OCT with electrically tunable focus and heads-up display for imaging of ophthalmic surgical maneuvers. Biomed Opt Express 5: 1877–1885.
- 27. Faber DJ, van der Meer FJ, Aalders MCG, van Leeuwen TG (2004) Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography. Optics Express 12: 4353–4365.
- 28. Tao YK, Ehlers JP, Toth CA, Izatt JA (2011) Visualization of vitreoretinal surgical manipulations using intraoperative spectral domain optical coherence tomography; 78890F-78890F-78810.
- 29. Klein T, Wieser W, Reznicek L, Neubauer A, Kampik A, et al. (2013) Multi-MHz retinal OCT. Biomed Opt Express 4: 1890–1908.
- 30. Grulkowski I, Liu JJ, Potsaid B, Jayaraman V, Lu CD, et al. (2012) Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers. Biomed Opt Express 3: 2733–2751.
- 31. Kang JU, Huang Y, Zhang K, Ibrahim Z, Cha J, et al. (2012) Real-time three-dimensional Fourier-domain optical coherence tomography video image guided microsurgeries. J Biomed Opt 17: 081403–081401.
- 32. Jian Y, Wong K, Sarunic MV (2013) Graphics processing unit accelerated optical coherence tomography processing at megahertz axial scan rate and high resolution video rate volumetric rendering. J Biomed Opt 18: 26002.
- 33. Kadonosono K, Itoh N, Uchio E, Nakamura S, Ohno S (2000) Staining of internal limiting membrane in macular hole surgery. Arch Ophthalmol 118: 1116–1118.
- 34. Machemer R, Parel JM (1978) An improved microsurgical ceiling-mounted unit and automated television. Am J Ophthalmol 85: 205–209.
- 35. Machemer R, Parel JM, Hickingbotham D, Nose I (1981) Membrane peeler cutter. Automated vitreous scissors and hooked needle. Arch Ophthalmol 99: 152–153.
- 36. Parel JM, Machemer R, Aumayr W (1974) A new concept for vitreous surgery. 4. Improvements in instrumentation and illumination. Am J Ophthalmol 77: 6–12.
- 37. Ehlers JP, Kernstine K, Farsiu S, Sarin N, Maldonado R, et al. (2011) Analysis of pars plana vitrectomy for optic pit-related maculopathy with intraoperative optical coherence tomography: a possible connection with the vitreous cavity. Arch Ophthalmol 129: 1483–1486.
- 38. Chavala SH, Farsiu S, Maldonado R, Wallace DK, Freedman SF, et al. (2009) Insights into advanced retinopathy of prematurity using handheld spectral domain optical coherence tomography imaging. Ophthalmology 116: 2448–2456.
- 39. Xu D, Yuan A, Kaiser PK, Srivastava SK, Singh RP, et al. (2013) A novel segmentation algorithm for volumetric analysis of macular hole boundaries identified with optical coherence tomography. Invest Ophthalmol Vis Sci 54: 163–169.
- 40. Chiu S, Li X, Nicholas P, Toth C, Izatt J, et al. (2010) Automatic segmentation of seven retinal layers in SDOCT images congrouent with expert manual segmentation. Optics express 18: 19413–19428.