Abstract
Recently an in vivo high-resolution backward-mode photoacoustic microscope was developed that shows potential for applications in dermatology and related cancer research. However, the limited depth of focus of the large-numerical-aperture (NA) ultrasonic lens employed in this system causes the image quality to deteriorate significantly in the out-of-focus region. To solve this problem, we devised and explored, for the first time to our knowledge, a virtual-detector-based synthetic-aperture focusing technique, combined with coherence weighting, for photoacoustic microscopy with such a large-NA transducer. Images of phantoms show that the proposed technique improves the lateral resolution from and increases the signal-to-noise ratio by up to , depending on the distance from the ultrasonic focal point. In vivo experiments show that the technique also provides a clearer representation of the vascular distribution in the rat’s scalp.
© 2006 Optical Society of America
Full Article | PDF ArticleMore Like This
De Cai, Zhongfei Li, Yao Li, Zhendong Guo, and Sung-Liang Chen
Opt. Express 25(2) 1421-1434 (2017)
Hao F. Zhang, Konstantin Maslov, Meng-Lin Li, George Stoica, and Lihong V. Wang
Opt. Express 14(20) 9317-9323 (2006)
Konstantin Maslov, Gheorghe Stoica, and Lihong V. Wang
Opt. Lett. 30(6) 625-627 (2005)