Abstract
A Fabry–Perot ultrasound sensor with nonhygroscopic dielectric mirrors made out of and for use in photoacoustic tomography is described. The sensor offers flat frequency response up to 36 MHz, low noise-equivalent pressure (70 Pa), and near-omnidirectional response up to 20 MHz as well as optical transparency for near-infrared illumination. A numerical model was developed to predict its frequency response, and the results were validated experimentally. An image of the human palm was acquired to demonstrate in vivo imaging capabilities.
© 2017 Optical Society of America