Abstract
Purpose
Intracranial aneurysm (IA) wall enhancement on post-contrast vessel wall magnetic resonance imaging (VW-MRI) is assumed to be a biomarker for vessel wall inflammation and aneurysm instability. However, the exact factors contributing to enhancement are not yet clarified. This study investigates the relationship between luminal enhancement and intra-aneurysmal flow behaviour to assess the suitability of VW-MRI as a surrogate method for determining quantitative and qualitative flow behaviour in the aneurysm sac.
Methods
VW-MRI signal is measured in the lumen of three patient-specific IA flow models and compared with the intra-aneurysmal flow fields obtained using phase-contrast magnetic resonance imaging (PC-MRI) and computational fluid dynamics (CFD). The IA flow models were supplied with two different time-varying flow regimes.
Results
Overall, the velocity fields acquired using PC-MRI or CFD were in good agreement with the VW-MRI enhancement patterns. Generally, the regions with slow-flowing blood show higher VW-MRI signal intensities, whereas high flow leads to a suppression of the signal. For all aneurysm models, a signal value above three was associated with velocity values below three cm/s.
Conclusion
Regions with lower enhancements have been correlated with the slow and high flow at the same time. Thus, further factors like flow complexity and stability can contribute to flow suppression in addition to the flow magnitude. Nevertheless, VW-MRI can qualitatively assess intra-aneurysmal flow phenomena and estimate the velocity range present in the corresponding region.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The prevalence of intracranial aneurysms (IA) in the adult population is estimated to be around 3% [1, 2]. Although IAs are generally associated with a relatively low risk of rupture, haemorrhage following the rupture of aneurysms is fatal in most cases. Therefore, incidental unruptured aneurysm IAs cause significant concerns [3]. Simultaneously, the treatment options available are associated with a variety of severe complications. Thus, a patient-specific IA rupture risk assessment is inevitable.
Previous studies have identified numerous factors that are associated and correlated with a high risk of rupture such as morphological parameters (e.g., size and shape) [4, 5] and flow characteristics [6].
Intracranial blood flow covers a large variety of different flow conditions (slow, stagnant or transitional [7, 8]). Especially, low flow conditions near the aneurysm’s wall have been associated with rupture sites [9]. Moreover, concentrated inflow jets and small impingement regions have been found to instabilize aneurysms [10]. Other flow-related parameters, such as the low shear area (LSA), the oscillatory shear index (OSI) and the neck inflow rate (NIR), contribute to the growth and rupture risk of the aneurysm [9, 11, 12].
Wall inflammation is a primary factor contributing to instable vessel wall constitution and hence instability of aneurysm and change of aneurysm shapes [13,14,15,16,17], but until now, wall inflammation is difficult to assess. However, several studies have suggested using wall enhancement in contrast-enhanced vessel wall magnetic resonance imaging (VW-MRI) as an indirect marker for wall inflammation [18,19,20]. Exemplarily, different flow conditions (e.g., at the neck, aneurysm body or dome) were associated with a varying degree of local enhancement [21]. Furthermore, VW-MRI in combination with flow MRI was used to correlate morphology and haemodynamics in IAs [22].
Although it is well known that VW-MRI is based on suppressing intraluminal signal, generating images with black blood and bright surrounding tissues, the suppression depends on local flow conditions and works best for higher flow regions. Therefore, inflammation and slow flow can contribute to enhanced luminal regions on VW-MRI [23].
Recent studies have discussed the interaction of slow-flowing blood and enhanced wall regions [3, 24,25,26]. The slow flow at inner boundaries could lead to an overestimation of the aneurysm wall enhancement. Alternatively, the signal originating from unsuppressed blood may be mistaken for the aneurysm wall [24, 25]. This complicates the distinction between the effect and extent of slow flow and wall inflammation on the enhancement in in vivo studies. However, as discussed above, slow flow itself might be a valuable marker for aneurysm instability. Flow in aneurysms can be measured non-invasively using phase-contrast magnetic resonance imaging (PC-MRI). However, the widespread of this technique in clinical facilities is limited by required long examination times. Therefore, we hypothesize that the ability to visualize slow flow by VW-MRI can be a useful feature rather than disadvantage.
Recently, the effect of slow-flowing blood on VW-MRI data alone was investigated in vitro using phantom models [24, 27]. However, the exact interaction of flow constitution and blood signal intensity and a possible quantitative correlation remains unclear. Therefore, this study investigates the relationship between flow characteristics and luminal enhancement in VW-MRI measurements by performing an advanced flow analysis. We aim to assess the suitability of VW-MRI as a surrogate method for determining quantitative and qualitative flow behaviour by correlating VW-MRI signal void and flow constitution. The focus is set on the possibility to extract features from VW-MRI that complement or even substitutes information from PC-MRI or numerical flow data. PC-MRI and VW-MRI images have been acquired in phantom models of three patient-specific intracranial aneurysms under varying inflow conditions. Detailed haemodynamic flow simulations based on computational fluid dynamics (CFD) of the respected patient models complement the data and allow for a more detailed analysis.
Materials and methods
Case selection
Since IAs show a large variety with respect to size and shape, the intraluminal flow patterns can drastically differ (e.g., stable versus complex haemodynamics). Therefore, three representative IA patients with different sizes and locations as well as in vivo VW-MRI enhancement were identified. Due to these differences a large range of flow regimes could be covered and investigated both experimentally and numerically. This enables a more generalized identification of the correlation between signal intensity and velocity values. The patients underwent 3D rotational angiography (3D RA, Allura XperFD 20/10, Philips, Best, the Netherlands) with a spatial resolution of 0.27 mm3 as part of the clinical care.
Afterwards, three aneurysm flow models (Table 1, Fig. 1) were designed and produced in-house from rigid material (Form 3, Formlabs, USA). First, the vessel lumen was segmented from patient 3D RA. Second, it was simplified, e.g., by cutting branches smaller than 1 mm. Finally, the outer wall (3 mm thick) was added and models were equipped with flow connectors [28]. For further details the interested reader is referred to [27].
MRI in vitro measurement set-up
A glycerol–water mixture (40:60) with 0.3 mmol/l of contrast agent (Gadovist, Bayer Vital, Germany) was pumped (Ismaltec MCP Standart, Cole Parmer, USA) through the models. The pump generated two different pulsating inflow profiles (high and low flow conditions) with an averaged flow rate of 160 and 266 ml/min for M1; 203 and 307 ml/min for M2; 208 and 311 ml/min for M3, respectively. Time-dependent flow and pressure were measured at the outlets of the models using a clamp-on transonic flow sensor (ME8PXL-M12, Transonic System Inc., USA) and a Luer-lock pressure sensor (PRESS-N-000, PendoTech, USA).
The in vitro MRI measurements were taken on a 3 T MR system (Ingenia CX, R5 V6.1, Philips Healthcare, Best, the Netherlands). The MR protocol comprises VW-MRI and PC-MRI (Table 1). PC-MRI acquired time-resolved three-dimensional and three-directional velocity distribution using a four-point phase-contrast encoding scheme [29].
Further details regarding the measurements and the underlying concept of black blood MRI can be found in [27].
Haemodynamic flow simulations
High-resolution image-based blood flow simulations were carried out to support the given PC-MRI data in information and resolution. The inner vascular lumens segmented from the patient's 3D RA were used to extract the simulation geometries. The extrusion length of in- and outlet cross sections ranged between 7 and 24 cm to match the distance between the aneurysm model and the flow or pressure sensors, respectively, providing realistic boundary conditions for the CFD simulations. The numerical meshes consisted of polyhedral and prismatic cells and had a base size of 0.1 mm. Spatial discretization was carried out using the finite volume solver STAR-CCM + 14.04 (Siemens Product Lifecycle Management Software Inc., Plato, TX, USA) [30]. The base size was enlarged for the extruded parts of the geometry to reduce the total number of required cells. In total, the resulting meshes consisted of approximately 9.8 million (model M1), 15.1 million (model M2) and 8.5 million (model M3) cells, respectively, guaranteeing a mesh-independent solution. The time-dependent flow rates measured using the transonic flow sensors were applied to the corresponding inlet cross-sections. In addition, the curves provided by the pressure sensors defined the outlet boundary conditions. All vessel walls were assumed to be rigid and no-slip boundary conditions were applied. Fluid properties were considered incompressible (ρ = 1114.5 kg/m3), and Newtonian (η = 3.72 mPas) and laminar flow conditions were assumed, which matches the experimental settings. Each of the six time-dependent blood flow simulations comprises three cardiac cycles (time step size Δt = 0.001 s) to obtain a periodic solution and only the last one was included in the analysis. Consistent with the PC-MRI measurements, 24 evenly distributed time points throughout the cardiac cycle were exported and averaged (Table 2).
Data processing and flow evaluation
To ensure quantitative comparison between the three modalities (VW-MRI, PC-MRI and CFD), several processing steps are needed. First, velocity, which is encoded in the raw phase difference PC-MRI data, was calculated (GTflow, Version 3.1.12, Gyrotools, Switzerland), considering eddy-current correction and phase-aliasing [31]. Second, the VWI signal intensity values of the vessel lumen data were normalized by the signal intensity of the stationary tissue. Third, the CFD results were manually co-registered with both imaging datasets for all three patient-specific aneurysm models. Finally, each aneurysm sac was virtually separated from the surrounding vessels. The mentioned processing steps as well as the qualitative evaluation were carried out in EnSight v10.2 (ANSYS Inc., Canonsburg, PA, USA).
The subsequent quantitative evaluation was performed in MATLAB R2020a (MathWork, USA) and was based on the voxel resolution of the PC-MRI data. Therefore, VW-MRI and CFD data were downsampled to 1 mm3. Furthermore, all voxels closer than 1 mm to the vessel wall, which is the acquired voxel size of PC-MRI data, were discarded in the quantitative analyses. Hence, the influence of partial volume effects near the vessel wall was minimized.
Cluster generation
To assess the relationship between signal intensity in the VW images and the measured velocity in each case, the signal values were clustered into ten groups. First, the VW-MRI signal values were combined into ten equidistant regions from the lowest to the highest measured signal value. Second, for each VW-MRI cluster the corresponding PC-MRI values were assigned. Finally, the average signal intensity and velocity calculated for each cluster were compared (Fig. 5). Note that the number of signal values is not equally distributed over the whole range (e.g., mostly low signal values exist in model M1). Therefore, the number of values belonging to one cluster differs within one model (see also Fig. 6).
Assessment of velocity fluctuations
The oscillatory velocity index (OVI) can quantify flow pattern variations and flow complexity during the cardiac cycle (OVI ≈ 0 when the velocity field direction remains steady during the cardiac cycle, OVI ≈ 0.5 for high temporal changes and strongly varying flow patterns) [32,33,34]
where fvi is the velocity vector comprising of the components and T is the duration of one cardiac cycle.
Results
The consideration of time- and magnitude-varying inflow profiles, as well as the use of differently sized and shaped intracranial aneurysms, located at the different arteries, ensures the coverage of a broad range of flow regimes in this study.
Luminal enhancement
VW-MRI reveals different levels and patterns of luminal enhancement among the models, which is qualitatively illustrated in Fig. 2. Model M1 (basilar tip aneurysm) experiences the highest suppression over the entire aneurysmal lumen. During the high flow condition, only a small non-suppressed zone near the aneurysm neck remains. Slightly higher signal intensities appear for the lower flow conditions in the same model. Here, inhomogeneous luminal enhancement appears in the centre of the aneurysm. Model M2 (internal carotid artery aneurysm) presents complex enhancement and suppression patterns. Signal void zones separate weak and strong enhanced regions for both flow conditions. Model M3 (MCA bifurcation aneurysm) possesses the highest enhancement. Nearly the complete aneurysm is enhanced during low flow conditions. With the higher flow rate, a regional suppression is observed near the aneurysm neck that appears to be formed like an inflow jet. The luminal enhancement's general characteristics were similar within the model supplied with the different flows, but changed drastically among different models. This can be attributed to the difference in size and shape. Moreover, the actual flow rate entering the aneurysm is highly different due to the presence of the side branches at models 2 and 3 and the structure of the parent vessels.
Flow characteristics
Overall, Fig. 3 shows a qualitatively good agreement in the flow behaviour between PC-MRI measurements and CFD simulations. In accordance with the VW-MRI results (Fig. 2), the general flow characteristics are similar within one model for low and high flow conditions. As indicated from the VW-MRI measurement, the intra-aneurysmal flow differs in magnitude and complexity among the models M1 – 3. The flow entering the aneurysm was different for each model due to varying flow rates, numbers of side branches and parent vessel structures. As calculated based on haemodynamic simulations, the aneurysm in model M1 is fed with the highest flow (low flow average 158 ml/min; high flow average 268 ml/min). This correlates with the strong suppression of blood flow observed in Fig. 2. Moreover, the inhomogeneous enhancement in the aneurysm sac and near the aneurysm neck can be found in the regions, where the lowest intra-aneurysmal flow is observed. Model M2 experiences an average flow rate entering the aneurysm of 56 ml/min for low flow conditions and 77 ml/min for high flow conditions as calculated based on haemodynamic simulations. The inflow jet position correlates with the lower enhanced regions for low and high flow conditions. Nevertheless, the origin of the signal void zones between enhanced regions remains unclear and requests a more detailed investigation of the intra-aneurysmal flow behaviour.
In accordance with the high enhancement seen for model M3 in the VW-MRI (recall Fig. 2), only a small fraction of the parental flow enters the aneurysm (0.1 and 0.2 ml/min). Nevertheless, the lower enhancement observed under the high flow conditions correlates with the position of the aneurysm jet (Fig. 3).
The streamlines illustrated in Fig. 4 indicate the flow complexity inside the aneurysmal lumens. Models M1 and M3 possess one recirculation zone (orange arrow), whereas the flow in model M2 seems to have a higher complexity. Here, flow division is observed at the impingement zone in the aneurysm sac (green arrow) and two recirculation zones (orange triangles) can be observed.
The OVI indicates the flow complexity throughout the cardiac cycle (Fig. 5). All aneurysm models show the highest temporal variation in the centre of the lumen. The isosurfaces represent an OVI value above 0.1. The OVI distribution is more heterogeneous for model M1 and M2 (low flow conditions), but confined for model M3 and M2 (high flow conditions). No direct colocalization between signal suppression zones (Fig. 1) and high temporal complexity can be observed.
Quantitative relation between intra-aneurysmal velocity and signal intensity
To assess the relationship between signal intensity in the VW images and the measured velocity, signal and velocity values in ten different clusters (see 2.4) are evaluated. Figure 6 indicates the velocity distribution within the clusters for the representative model M2. Low flow can be observed in all clusters, thus in all signal regimes, whereas high flow can only be found in regions correlating with low signal. Therefore, low signal intensity areas do not necessarily correspond to high flow and low flow does not necessarily correspond to a high signal. The effect gets smaller with increasing signal intensity or higher flow, respectively. Similar observations can be found for the other models and flow regimes.
Figure 7 illustrates mean velocity measured using PC-MRI plotted over mean signal intensity measured for all models within the generated clusters. For each model, a similar dependence between velocity and signal enhancement can be found. Clusters with lower average signal intensities are associated with higher average velocity values. In the signal range between three and six, the average velocity for all models is in the same range (3–0 cm/s) and linearly decreases, although for model M1 the enhancement is never above four. In the signal range between zero and three, the average velocity decreases exponentially. Nevertheless, this exponential value is highly different among the models.
Discussion
This study quantifies the relationship between luminal signal intensity based on VWI and the velocity distribution inside intracranial aneurysms. Three representative patient-specific IAs that differ in size and shape, as well as in intraluminal flow magnitude, have been evaluated. Qualitatively, different enhancement patterns as well as flow fields have been observed. The highest signal intensity and lowest inflow rate occur in model M3, while the lowest enhancement and largest inflow rate were found in model M1. The observed flow phenomena, detected in good agreement using PC-MRI or CFD, match with VW signal intensity. Inflow jets and low flow regions colocalize with low and high intensity zones. Furthermore, complex enhancement patterns, including signal void zones in between enhanced regions, indicate a higher complexity of the intra-aneurysmal flow (model M2). These findings support the hypothesis that behaviour and characteristics of the intra-aneurysmal flow highly influence the flow suppression and consequently signal intensity observed by the VWI measurements.
The qualitative results are supported by a quantitative evaluation of the relationship between signal intensity and measured velocity values. Although the overall trend indicates that higher signal intensity is generally associated with lower intra-aneurysmal flow, low signal intensity can also be found in regions corresponding to all other flow regimes. More specifically, luminal flow enhancement can be almost exclusively found for slow flow areas, but the effect is not very distinct in signal-suppressed regions.
The nature of the blood flow within IAs can be complex and highly patient-specific. Thus, phenomena contributing to flow suppression can occur in parallel. Based on the results of the associated study, it is assumed that the suppression can not only be attributed to high flow, but further factors associated with complex spatial behaviour have an additional impact. Nevertheless, no direct conclusion can be drawn from the temporal complexity (e.g., based on the evaluation of OVI) to regions that face a signal suppression during VW-MRI.
Previous studies have discussed the role of signal intensity and argued about fact or artefact [3, 19, 25, 26, 35]. This study encourages using the enhancement patterns as a feature and potential biomarker to identify pathological changes in the luminal flow behaviour. Due to the high complexity of the aneurysmal flow, VWI cannot completely replace more sophisticated measurement methods like PC-MRI on a quantitative basis. Nevertheless, it can give an estimate of the overall flow range and magnitude measured. Moreover, the enhancement patterns can be used for the characterization of flow phenomena and comparison of flow in patients with multiple aneurysms. Advantages of VW-MRI in comparison to PC-MRI are the smaller measurement times and reduced number of associated post-processing steps. Nevertheless, the potential relationship between luminal enhancement and IA rupture requires further in-depth investigation. However, this study contributes to an improved understanding of previous observations and occurring flow-related phenomena.
Apart from the findings, it is important to emphasize that this study has several limitations: Due to the complex manufacturing and measurement process, the number of models was limited to three. Furthermore, the high difference in size and shape of the models has ensured the investigation of a broad range of flow regimes. Nevertheless, this leads to a lack of information about the behaviour in aneurysms covering a similar flow behaviour. Additionally, the use of a blood-mimicking fluid could lead to different signal intensities in comparison to real blood measurement. Finally, no information about wall enhancement could be given because of the in vitro character of the study.
Future studies will include the VW-MRI results within similar aneurysm types to ensure comparability and representativeness of the results, including in vivo investigations. Furthermore, a strong focus will set on the detailed identification and characterization of additional factors that contribute to flow suppression or signal enhancement.
Conclusions
This study has demonstrated the influence of intra-aneurysmal flow behaviour on VW-MRI signal intensity. It supports the use of intra-aneurysmal flow-related signal enhancement as a feature to characterize the overall flow behaviour. VW-MRI can help to qualitatively assess flow phenomena, including inflow jets, recirculation zones and different flow regimes. It is well suited to compare the flow regimes in different aneurysms. On a qualitative basis, VW-MRI can be an alternative to complex PC-MRI measurements or support spatially lower-resolved measurements in detail and information. The investigation of signal intensity patterns can help to identify pathological changes in the aneurysmal lumen.
References
Wiebers DO (2003) Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. The Lancet 362(9378):103–110
Etminan N, Rinkel GJ (2016) Unruptured intracranial aneurysms: development, rupture and preventive management. Nat Rev Neurol 12(12):699
Cebral JR, Detmer F, Chung BJ, Choque-Velasquez J, Rezai B, Lehto H, Tulamo R, Hernesniemi J, Niemela M, Yu A (2019) Local hemodynamic conditions associated with focal changes in the intracranial aneurysm wall. Am J Neuroradiol 40(3):510–516
Leemans EL, Cornelissen BMW, Said M, Rvd B, Slump CH, Marquering HA, Majoie CBLM (2019) Intracranial aneurysm growth: consistency of morphological changes. Neurosurg Focus FOC 47(1):E5. https://doi.org/10.3171/2019.4.FOCUS1987
Leemans EL, Cornelissen BMW, Slump CH, Majoie CBLM, Cebral JR, Marquering HA (2019) Comparing morphology and hemodynamics of stable-versus-growing and grown intracranial aneurysms. Am J Neuroradiol 40(12):2102–2110. https://doi.org/10.3174/ajnr.A6307
Detmer FJ, Chung BJ, Jimenez C, Hamzei-Sichani F, Kallmes D, Putman C, Cebral JR (2019) Associations of hemodynamics, morphology, and patient characteristics with aneurysm rupture stratified by aneurysm location. Neuroradiology 61(3):275–284
Balasso A, Fritzsche M, Liepsch D, Prothmann S, Kirschke JS, Sindeev S, Frolov S, Friedrich B (2019) High-frequency wall vibrations in a cerebral patient specific aneurysm model. Biomed Eng Biomedizinische Technik 64(3):275–284
Xu L, Saito A, Yokoyama Y, Sato K, Sasaki T, Yamaguchi R, Sugawara M, Liu H (2016) Low-frequency harmonics in inlet flow rate play a crucial role in inducing flow instabilities in terminal cerebral aneurysms. J Biomech Sci Eng 11 (3):16–00117–00116–00117
Liang L, Steinman DA, Brina O, Chnafa C, Cancelliere NM, Pereira VM (2019) Towards the clinical utility of CFD for assessment of intracranial aneurysm rupture – a systematic review and novel parameter-ranking tool. J NeuroInterventional Surg 11(2):153–158. https://doi.org/10.1136/neurintsurg-2018-014246
Cebral JR, Mut F, Weir J, Putman CM (2011) Association of hemodynamic characteristics and cerebral aneurysm rupture. Am J Neuroradiol 32(2):264–270
Chung BJ, Mut F, Putman CM, Hamzei-Sichani F, Brinjikji W, Kallmes D, Jimenez CM, Cebral JR (2018) Identification of hostile hemodynamics and geometries of cerebral aneurysms: a case-control study. Am J Neuroradiol 39(10):1860–1866
Berg P, Beuing O (2018) Multiple intracranial aneurysms: a direct hemodynamic comparison between ruptured and unruptured vessel malformations. Int J Comput Assist Radiol Surg 13(1):83–93
Tulamo R, Frösen J, Hernesniemi J, Niemelä M (2018) Inflammatory changes in the aneurysm wall: a review. J Neurointerventional Surg 10(Suppl 1):i58–i67
Turkmani AH, Edwards NJ, Chen PR (2015) The role of inflammation in cerebral aneurysms. Neuroimmunol Neuroinflamm 2:102–106
Frösen J, Piippo A, Paetau A, Kangasniemi M, Niemelä M, Hernesniemi J, Jääskeläinen J (2006) Growth factor receptor expression and remodeling of saccular cerebral artery aneurysm walls: implications for biological therapy preventing rupture. Neurosurgery 58(3):534–541
Frösen J, Piippo A, Paetau A, Kangasniemi M, Niemelä M, Hernesniemi J, Jääskeläinen J (2004) Remodeling of saccular cerebral artery aneurysm wall is associated with rupture: histological analysis of 24 unruptured and 42 ruptured cases. Stroke 35(10):2287–2293
Larsen N, Von Der Brelie C, Trick D, Riedel C, Lindner T, Madjidyar J, Jansen O, Synowitz M, Flüh C (2018) Vessel wall enhancement in unruptured intracranial aneurysms: an indicator for higher risk of rupture? High-resolution MR imaging and correlated histologic findings. Am J Neuroradiol 39(9):1617–1621
Edjlali M, Guédon A, Hassen WB, Boulouis G, Benzakoun J, Rodriguez-Régent C, Trystram D, Nataf F, Meder J-F, Turski P, Oppenheim C, Naggara O (2018) Circumferential thick enhancement at vessel wall MRI Has high specificity for intracranial aneurysm instability. Radiology 289(1):181–187. https://doi.org/10.1148/radiol.2018172879
Larsen N, Flüh C, Saalfeld S, Voß S, Hille G, Trick D, Wodarg F, Synowitz M, Jansen O, Berg P (2020) Multimodal validation of focal enhancement in intracranial aneurysms as a surrogate marker for aneurysm instability. Neuroradiology 62(12):1627–1635. https://doi.org/10.1007/s00234-020-02498-6
Zwarzany Ł, Tyburski E, Poncyljusz W (2021) High-resolution vessel wall magnetic resonance imaging of small unruptured intracranial aneurysms. J Clin Med 10(2):225
Hadad S, Mut F, Chung B, Roa J, Robertson A, Hasan D, Samaniego E, Cebral J (2020) Regional Aneurysm Wall Enhancement is Affected by Local Hemodynamics: A 7T MRI Study. American Journal of Neuroradiology
Zhang M, Peng F, Li Y, He L, Liu A, R. L, (2021) Associations between morphology and hemodynamics of intracranial aneurysms based on 4D flow and black-blood magnetic resonance imaging. Quant Imaging Med Surg 11(2):597–607. https://doi.org/10.21037/qims-20-440
Hu P, Yang Q, Wang D-D, Guan S-C, Zhang H-Q (2016) Wall enhancement on high-resolution magnetic resonance imaging may predict an unsteady state of an intracranial saccular aneurysm. Neuroradiology 58(10):979–985
Cornelissen BM, Leemans EL, Coolen BF, Peper ES, van den Berg R, Marquering HA, Slump CH, Majoie CB (2019) Insufficient slow-flow suppression mimicking aneurysm wall enhancement in magnetic resonance vessel wall imaging: a phantom study. Neurosurg Focus 47(1):E19
Cornelissen BM, Leemans EL, Slump CH, Marquering HA, Majoie CB, van den Berg R (2019) Vessel wall enhancement of intracranial aneurysms: fact or artifact? Neurosurg Focus 47(1):E18
van den Berg R (2020) Intracranial aneurysm wall enhancement: fact or fiction? Neuroradiology 62:269–270. https://doi.org/10.1007/s00234-020-02367-2
Pravdivtseva M, Gaidzik F, Berg P, Hoffman C, Rivera-Rivera L, Medero R, Bodart L, Roldan-Alzate A, Speidel M, Johnson K, Wieben O, Jansen O, Hövener J-B, Larsen N (2020) Pseudo-Enhancement in Intracranial Aneurysms on Black Blood MRI: Effects of flow rate, spatial resolution, and additional flow suppression. J Magn Reson Imaging 54:888–901. https://doi.org/10.1002/jmri.27587
Pravdivtseva M, Peschke E, Lindner T, Wodarg F, Hensler J, Gabbert D, Voges I, Berg P, Barker AJ, Jansen O, Hövener J-B (2020) 3D-printed, patient-specific intracranial aneurysm models: from clinical data to flow experiments with endovascular devices. Med Phys 48:1469–1484. https://doi.org/10.1002/mp.14714
Pelc NJ, Bernstein MA, Shimakawa A, Glover GH (1991) Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging 1(4):405–413
Berg P, Saalfeld S, Voß S, Beuing O, Janiga G (2019) A review on the reliability of hemodynamic modeling in intracranial aneurysms: why computational fluid dynamics alone cannot solve the equation. Neurosurg Focus 47(1):E15
Lankhaar J-W, Hofman MBM, Marcus JT, Zwanenburg JJM, Faes TJC, Vonk-Noordegraaf A (2005) Correction of phase offset errors in main pulmonary artery flow quantification. J Magn Reson Imaging 22(1):73–79
Bouillot P, Brina O, Ouared R, Lovblad K-O, Farhat M, Pereira VM (2014) Particle imaging velocimetry evaluation of intracranial stents in sidewall aneurysm: hemodynamic transition related to the stent design. PLoS One 9(12):e113762
Sano T, Ishida F, Tsuji M, Furukawa K, Shimosaka S, Suzuki H (2017) Hemodynamic differences between ruptured and unruptured cerebral aneurysms simultaneously existing in the same location 2 case reports and proposal of a novel parameter oscillatory velocity index. World Neurosurg 98:868. e865–868. e810
Tanioka S, Ishida F, Kishimoto T, Tsuji M, Tanaka K, Shimosaka S, Toyoda M, Kashiwagi N, Sano T, Suzuki H (2019) Quantification of hemodynamic irregularity using oscillatory velocity index in the associations with the rupture status of cerebral aneurysms. J Neurointerventional Surg 11(6):614–617
Bart MWC, Eva LL, Cornelis HS, Henk AM, Charles BLMM, van den René B (2019) Vessel wall enhancement of intracranial aneurysms: fact or artifact? Neurosurgical Focus FOC 47(1):E18. https://doi.org/10.3171/2019.4.FOCUS19236
Funding
Open Access funding enabled and organized by Projekt DEAL. This study was funded by the Federal Ministry of Education and Research in Germany within the Forschungscampus STIMULATE (grant number 13GW0473A), the German Research Foundation (grant number BE 6230/2–1) and the international Graduate School MEMoRIAL at University of Magdeburg, Germany, kindly supported by the European Structural and Investment Funds (ESF, project number ZS/2016/08/80646) and the Research Training Group "Materials4Brain" (GRK2154; P2). Further support by the cluster of excellence precision medicine in inflammation PMI 1267 and the faculty of medicine for funding the core facility MOIN CC is acknowledged.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Availability of data and material
Processed data and code is available upon request from the corresponding author.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gaidzik, F., Pravdivtseva, M., Larsen, N. et al. Luminal enhancement in intracranial aneurysms: fact or feature?—A quantitative multimodal flow analysis. Int J CARS 16, 1999–2008 (2021). https://doi.org/10.1007/s11548-021-02486-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-021-02486-y