Abstract
Purpose
The introduction of novel endoscopic instruments is essential to reduce trauma in visceral surgery. However, endoscopic device development is hampered by challenges in respecting the dimensional restrictions, due to the narrow access route, and by achieving adequate force transmission. As the overall goal of our research is the development of a patient adaptable, endoscopic anastomosis manipulator, biomechanical and size-related characterization of gastrointestinal organs are needed to determine technical requirements and thresholds to define functional design and load-compatible dimensioning of devices.
Methods
We built an experimental setup to measure colon tissue compression piercing forces. We tested 54 parameter sets, including variations of three tissue fixation configurations, three piercing body configurations (four, eight, twelve spikes) and insertion trajectories of constant velocities (5 mms−1, 10 mms−1,15 mms−1) and constant accelerations (5 mms−2, 10 mms−2, 15 mms−2) each in 5 samples. Furthermore, anatomical parameters (lumen diameter, tissue thickness) were recorded.
Results
There was no statistically significant difference in insertion forces neither between the trajectory groups, nor for variation of tissue fixation configurations. However, we observed a statistically significant increase in insertion forces for increasing number of spikes. The maximum mean peak forces for four, eight and twelve spikes were 6.4 ± 1.5 N, 13.6 ± 1.4 N and 21.7 ± 5.8 N, respectively. The 5th percentile of specimen lumen diameters and pierced tissue thickness were 24.1 mm and 2.8 mm, and the 95th percentiles 40.1 mm and 4.8 mm, respectively.
Conclusion
The setup enabled reliable biomechanical characterization of colon material, on the base of which design specifications for an endoscopic anastomosis device were derived. The axial implant closure unit must enable axial force transmission of at least 28 N (22 ± 6 N). Implant and applicator diameters must cover a range between 24 and 40 mm, and the implant gap, compressing anastomosed tissue, between 2 and 5 mm.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Since the last 15 years, current research efforts focus on enabling endoscopic interventions not only within the access lumen itself, but also on other organs of the abdominal cavity. Particularly interesting are interventions of daily clinical routine, such as the reconnection of bowel endings after colorectal resection. The indications are manifold (colorectal cancer, inflammatory disease, etc.) and it can be assumed that more than 1 million individuals have to undergo colorectal resection every year [1, 2]. Currently, for wide parts of the colon, anastomosis creation is the procedure step, which limits a further reduction of trauma the most. To achieve a smooth healing process, a direct serosa apposition of inverting end-to-end anastomoses is deemed necessary [3].
To securely connect two bowel endings and to prevent slippage, we are currently developing a compression-based implant which uses piercing of the bowel in the compression zone. However, endoscopic device development is hampered by the restrictions of limited space available and high functional demands. Force transmission to the endoscope tip is hereby a particular challenge. Accordingly, biomechanical characterization of gastrointestinal organs is needed to determine technical requirements and thresholds that define the functional design and load-compatible dimensioning. This also includes the assessment of size-related anatomical properties for dimensioning scalability with respect to lumen diameter and bowel wall thickness. Focus of the present study was to develop a system to reliably assess tissue penetration forces and the investigation of geometrical characteristics. By application of this innovative test rig, we were able to examine various implant- and applicator-related design specifications and their impact on puncturing forces.
State of the art
The mechanical characterization of biological tissue is a central focus of various research studies. Potential applications include the development of force feedback applications or VR-based surgical simulators, the invention of new instruments or the input to in silico approaches for the investigation of surgical procedures, devices and therapeutic outcome prognosis [4]. Biomechanical computation holds the great potential of flexibility, time efficiency and the reduction, refinement and replacement of animal experiments in biomedical research [4].
Common experimental setups to derive the relevant parameters include uni- and biaxial tensile or indentation tests. Hereby, diverse research groups focus on the biomechanical characterization of porcine or human gastrointestinal tissues.
As we are interested in tissue puncture tests, related experimental setups will be primarily examined in the following. Typical research foci include electrode implantation into the brain or investigations of mechanical tissue interaction during single-body insertion.
Heijnsdijk et al. measured the perforation forces for porcine large bowel, pinching the tissue between two opposingly arranged metal hemispheres with a diameter of 1.5 mm. They determined a mean force of 13.5 ± 3.7 N [5]. Research groups around Abolhassani, Okamura, Bao, Jiang, Frick and Butz examined single-needle insertion forces for different tissue types and process relevant parameters affecting the measurement values [6,7,8,9,10,11,12]. Typical setups consisted of 1 or 2 DOF testing machines to control insertion, and 6 DOF electronic force transducers [6, 7, 9, 10]. Factors to be considered during puncture experiments include the type of tissue [9, 10], tissue fixation [13] and pre-conditioning [11, 12], needle geometry (body thickness/length, tip geometry) [7, 9, 10], puncturing device insertion trajectory (comprising velocity, acceleration and rotation) [6, 9, 10] and interrupted or continuous insertion process [10].
Kwon et al. [14] developed the BMPM system (Biomaterial property measurement system) to enable measurement of deformation and puncture forces of esophageal, colonic and gastric tissues in vivo. However, experiments described were performed in vitro on porcine tissue. For piercing colon with a needle, a peak force of 0.056 N was measured.
For anastomotic device-tissue interaction, Schell came closest to answering our question with his investigations. In 2005, he developed a semicircular staple suture device for transanal endoscopic microsurgery [15]. For this purpose, the forces occurring during stapling with 17 clamps instead of 24 were measured. The force ratio from handle to staple magazine was specified to be 9.3. A force of 202 N was obtained at the lever and 1885 N at the stapler head. However, there was no specification of the experimental setup given. As the other groups’ experiments comprised only investigations with one single puncturing body and as we need to provide more penetration points for safe and non-slip positioning of the tissue in the anastomosis zone, the results are not sufficient for the design and dimensioning of an endoluminal applicator.
Therefore, the purpose of the present study was to investigate various implant- and applicator-related design specifications concerning tissue fixation to the applicator and tissue puncturing by the implant. We investigated whether different fixation point arrangements, amounts of tips, as well as varying insertion speeds and accelerations of the implant influence insertion force and process.
For our experiments, we chose porcine tissue, as the physiology and anatomy (including length) of the pig’s gastrointestinal tract, as well as its digestive function, blood flow characteristics and tissue mechanics have a very close resemblance to those of an adult human [16, 17]. Heijnsdijk et al. [5] compared perforation forces of human and pig small bowel and didn’t detect any difference. Kararli assessed the gastrointestinal anatomy, physiology and biochemistry of humans in comparison with commonly used laboratory animals and suggested to use porcine tissue for colon-related experiments [18]. Variations of anatomy are present with respect to the configuration of the colon. Pigs have a short transverse and a descending colon, but no sigmoid flexure. Furthermore, the cecum and the colon ascendens are coiled together and loops adhere to each other in the spiral colon. However, these differences in anatomy didn’t have any influence on our methods as we performed our experiments with explanted samples. Due to the high degree of resemblance, we decided to define the design specification for our system, based on the experimental results, without any further transformation.
Material and methods
Our experimental setup consisted of two identical tissue fixation entities (applicator dummies) mounted on a centering rail (Fig. 1a-1) facing each other in 4 mm distance, a disc with segmental, symmetrically arranged spikes (implant dummy) (Fig. 1a-2), and a Sauter, FH 100, calibrated force gauge (Fig. 1a-3). The FH 100 was fixed onto a slide (Fig. 1a-4) movable translationally (1DOF) along a guide rail (Fig. 1a-5). The movement was enabled by a wedge toothed belt tensioned onto two gears, of which one was driven by a stepper motor (Sanyo Denki, StepSyn Type 103H7126-0740) (Fig. 1a-6). An Arduino (Arduino Uno) with a motorshield (Arduino Motor shield Rev3) was used to control motor speed and acceleration, thus \(\frac{{{\text{steps}}}}{{{\text{min}}}}\). To ensure sufficient power supply, an external power source was connected and powered with 5 V. The piercing body was fixed to the force gauge (Fig. 1).
Tissue fixation
Each specimen holder featured openings through which the implant tips were passed and suction surfaces to atraumatically fix the tissue using vacuum [19]. The sample holder had a diameter of 35 mm and the total suction area of all entities was ~ 130 mm2. Connecting a vacuum pump to each of the specimen holders, with a maximum achievable negative pressure of 80 kPa with respect to atmospheric pressure, the fixation force per side was \(F_{{{\text{fixation}}}} = \Delta p*A = 10.4 \,{\text{N}}\) (\(\Delta p\):Difference between atmospheric and system pressure; A: surface of suction area). Three variations of suction surface configurations (Fig. 2) were investigated to find out whether the number and arrangement of tissue fixation points effects the penetration process/force. Vacuum configuration1 (Vac1I) featured suction surfaces exclusively in the center. For vacuum configurations2 (Vac2I), suction surfaces were split and located to both sides of the frontal spike apertures, to firmly stretch the tissue over the openings. The same effect was intended for configuration3 (Vac2A), for which suction surfaces were located toward the lumen center at frontal face and on the sheath.
Spikes
The implants, with a diameter of 26.5 mm, comprised four equidistant segments arranged symmetrically on a plate claiming each 36° of the needle ring. Along these fixed segments, one to three tips were applied, resulting in configurations with four, eight and twelve tips. The podiums ensured sufficient needle length for piercing two tissue layers, when advancing from the first into the second specimen holder (Fig. 3b). On these platforms, the spikes were arranged symmetrically to the center (Fig. 3a). The tip geometry was designed according to Okamura et al., who detected lowest forces for triangular/tetrahedral tip shape [7] (Fig. 3a). We used needles with a tip angle of 15°, a length of 20 mm and a diameter of 1.7 mm. The needles had to be of sufficient length to compensate for indentation that typically occurs in biological tissue due to viscoelastic properties. All piercing bodies and specimen holders were 3D-printed using Formlabs2 from standard WhiteFootnote 1 or Clear resin.Footnote 2
Trajectory
We specified trajectory velocities and accelerations in orientation to Abolhassani et al. [6]. By variation of motor velocity and acceleration, three constant axial velocities of \(v_{1} = \) 5 mms−1, \(v_{2} =\) 10 mms−1 and \(v_{3} =\) 15 mms−1 and three constant axial accelerations of \(a_{1} = \) 5 mms−2 \( a_{2} =\) 10 mms−2 a3 = 15 mms−2 were examined, for which final velocities of 12.5 mms−1, 17.7 mms−1and 21.6 mms−1 were achieved when reaching the piercing plane. The corresponding Arduino script motor velocities were 4 rpms−1, 8 rpms−1 and 12 rpms−1, and the motor accelerations 10 rpms−2, 100 rpms−2 and 150 rpms−2.
Figure 4 provides an overview over all parameter sets assessed within our study. For each parameter set, experiments with n = 5 samples were performed.
Sample preparation
Pig colons were obtained day-fresh from the local butcher, cooled and kept moist during the day until use. Fat tissue was removed (Fig. 5a) to realistically simulate bowel margins readily prepared for anastomosis closure. For the experiments, colon was cut perpendicular to the Taenia Libera into approximately 30 mm long tubular segments (Fig. 5b). For each parameter set, five sample pairs were prepared and stored in two kidney shells covered with saline (NaCl 0.9) [20] soaked bandages, numbered according to the sequence in the experimental procedure. The remaining prepared specimens were stored in a cool box.
Each specimen couple was assigned an identification code (ID) comprising: [vacuum configuration]_[number of spikes] _[velocity value/acceleration_value] _[Trajectory specification]_[sample number within parameter set] ([1I/2I/2A]__[4/8/12]_[5/10/15]_[v/a]_[1/2/3/4/5]), by which the recorded characteristics diameter and tissue thickness were correlated with the measured piercing forces. For example, sample three for Vac2I, pierced with eight tips and a constant trajectory velocity of 15 mms−1 had ID: 2I815v3.
Before the experiment, each specimen was draped flat on a base plate to measure half of the circumference, \(u_{0.5} = \frac{U}{2}\), in the center of each probe (Fig. 5c) and respective diameter was calculated by \(D = 2*\frac{{u_{0.5} }}{\pi }\), subsequently. The pierced tissue thickness of each specimen pair was assessed by means of indirect optical measurement.
Experiments
Prior to each parameter set, a script specifying the trajectory was uploaded to the Arduino board. One of the prepared bowel segments was mounted onto each of the specimen holders (Fig. 1), so that the fixation surfaces were inside the sample lumen. The open bowel ends were formed into sausage-endings and adjusted around the circumference until all suction apertures were covered. Switching on the vacuum pumps, a uniform negative pressure established on both sides. Using a MATLAB script, the compression forces occurring during tissue piercing measured with the FH 100 were recorded over time. After each trial, the specimens were replaced.
Statistical analysis
To investigate the influences of spike amount variations, different fixation point arrangements and trajectories, we statistically analyzed our results. Therefore, the Gaussian distribution of each parametric set was checked with the Shapiro–Wilk test based on a significance level of \(\alpha_{N} =\) 0.05. A parametric three-way ANOVA (A), as well as the non-parametric KRUSKAL–WALLIS (KW) test were used, whereby \(H_{0} \) hypotheses were rejected with respect to a significance level of \( \alpha_{A/KW} = 0.01667 \) for both tests.
Results
Force–time–correlation
For each experiment, occurring forces were plotted over time (Fig. 6). This correlation is qualitatively equal for all vacuum configurations, peak numbers, and trajectory specifications. As soon as tissue penetration begins, the force rises sharply (1) resulting in the peak force (2), and just as abruptly thereafter falls back (3) to a plateau (4) until force finally drops back to zero (5). A qualitative difference was observed in the plateau slope, which was not always horizontal, but sometimes showed a decrease in force over time.
Maximum and minimum forces
The mean (n = 5) and single measurement maximum and minimum peak forces for all implant configurations (four, eight, twelve spikes), over all assessed parameters, were evaluated and reported in Table 1.
Statistical analysis
Seven out of 54 parameter sets were not normally distributed. Therefore, non-parametric and parametric tests were used for the statistical analysis.
Trajectories
To identify possible differences between tested velocities and accelerations, our first hypothesis was: \(H_{0} :\) There is no difference in occurring insertion forces for groups \(v_{1}\), \(v_{2}\), \(v_{3}\), \(a_{1}\), \(a_{2}\) and \(a_{3}\). This assumption was confirmed with a p valueA of 0.513 and a p valueKW of 0.2362, thus there was no statistically significant difference between the groups (Fig. 7).
Vacuum configuration
Second hypothesis: \(H_{0} :\) There is no difference in occurring insertion forces depending on the arrangement of fixation points (different vacuum configurations used). The ANOVA test revealed a p valueA of 0.501 and the KRUSKAL–WALLIS test a p valueKW of 0.845, thus there was no statistically significant difference for the three vacuum configurations (Fig. 8).
Amount of spikes
Third hypothesis: \(H_{0} :\) There is no difference in occurring insertion forces depending on the amount of spikes (4, 8, 12 spikes). With p valuesA/KW of < 2e-16 for ANOVA and the KRUSKAL–WALLIS test, this hypothesis was rejected with respect to \(\alpha_{{\text{A/KW}}} = 0.01667.\) The insertion force measured increased significantly with an increasing number of implant spikes (Fig. 9).
Plotting the correlation between the mean value of peak insertion forces and the number of spikes, we observed a consistent coherency for all velocities, accelerations and vacuum configurations (Fig. 10).
Diameter and thickness of pierced tissue
We evaluated a mean diameter of 31.3 ± 4.6 mm (n = 540), and a mean pierced tissue thickness of 3.8 ± 0.7 mm (n = 270). The 5th and 95th percentiles, \(x_{p = 0.05}\) and \( x_{p = 0.95}\) were calculated by \(x_{p} = 0.5*\left( {x_{n*p} + x_{n*p + 1} } \right)\). The 5th percentile for the lumen diameter is \(x_{0.5}^{{{\text{diameter}}}} = 24.1\,{\text{mm}}\), and the 95th is \(x_{0,95}^{{{\text{diameter}}}} = 40.1\,{\text{mm}}\). Same procedure revealed the 5th and 95th percentile for the pierced tissue thickness to be \(x_{0.5}^{{{\text{thickness}}}} = 2.8\,{\text{ mm}}\), and \(x_{0,95}^{{{\text{thickness}}}} = 4.8\,{\text{mm}}\), respectively.
Discussion
Improving colorectal surgery is of particular interest, due to the high numbers of patients affected every year [1, 2]. To reduce trauma in clinical daily life, new instruments are of particular interest, as for example innovative anastomotic devices. However, there are few studies characterizing the mechanical tissue-system interaction, data, which would be relevant for dimensioning and design of such instruments. It was our goal, to fill this gap, give orientation values and provide general design and dimensioning guidelines for new instruments and colorectal anastomosis system development.
Force–time-correlation
Profile phases are explained under consideration of other research groups’ findings. In contrast to Abolhassan et al. [6, 7, 9], we evaluated the force–time instead of the force–displacement correlation. Needle, thus tissue displacement and the time relate linearly to each other, which is why comparison of the curves is still intuitively possible. Phase (1) corresponds to indentation of viscoelastic tissue. Stress applied to the tissue, and therefore measured force, rises with increasing piercing body displacement, respectively, time. In our profile, this correlation looks linear. Comparing the results to Bao and Okamura, we observe polynomial curve progressions, which are described by second or third-order polynomial spring model [7, 9]. However, they used massive specimens (i.e., whole liver) instead of tissue slices. In a specimen block, subjacent tissue layers strongly influence the overall stiffness, which can therefore not be considered constant over entire insertion depth. The results of Dimaio and Abolhassani, based on experiments with thin tissue samples, are more comparable to ours. Dimaio used a homogeneous linear elastostatic model to predict forces during soft tissue puncture [21] and for Abolhassani the displacement-force curve approaches a linear progression as well. However, our slope is much higher than for 1:1 depiction of indentation, as the process itself, depending on the transversal speed, only takes a few miliseconds. The peak indicates the event of puncture [6, 7, 9] (2), occurring when the tissue stress, created by the piercing bodies, exceeds a critical threshold and micro-cracks propagate [22], resulting in tissue failure and an abrupt drop of force recorded (3). While for Abolhassani, the force remains approximately zero after the drop, it rises again for Bao and Okamura [7, 9], as frictional forces occur with further penetration into specimen block. At the plateau (4), the sleigh remains in its position for 3000 ms. Variations in the plateau slope may occur due to viscoelastic relaxation. In the last phase (5), the force drops back to zero as the sleigh is pulled back, out of the tissue.
Trajectories
The situation was similar for the tested insertion trajectories, as our analyses did not show any statistically significant difference between the groups of \(v_{1}\), \(v_{2}\), \(v_{3}\), \(a_{1}\), \(a_{2}\) and \(a_{3}\). Considering \(F = m*a\) and \(F\Delta t = m*\Delta v\), this result does not match the initial expectation. This could be due to the fact that the differences between the tested groups were very small. For our results, trajectory specification within these ranges of piercing velocities and accelerations didn’t effect on the required piercing forces.
Vacuum configuration
\(H_{1}\) was based on the assumption that splitting the suction areas would stretch the tissue over the spike apertures reducing tissue indentation during piercing and making the process more independent of viscoelastic tissue property variations. We supposed an increase in tissue tension would in turn lead to higher piercing forces [11, 12]. However, with respect to our results, we saw, that the fixation point arrangement did not effect on the forces required to pierce the tissue.
Piercing force and influence of implant configurations
We measured maximum mean peak forces for four, eight and twelve spikes, of 6.4 ± 1.5 N, 13.6 ± 1.4 N and 21.7 ± 5.8 N, respectively. Schell used 17 staple clamps to measure the forces, resulting in 34 insertion points. He detected 202 N at the handle and 1885 N at the stapler tip, which is more than 80 times higher than our highest mean peak force of 21.7 N. As the experimental setup was not specified, a comparison is hardly possible [15]. However, we assume that a large part of the applied force in this case was converted into plastic deformation of the staples and did not occur due to stapler-tissue interaction.
Statistical analysis showed a significant increase in insertion forces with an increase in amount of spikes (Fig. 9). With increasing number of insertion bodies, the characteristic spike-tissue interaction force of each single body accumulated. The characteristic profile phases (Fig. 7) observed for the insertion of a single needle [6, 7], as well as for the implant configurations with four, eight and twelve tips, are all qualitatively equal, with only difference in peak force height. Furthermore, since the tips were not distributed evenly over the entire circumference of the piercing body, but only within the segments, the density per segment increased with rising tip amount. Once the tips contact tissue surface, micro-injuries in the surface are created [22], acting as anchor points. As spikes are pushed forward, indentation occurs [6, 7, 21]. Tissue tension is therefore increased between individual tips/attachment points. This pre-tensioning effect may contribute to the increase in insertion force. This correlation was also discovered by Frick and Butz [11, 12]. Implanting electrodes into the cerebral cortex of rats, Jensen et al. also detected an increase in force with increasing number of electrode shafts inserted [23]. By plotting the dependence between the number of spikes and the maximum force, we observed that the slope is nearly equal for all vacuum configurations and trajectory specifications, which led us to the conclusion, that these parameters do not influence the coherency. The derived correlation can be used to optimize the closing mechanism and reduce compression forces required, as orientation values for alternative implant configurations can be interpolated.
However, the measured maximum mean peak forces already do serve as orientation benchmarks for the design of our endoscopic anastomosis implant closing unit. Based on this, a force transmission mechanism is selected and dimensioned accordingly. To ensure high anastomosis stability, we decided to design an implant with 12 tips, for now. Therefore, it will be necessary to enable an axial force transmission of at least 28 N (22 ± 6 N) at the endoscope tip.
Wall thickness and lumen diameter
The colon lumen diameter limits the acceptable size of our application device and determines the required range of the implant expansion unit. The colonic wall thickness is of particular interest, to achieve an optimal compression pressure, thus providing sufficient stability without compromising blood supply of the bowel margins. The 5th and 95th percentiles, \(x_{p}\), are usually taken for ergonomic dimensioning of products to cover the majority of a population of interest. These benchmarks give us preliminary orientation of an upper and lower limit to be addressed for the size adaptability of our system. The average colon diameter of humans is 5 cm, and therefore comparable to our results [18]. Based on our findings, we conclude that the diameters of implant and applicator will be in a range between 24 and 40 mm, and the implant gap, compressing anastomosed tissue, between 2 and 5 mm.
Higher velocities and accelerations, with larger variations between groups, and more complex implant geometries and closure mechanisms will be assessed in the next step.
Notes
Formlabs: https://formlabs-media.formlabs.com/datasheets/1801089-TDS-ENUS-0P.pdf; last accessed:10.01.2020.
Formlabs; https://formlabs-media.formlabs.com/datasheets/Clear_Resin_Technical.pdlast accessed:10.01.2020.
References
Etzioni DA, Mack TM, Beart RW, Kaiser AM (2009) Diverticulitis in the United States: 1998–2005: changing patterns of disease and treatment. Ann Surg 249:210–217. https://doi.org/10.1097/SLA.0b013e3181952888
Chawla N, Butler EN, Lund J, Warren JL, Harlan LC, Yabroff KR (2013) Patterns of colorectal cancer care in Europe, Australia, and New Zealand. J Natl Cancer Inst Monogr. https://doi.org/10.1093/jncimonographs/lgt009
Thornton FJ, Barbul A (1997) Healing in the gastrointestinal tract. Surg Clin N Am 549–573
Carniel EL, Toniolo I, Fontanella CG (2020) Computational biomechanics: in-silico tools for the investigation of surgical procedures and devices. Bioengineering 7:48
Heijnsdijk EAM, van der Voort M, de Visser H, Dankelman J, Gouma DJ (2003) Inter- and intraindividual variabilities of perforation forces of human and pig bowel tissue. Surg Endosc Interven Tech 17:1923–1926. https://doi.org/10.1007/s00464-003-9002-3
Abolhassani N, Patel R, Moallem M (2004) Trajectory generation for robotic needle insertion in soft tissue. https://doi.org/10.1109/IEMBS.2004.1403782.
Okamura AM, Simone C, O’leary MD (2004) Force modeling for needle insertion into soft tissue. IEEE Trans Biomed Eng 51:1707–1716. https://doi.org/10.1109/TBME.2004.831542
Kumar S, Shrikanth V, Amrutur B, Asokan S, Bobji MS (2016) Detecting stages of needle penetration into tissues through force estimation at needle tip using fiber Bragg grating sensors. J Biomed Opt 21:127009. https://doi.org/10.1117/1.JBO.21.12.127009
Bao X, Li W, Lu M, Zhou ZR (2016) Experiment study on puncture force between MIS suture needle and soft tissue. Biosurf Biotribol 2:49–58. https://doi.org/10.1016/j.bsbt.2016.05.001
Jiang S, Li P, Yu Y, Liu J, Yang Z (2014) Experimental study of needle–tissue interaction forces: effect of needle geometries, insertion methods and tissue characteristics. J Biomech 47:3344–3353. https://doi.org/10.1016/j.jbiomech.2014.08.007
Frick TB, Marucci DD, Cartmill JA, Martin CJ, Walsh WR (2001) Resistance forces acting on suture needles. J Biomech 34:1335–1340. https://doi.org/10.1016/S0021-9290(01)00099-9
Butz KD, Griebel AJ, Novak T, Harris K, Kornokovich A, Chiappetta MF, Neu CP (2012) Prestress as an optimal biomechanical parameter for needle penetration. J Biomech 45:1176–1179. https://doi.org/10.1016/j.jbiomech.2012.01.049
Svendsen KH, Thomson G (1984) A new clamping and stretching procedure for determination of collagen fiber stiffness and strength relations upon maturation. J Biomech 17:225–229. https://doi.org/10.1016/0021-9290(84)90013-7
Kwon J, Park S, Lee MG, Kim Y, Kim B, Park J-O, Park J-H (2004) Design and fabrication of a bio-material property measurement system
Schell CW (2005) Die endoluminale Vollwandresektion im Dickdarm in Klammernahttechnik: Technologische Entwicklung und tierexperimentelle Ergebnisse. Tübingen
Pommergaard HC, Rosenberg J, Schumacher-Petersen C, Achiam MP (2011) Choosing the best animal species to mimic clinical colon anastomotic leakage in humans: a qualitative systematic review. Eur Surg Res 47:173–181
Carniel EL, Gramigna V, Fontanella CG, Stefanini C, Natali AN (2014) Constitutive formulations for the mechanical investigation of colonic tissues. J Biomed Mater Res Part A 102:1243–1254
Kararli TT (1995) Comparison of the gastrointestinal anatomy, physiology, and biochemistry of humans and commonly used laboratory animals. Biopharm Drug Dispos 16:351–380
Tachi K, Furukawa KS, Koshima I, Ushida T (2011) New microvascular anastomotic ring-coupling device using negative pressure. J Plast Reconstr Aesthet Surg 64:1187–1193. https://doi.org/10.1016/j.bjps.2011.04.008
Carniel EL, Gramigna V, Fontanella CG, Frigo A, Stefanini C, Rubini A (2014) an Natali, Characterization of the anisotropic mechanical behaviour of colonic tissues: experimental activity and constitutive formulation. Exp Physiol 99:759–771. https://doi.org/10.1113/expphysiol.2013.076091
DiMaio SP, Salcudean SE (2002) Simulated interactive needle insertion. In: Proceedings 10th symposium on haptic interfaces for virtual environment and teleoperator. pp 344–351. https://doi.org/10.1109/HAPTIC.2002.998979.
Mahvash M, Hayward V (2001) Haptic rendering of cutting: a fracture mechanics approach 1545–1143
Jensen W, Yoshida K, Hofmann UG (2006) In-vivo implant mechanics of flexible, silicon-based ACREO microelectrode arrays in rat cerebral cortex. IEEE Trans Biomed Eng 53:934–940. https://doi.org/10.1109/TBME.2006.872824
Funding
Open Access funding enabled and organized by Projekt DEAL. Funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation)—386233407 (‘CONNECT’).
Author information
Authors and Affiliations
Contributions
The conceptualization, writing, visualization and editing of the original draft were done by JS. The development of material and methods and experimental data assessment was done by JS and IP. The investigation of literature used for this paper was done by JS and IP, and draft was reviewed and revised by: JS, SF, PM, DW, ME.
Corresponding author
Ethics declarations
Conflict of interest
The content of this paper has been prepared solely and independently by the listed authors, all of whom have no disclosure to declare.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Steger, J., Patzke, I., Berlet, M. et al. Design of a force-measuring setup for colorectal compression anastomosis and first ex-vivo results. Int J CARS 16, 1335–1345 (2021). https://doi.org/10.1007/s11548-021-02371-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-021-02371-8