- Research
- Open access
- Published:
Automated left atrial time-resolved segmentation in MRI long-axis cine images using active contours
BMC Medical Imaging volume 21, Article number: 101 (2021)
Abstract
Background
Segmentation of the left atrium (LA) is required to evaluate atrial size and function, which are important imaging biomarkers for a wide range of cardiovascular conditions, such as atrial fibrillation, stroke, and diastolic dysfunction. LA segmentations are currently being performed manually, which is time-consuming and observer-dependent.
Methods
This study presents an automated image processing algorithm for time-resolved LA segmentation in cardiac magnetic resonance imaging (MRI) long-axis cine images of the 2-chamber (2ch) and 4-chamber (4ch) views using active contours. The proposed algorithm combines mitral valve tracking, automated threshold calculation, edge detection on a radially resampled image, edge tracking based on Dijkstra’s algorithm, and post-processing involving smoothing and interpolation. The algorithm was evaluated in 37 patients diagnosed mainly with paroxysmal atrial fibrillation. Segmentation accuracy was assessed using the Dice similarity coefficient (DSC) and Hausdorff distance (HD), with manual segmentations in all time frames as the reference standard. For inter-observer variability analysis, a second observer performed manual segmentations at end-diastole and end-systole on all subjects.
Results
The proposed automated method achieved high performance in segmenting the LA in long-axis cine sequences, with a DSC of 0.96 for 2ch and 0.95 for 4ch, and an HD of 5.5 mm for 2ch and 6.4 mm for 4ch. The manual inter-observer variability analysis had an average DSC of 0.95 and an average HD of 4.9 mm.
Conclusion
The proposed automated method achieved performance on par with human experts analyzing MRI images for evaluation of atrial size and function.
Introduction
In the United States, the lifetime risk of developing atrial fibrillation (AF) is 1 in 4, in people over 40 years old [1], and 6 million Americans are living with left-ventricle (LV) heart failure [2]. AF is an abnormal heart rhythm characterized by a rapid and irregular heartbeat, produced by ectopic beats originating from the left atrium (LA) [3]. LV heart failure, with or without reduced ejection fraction (EF), is a condition in which the heart fails to produce adequate blood flow to the body. The evaluation of AF, stroke risk, heart failure, and other cardiomyopathies is potentially enhanced by assessment of atrial volumes, function, and strain. For example, larger LA volumes and lower strain and LA EF were independent predictors of AF development in recent studies [4, 5], including sub-clinical AF [6]. LA function may improve the diagnostic accuracy and prognostic value of diastolic dysfunction in magnetic resonance imaging (MRI) [7]. LA evaluation plays an important role in imaging of AF, stroke risk related to AF, and diastolic dysfunction evaluation [8]. Despite its importance, values for LA volumes and function are not often reported clinically, due in part to the lack of available image processing tools for the LA.
Segmentation of the LA is required for evaluation of LA size and function [9], which are relevant diagnostic and prognostic imaging biomarkers for diverse cardiovascular conditions. LA strain evaluation is a more recent and potentially superior biomarker of cardiac disease, compared to LA size and function. Evaluation for all of these measures requires LA segmentation [10,11,12,13], and for strain the segmentation must be highly accurate and include all time-frames.
Cardiac MRI is considered the reference standard for cardiac volume assessment, offering accurate evaluation of LV and LA structure and function [14]. Cardiac MRI is routinely used to acquire time-resolved 2-chamber (2ch) and 4-chamber (4ch) long-axis cines of the heart (multiple 2D slices), providing high-contrast cardiac images at multiple points in the cardiac cycle. From these cine data, LA volumes can be measured, using the bi-plane method [15], and function can be estimated, based on the changes in LA volume over the heartbeat.
Volumes and strain during the entire cardiac cycle are essential, both for calculating volume-flow rates and strain rates, which require time derivatives, and for evaluation during multiple phases of the cardiac cycle, which include reservoir and conduit phase (filling in LV systole, and passive emptying in early diastole) and the active emptying phase. For the most part, LA segmentations are currently being performed manually, which is time-consuming and observer-dependent, underscoring the need for automated segmentation algorithms.
Advances in automated heart segmentation have primarily focused on the LV [16], secondarily on the right ventricle [17], whereas an automated segmentation method for the LA has yet to be routinely used in clinical practice [16] for several reasons. The LA chamber is typically smaller than the LV, with a thin myocardial border and a very variable shape across subjects. Boundaries are not clearly defined, primarily (i) the boundary between the LA cavity and pulmonary vein (PV) ostia, (ii) between the LA cavity and the left atrial appendage (LAA), and (iii) the mitral valve (MV) boundary, which separates the LA from the LV. Moreover, the MV has a dynamic motion throughout the cardiac cycle and is almost invisible in diastole.
Most LA segmentation methods are focused on whole heart images [18], with only a few methods developed for LA cine. Regarding these, one method uses feature tracking to propagate LA contours, once manually initialized on end-diastolic and/or end-systolic frames [10, 19], but this still demands user interaction. Machine learning techniques have been used to segment the LA in long-axis cine images, integrating a modified U-net based on deep convolutional neural networks with the unscented Kalman filter to impose temporal consistency in 2, 3 and 4-chamber views [20]. Another convolutional neural network architecture with a VGG-16 framework was proposed to predict a pixel-wise label map for the LA [21], but it was only trained on end-diastolic and end-systolic frames and with healthy volunteers. Regarding segmentation performance, the proposed methods provided high accuracy. However, from visual inspection it is clear that none of these methods provided a smooth delineation of the PV and LAA and a flat cut across the dynamic MV, nor did they present clinically validated, derived LA parameters (i.e., volumes, EFs, or strains) from the generated contours with validation against manual segmentation. Therefore, an automated method capable of segmenting all cine frames, with high accuracy, including validation of clinically utilized parameters, is still needed.
In this paper, we propose an automated LA time-resolved segmentation method for long-axis cine images of the 2ch and 4ch via active contours. The method combines MV tracking (fully or semi-automated), automated threshold calculation, edge detection on a radially resampled image, cost image optimization based on a Dijkstra algorithm and post-processing involving smoothing and interpolation. The resulting volumes and functional parameters are then compared to manual segmentation.
Methodology
We propose an automated LA segmentation based on active contours on a polar grid. The input data passed through four key steps: (i) preprocessing, (ii) polar mapping, (iii) edge detection and reconstruction, and (iv) Cartesian mapping. Figure 1 depicts the segmentation workflow in both views from a subject.
Input data
The LA segmentation method used long-axis cine images of the 2ch and 4ch views, acquired in a breath-hold with a balanced steady-state free precession sequence. Each cardiac cycle was composed of 30 images, representing time frames. Each image had an average spatial resolution of \(2 \times 2 \times 8\) \(\hbox {mm}^3\), and the MRI sequence acquisition parameters were a repetition time of 3 ms, an echo time of 1.5 ms, and a flip angle of 60\(^{\circ }\).
Preprocessing
The MV insertion points were obtained in a semi-automated manner, using feature tracking [22], and in a fully-automated manner, using residual neural networks [23], for comparison. The first approach was based on template tracking by normalized cross-correlation and a priori information by principal component analysis, and required a manual initialization of the MV points at end-diastole. The second approach employed a deep learning-based landmark annotation with no user interaction, whose sensitivity was tested (see Reproducibility due to variability in MV point placement section). These points define the MV plane, which separates the LA from the LV. Figure 1a illustrates the placement of the MV points. For the case of 2ch, the red and green points correspond to the anterior and inferior points, respectively, whereas for the case of 4ch, they correspond to the lateral and septal points.
To avoid false edge detection within the blood pool, all blood pool pixels with similar intensity were assigned under a single threshold value. To implement this threshold, the input image was considered as a Gaussian mixture of two-component pixel intensities, corresponding to sections of low intensity (muscle tissue), and high intensity (blood tissue). The Gaussian mixture parameters were obtained by an adapted Expectation-Maximization algorithm [24]. The threshold value was defined as the mean of the highest Gaussian distribution. Figure 1b shows the preprocessing step on the same subject, in which all pixels above the threshold were set to the threshold value.
Polar mapping
The cine images were mapped from a Cartesian grid (x, y) to a polar grid \((\theta ,r)\). Under this constraint, due to the geometric LA shape, other borders within the LA, related to the PV ostia and LAA, would be superimposed by the LA myocardium and thereby excluded using active contours.
The MV points were used to delimit the dynamic mitral valve and to initialize the polar mapping step in each frame. The reference point, illustrated in blue in Fig. 1b, was set in the middle of the mitral valve axis defined by the MV annotations. The radial range \(N_r\) covered twice the mitral valve length, whereas the angular range \(N_\theta\) was uniformly limited by the MV points, meaning it mapped from anterior to inferior point for 2ch or from lateral to septal point for 4ch, in other words, following the same color bar of Fig. 1, from red to green point. Due to the MV points’ dynamic behavior along the cardiac cycle, the reference point was updated for every time frame.
This task was performed in both views with a difference in the reference point position. This point was set exactly in the middle of the MV annotations for 2ch. However, for 4ch, the point was perpendicularly moved above the mitral valve plane for a quarter of its length, as illustrated in Fig. 1b. This change was due to the prolonged and curvilinear LA shape on this view, which was captured by this adjustment. The resampled LA shape in both cases resembled a “V” shape, as shown in Fig. 1c.
The polar map was generated with sampling intervals \(\Delta r\) of 0.25 pixels and \(\Delta \theta\) of \(1^{\circ }\), to resample the LA region from Cartesian coordinates, for each frame. The radial range varied from 150 to 200 rows depending on the mitral valve length, whereas the angular range was fixed to 180 or 230 columns, for 2ch and 4ch, respectively, as a result of the different reference point positions. The resultant image was interpolated with a bilinear resampling and its resolution was controlled by the sampling intervals.
Edge detection and reconstruction
A Canny edge detection technique [25] was used to robustly find the edges in each polar image. This resulted in a 2D matrix, considered to be the cost image \(I_\text {c}(i,j)\), with white pixels representing detected edges, with the MV points and reference line in polar coordinates as shown in Fig. 1d.
The presumed LA edge was found from \(I_\text {c}(i,j)\) using an optimization algorithm based on active contours, in each polar image. This active contour f was modeled as a physical string and defined as the minimization of its internal energy, \(E_\text {int}(f)\), and the energy of its position on the image, \(E_\text {im}(f,I_\text {c})\), by
The use of active contours facilitated the manipulation of physical behavior with the tuning parameters elasticity, \(\alpha\), and rigidity, \(\beta\). The elasticity controlled the amount of stretch and penalized changes from point to point in the contour, while the rigidity controlled the amount of curvature [26]. These tuning parameters defined the internal energy, first part of Eq. (1), as
An optimization problem positioned in a Cartesian grid in this case would yield local optimums, i.e., non-convex problem, which may not be the best solution. The polar mapping enabled a convex problem restricting the active contour to a proper discrete function of its own \(f:\text {N}^1\rightarrow \text {N}^1\), restricted to only pass through each column (\(\theta\)) in the grid once. With the reformulation of the optimization problem, described and solved with dynamic programming in [27], \(E_\text {int}(f)\), or Eq. (2), was discretized with approximate derivatives as
where \(N_r\) is the radial range previously described, and the index of f belongs to \(1\dots N_\theta\), being \(N_\theta\) the angular range, whereas \(E_\text {im}(f,I_\text {c})\), second part of Eq. (1), was discretized as
The discretized line f goes from the starting MV point (red) to the ending MV point (green) in the polar image, as illustrated in Fig. 1d. This optimization problem works as follows: (i) the algorithm keeps track of the set of included nodes (\(\text {N}\)), (ii) in each iteration the set \(\text {N}\) is extended with one node until all nodes are included and the algorithm terminates, (iii) nodes are selected so that the total cost from the source node to the included node is minimal, in the same way it is chosen in Dijkstra’s shortest path algorithm, (iv) the cost for including a node is computed from Eq. (3) and (4), (v) for each included node the algorithm also keeps track on the “parent” node, (vi) tentative nodes to be included the set \(\text {N}\) are stored in a heap data structure, (vii) the algorithm is initialized with an empty set \(\text {N}\), (viii) the optimal path through the image is finally captured by backtracking nodes from end node to start node (see [27] for more details and proof on optimality). This reformulation had the main advantage of maintaining the physical interpretation of treating the contour as a physical string, including its tuning parameters, and optimal solution. As the MV points position sometimes differed from the myocardial edge, the closest edge pixels to the MV points detected in the starting and ending columns were considered to be the contour tips. The \(E_\text {int}\) parameters were set as \(\alpha = 0.02\) and \(\beta = 0.0002\).
A post-processing technique was performed to smooth the surfaces and ensure false peaks were not detected. As the LA in the polar grid resembled a “V” shape, any detected non-predominant peaks in \(f(\theta ,r)\) were filtered. All reconstructed contours from a cine image were stacked in a 2D matrix \(F(\theta ,t)\), with rows representing the contour in each time frame and with columns representing the radial position of each active contour. This matrix was smoothed with a median filter. Figure 2 shows an example from a 2ch segmentation before and after this post-processing technique, with \(F(\theta ,t)\) from the whole cardiac cycle and \(f(\theta ,r)\) from a time frame. The result of the post-processing technique shows a smoothed \(F(\theta ,t)\) and a \(f(\theta ,r)\) without the false peak marked in purple.
Cartesian mapping
Once each contour was reconstructed and smoothed, it was converted to Cartesian coordinates for the output LA segmentation mask. With the initial reference point and the sampling intervals, the radial magnitude defined the distance from the reference point, and the angular magnitude marked the direction. This mask presented a dynamic cut with the mitral valve, delineated in a D-shaped form, and smooth separations from PV and LAA, as illustrated in Fig. 1e.
Evaluation
The method was compared against manual segmentation in 37 patients (15 females, age 53 ± 15 years), scanned on a 1.5T clinical MRI (Siemens Healthcare, Erlangen) for diverse cardiovascular indications, mostly paroxysmal AF, who were in sinus rhythm during the MRI cine acquisitions. The subjects were enrolled as part of an IRB-approved chart-review study. This data comprised a total of 1110 images (30 time frames for each patient) for each chamber view separately, with their corresponding manual segmentation performed with a standard operating protocol [28] using the software Segment [29]. For inter-observer variability analysis, a second observer performed manual segmentations at end-diastole and end-systole on all subjects.
Segmentation accuracy was evaluated using the Dice similarity coefficient (DSC) [30], mean contour distance (MCD), and Hausdorff distance (HD) [31]. The DSC was measured as the overlap between two segmentation masks, MCD and HD were measured as the distance error between two segmentation contours with the mean and the maximum values, respectively.
For clinical relevance, LA EDV, LA ESV, LA EF, and LA GLS were calculated from the automated and manual segmentation, as described by others [10, 14, 15]. LA volume (ml) along time t was calculated with the bi-plane method as
where the areas were obtained with the segmentation contours, and the lengths were calculated on the longitudinal axis as the distance of the perpendicular line measured from the MV center to the superior aspect of the LA. LA EDV (ml) and LA ESV (ml) were measured as the minimum and maximum values of LA volume, respectively, and LA EF (%) as
LA strain (%), the change in length normalized by the initial \(t=1\) length, was calculated using
where the perimeters were obtained with the segmentation contours excluding the MV. LA GLS (%) was measured as the maximum value of LA strain.
Reproducibility due to variability in MV point placement
As the position of MV points automatically initialized our LA segmentation process, differences in input points will influence segmentation results. The importance of MV point placement was assessed by adding a variability to the these initialization points. The variability was chosen as the manual inter-observer variability of MV points placement, calculated from 10 subjects to be 1.5 ± 0.7 mm. To study the initialization influence of MV point placement, all the MV points were randomly varied following this inter-observer variability, 100 times. For each simulation, the entire segmentation process was performed, and the segmentation accuracy and the clinical error were assessed against the first automatically obtained segmentations. We also evaluated the LA segmentation accuracy using our automated method for MV point placement [23].
Comparison to an automated LA segmentation method
The proposed method’s performance was compared against another automated method capable of LA volumes derivation, provided by the commercially available cardiac MR software \(\text {CVI}_\text {42}\) (Circle Cardiovascular Imaging, Calgary, Canada). A subset of 26 subjects (1560 images) was used for comparison between manual measures and automated measures of LA EDV and LA ESV. LA EF was manually derived using Eq. (6). LA GLS was not reported by the software. Segmentation masks were not extractable.
Implementation
The LA segmentation method was developed in MATLAB R2019a (Mathworks, Natick, Massachusetts) and implemented in the medical image analysis software Segment v3.1 R8109 [29] (http://segment.heiberg.se), and is freely available for research purposes.
Statistical analysis
Clinical metrics comparisons were performed using linear regression analysis, and Bland-Altman plots with manual measurements as a reference, including the bias with limits of agreement ± 1.96 standard deviation [32]. For the analysis, all clinical metrics were obtained for each set, by the observers and the method, and mean error, intra-class correlation coefficient (ICC) with a confidence interval of 95%, correlation value (R), and coefficient of variance (CoV) were calculated.
Results
The automated LA segmentation process was evaluated for accuracy, regarding both segmentation accuracy and accuracy of the clinical metrics, using statistical analysis. The total segmentation processing time in both chambers took 20 s per patient, compared with about 40 min from a manual segmentation.
Case study of an automated LA segmentation
An example of an automated LA segmentation in end-diastole, mid-systole, end-systole, and mid-diastole, in both views, is illustrated in Fig. 3a, showing excellent agreement with the corresponding manual segmentation. The LA volume and LA strain plots throughout the normalized cardiac cycle of the same subject compare the expert and automated values, as shown in Fig. 3b. This case presents the typical appearance of the LAA and PV ostia in the acquired images and their exclusion from the segmentation. A movie is available as additional file illustrating typical segmentation results across a cardiac cycle with time-resolved volume and strain curves (see Additional file 1).
Segmentation accuracy
The DSC, MCD, and HD between manual and automated segmentation in all the evaluation set are reported in Table 1. The automated segmentation method achieved an excellent DSC of 0.96 and 0.95, an excellent MCD of 1.2 and 1.3 mm, and a good HD of 5.5 and 6.4 mm, for 2ch and 4ch, respectively. The inter-observer agreement for manual segmentation, reported in Table 2, achieved an excellent DSC of 0.95 and 0.95, an excellent MCD of 1.4 and 1.2 mm, and a good HD of 5.0 and 4.8, for 2ch and 4ch, respectively.
Clinical metric accuracy
The accuracy of clinical metrics for the LA ESV, LA EDV, LA EF, and LA GLS of the automated method are shown in Table 3 compared against the manual metrics with excellent ICC and low bias. The inter-observer agreement is shown in Table 4. The regression and Bland-Altman plots for the LA parameters between the automated and manual measurements are presented in Fig. 4a, where an excellent correlation and good agreement were observed for each of the four parameters, with a lower accuracy for LA GLS. Figure 4b shows the corresponding inter-observer variability of these metrics. All reported correlation values are significant (p<0.0001).
Reproducibility due to variability in MV point placement
The reproducibility was studied through the evaluation of the MV points, obtaining the following errors (bias ± SD) on the clinical metrics: 0.31 ± 2.26 ml, − 0.84 ± 5.32 ml, − 0.52 ± 3.24%, and − 0.45 ± 3.43%, for the relative errors of LA EDV, LA ESV, LA EF, and LA GLS, respectively. This variation contributed an insignificant impact on the segmentation accuracy, which yielded a DSC of 0.97 ± 0.02, an MCD of 0.54 ± 0.44 mm, and an HD of 2.18 ± 2.19 mm, compared against the first automated results.
The agreement on the LA segmentation results between the initialization by the semi-automated [22] and the automated [23] method to track the MV points on the clinical metrics was: 0.52 ± 3.89 ml, 0.87 ± 5.06 ml, − 0.13 ± 2.85%, and − 0.03 ± 2.94% for LA EDV, LA ESV, LA EF, and LA GLS, respectively. The MV point placement presented minimal impact on the segmentation accuracy, yielding a DSC of 0.97 ± 0.02, an MCD of 0.48 ± 0.49 mm, and an HD of 2.08 ± 2.19 mm.
Comparison with an automated LA segmentation method
\(\text {CVI}_\text {42}\) produced a clinical-metric error (bias ± SD) against manual measures of 7.83 ± 10.72 ml, 11.07 ± 14.06 ml, and − 3.36 ± 9.19%, with an agreement (ICC) of 0.95, 0.91, and 0.80 for LA EDV, LA ESV, and LA EF, respectively. Although this learning-based method performance is strong, it was outperformed by our proposed method, as seen in Table 3. However, we were only able to compare LA volumes, not segmentations, and LA volumes depend strongly on the evaluation of LA length (also automatically calculated by \(\text {CVI}_\text {42}\)), which may be a source of discordance.
Discussion
In this study, a fast, automated method for time-resolved segmentation of the LA in 2ch and 4ch views from standard long-axis cine images was developed and evaluated. The measured values for LA volumes and EF were in agreement with other studies [33]. The proposed method showed excellent agreement with manual segmentation by expert readers. The method extracts clinically relevant LA functional parameters with precision approaching that of expert readers, 120 times faster, and with minimal sensitivity to user input error. This enables accurate, fast, and reproducible assessment of LA volumes and function in clinical routine. Although the approach requiring manual MV point insertion is not fully automated, we developed an automated MV tracking method for obtaining these points using an in-house deep learning-based network [23]. We tested its use in this LA segmentation method, demonstrating that a fully-automated LA segmentation method provides the same accuracy as the semi-automated version. Indeed, the accuracy of this LA segmentation method is not sensitive to variability in the MV points initialization.
While this paper mainly evaluates LA systolic parameters, the results should be similar for diastolic time-points. The proposed method is able to improve the value of cardiac MRI in clinical practice as it allows a fast assessment of phasic LA function. Indeed, LA function, especially in diastole, is rarely evaluated in cardiac MRI clinical routine. As shown in Fig. 4a, the method provides low bias for volumetric values (LA EDV and LA ESV), and functional metrics (LA EF and LA GLS). The slight overestimation in volumes is mainly caused by (i) the exact, automatic delimitation in the LA edge, bigger by one or two-layer pixels than the manual segmentation as its delimitation is observer-dependent, and by (ii) the difference in excluding the PV or LAA, which also influenced a higher HD. The most significant CoV is for LA strain, which is more sensitive than volume due to errors in perimeter delineation, especially with the valve positioning. However, the automatic strain measurement might be better and more reproducible, since manual contouring suffers from inconsistencies between time frames. Despite these differences, the CoVs of these clinical metrics are on par with other reported variations, with a range between 5.7 and 28% [11, 34, 35].
The automated method compared well to other segmentation techniques. For instance, an active contour LA segmentation method which automatically initializes a seed region with the Hough transform technique was evaluated for volumetric 3D MRI of the LA, with a good DSC of 0.82 ± 0.06 (n = 12) [36]. Also applied to 3D volumetric MRI, a multi-view convolutional neural network architecture with an adaptive fusion strategy yielded an excellent DSC of 0.95 (n = 20) [37]. Furthermore, the methods presented in a benchmark study [18] achieved similar performance. However, these methods were targeted to single time frame volumetric imaging of the LA, which is an entirely different application, and if they were applied to cine images, they would not delimit the segmentation with the PV, LAA and the MV boundaries. Other methods, similar to ours, sought to segment the LA in long-axis cine images, as noted above. These include a convolutional-neural-network method with the unscented Kalman filter which yielded an excellent DSC of 0.94 ± 0.04 for 2ch and 0.94 ± 0.08 for 4ch (n = 20) [20], and a similar method with a VGG-16 framework which also yielded an excellent DSC of 0.93 ± 0.05 for 2ch and 0.95 ± 0.02 for 4ch (n = 600) [21]. However, none of these methods reported clinical metrics nor were evaluated in a clinical population. We evaluated commercially available clinical cardiac MRI software capable of automated LA volume derivation (\(\text {CVI}_\text {42}\)) on our data sets, and found high performance, although it was still lower than our proposed method and unable to derive LA strain values.
The segmentation accuracy of the automated method was similar to the manual inter-observer accuracy, as shown in Fig. 4b. The inter-observer variability displayed a lower standard deviation but higher bias, which is inherent to each observer. The accuracy in the 4ch view was slightly lower than the inter-observer accuracy, meaning the 4ch view is more challenging to segment by this method. However, the automated method segmented the LA in the same manner in every frame, which improves the segmentation consistency along the cardiac cycle.
In the proposed workflow, the edge detection task was a key step prior the edge reconstruction. Within the vast range of edge detection techniques [38], in this application the chosen Canny’s technique achieved a clearer cost image preserving the physiological edges while reducing the imaging artifact noise, compared to Sobel, Prewitt and Roberts edge detection techniques, which were less computationally expensive but achieved a lower overall performance. Future work on this matter would involve the use of an edge detection technique and an adapted problem optimization in a three dimensional space, i.e., a stack of all cine frames.
Limitations
The main limitation of this study is the small sample size, which did not permit an analysis of how specific imaging issues, such as off-resonance artifacts within the LA blood pool or non-standard slice-prescription, affected segmentation accuracy. Another limitation might be a lack of healthy subjects, since our cohort consisted of mainly AF patients. However, this represents the types of patients who would require LA function evaluation. In our experience, analysis of patients is generally more demanding compared to controls.
Another limitation of the method is the need of initialization with the MV points, although this could be automated with residual neural networks [23]. An initialization task is commonly employed in segmentation problems, e.g. to initiate a thresholding task [39] or a region-growing process [40]. We showed that the inter-observer variability of MV placement, including the automated approach we have developed, minimally impacted the segmentation process.
Another limitation of this work is that total segmentation time which, while a matter of only seconds, was still longer than learning-based methods [20, 21] which require less than a second. Although this difference in time is a pitfall, compared to human labor our method is valuable and an attractive option.
Conclusion
The developed automated method performs time-resolved segmentation of the LA in cardiac MRI cine images using active contours in a polar grid. The method performs well in assessing LA volumes and strains against manual measurements in a patient population. Furthermore, the method yielded clinical metrics in line with inter-observer variability between expert readers.
Our LA segmentation method, based on active contours automatically initialized by mitral annular points placement, may be introduced as a accurate, fast, and reproducible method for measuring phasic LA volume and strain in cardiac MRI.
Availability of data and materials
The implemented software will be made freely available to researchers provided that they properly cite the current paper describing the method. The data sets used and/or analysed during the current study available from the corresponding author on reasonable request.
Abbreviations
- 2ch:
-
2-Chamber
- 4ch:
-
4-Chamber
- AF:
-
Atrial fibrillation
- CoV:
-
Coefficient of variation
- DSC:
-
Dice similarity coefficient
- EDV:
-
End-diastolic volume
- EF:
-
Ejection fraction
- ESV:
-
End-systolic volume
- GLS:
-
Global longitudinal strain
- HD:
-
Hausdorff distance
- ICC:
-
Intra-class correlation coefficient
- LA:
-
Left atrium
- LAA:
-
Left atrial appendage
- LV:
-
Left ventricle
- MA:
-
Mitral annular
- MCD:
-
Mean contour distance
- MRI:
-
Magnetic resonance imaging
- PV:
-
Pulmonary vein
- R:
-
Correlation value
References
Lloyd-Jones DM, Wang TJ, Leip EP, Larson MG, Levy D, Vasan RS, D’Agostino RB, Massaro JM, Beiser A, Wolf PA, Benjamin EJ. Lifetime risk for development of atrial fibrillation: the Framingham Heart Study. Circulation. 2004;110(9):1042–6.
Benjamin EJ, Muntner P, Alonso A, Bittencourt MS, Callaway CW, Carson AP, Chamberlain AM, Chang AR, Cheng S, Das SR, Delling FN, Djousse L, Elkind MSV, Ferguson JF, Fornage M, Jordan LC, Khan SS, Kissela BM, Knutson KL, Kwan TW, Lackland DT, Lewis TT, Lichtman JH, Longenecker CT, Loop MS, Lutsey PL, Martin SS, Matsushita K, Moran AE, Mussolino ME, O’Flaherty M, Pandey A, Perak AM, Rosamond WD, Roth GA, Sampson UKA, Satou GM, Schroeder EB, Shah SH, Spartano NL, Stokes A, Tirschwell DL, Tsao CW, Turakhia MP, VanWagner LB, Wilkins JT, Wong SS, Virani SS. Heart disease and stroke statistics-2019 update: a report from the American Heart Association. Circulation. 2019;139(10):56–528.
January CT, Wann LS, Alpert JS, Calkins H, Cigarroa JE, Cleveland JC, Conti JB, Ellinor PT, Ezekowitz MD, Field ME, et al. AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the heart rhythm society. J Am Coll Cardiol. 2014;64(21):1–76.
Habibi M, Zareian M, Venkatesh BA, Samiei S, Imai M, Wu C, Launer LJ, Shea S, Gottesman RF, Heckbert SR, et al. Left atrial mechanical function and incident ischemic cerebrovascular events independent of AF: insights from the MESA study. JACC Cardiovasc Imaging. 2019;12(12):2417–27.
Park JJ, Park JH, Hwang IC, Park JB, Cho GY, Marwick TH. Left atrial strain as a predictor of new-onset atrial fibrillation in patients with heart failure. JACC Cardiovasc Imaging. 2020;13(10):2071–81.
Bertelsen L, Diederichsen SZ, Haugan KJ, Brandes A, Graff C, Krieger D, Kronborg C, Køber L, Højberg S, Vejlstrup N, et al. Left atrial volume and function assessed by cardiac magnetic resonance imaging are markers of subclinical atrial fibrillation as detected by continuous monitoring. EP Europace. 2020;22(5):724–31.
Gopalakrishnan P, Biederman R. Impact of the 2016 ase/eacvi guidelines on diastolic function reporting in routine clinical practice. Echocardiography. 2020;37(4):546–53.
Cameli M, Mandoli GE, Loiacono F, Dini FL, Henein M, Mondillo S. Left atrial strain: a new parameter for assessment of left ventricular filling pressure. Heart Fail Rev. 2016;21(1):65–76.
Olsen FJ, Bertelsen L, de Knegt MC, Christensen TE, Vejlstrup N, Svendsen JH, Jensen JS, Biering-Sørensen T. Multimodality cardiac imaging for the assessment of left atrial function and the association with atrial arrhythmias. Circ Cardiovasc Imaging. 2016;9(10):004947.
Evin M, Cluzel P, Lamy J, Rosenbaum D, Kusmia S, Defrance C, Soulat G, Mousseaux E, Roux C, Clement K, et al. Assessment of left atrial function by mri myocardial feature tracking. J Magn Resonan Imaging. 2015;42(2):379–89.
Lamy J, Soulat G, Evin M, Huber A, De Cesare A, Giron A, Diebold B, Redheuil A, Mousseaux E, Kachenoura N. Scan-rescan reproducibility of ventricular and atrial mri feature tracking strain. Comput Biol Med. 2018;92:197–203.
Freed BH, Daruwalla V, Cheng JY, Aguilar FG, Beussink L, Choi A, Klein DA, Dixon D, Baldridge A, Rasmussen-Torvik LJ, et al. Prognostic utility and clinical significance of cardiac mechanics in heart failure with preserved ejection fraction: importance of left atrial strain. Circ Cardiovasc Imaging. 2016;9(3):003754.
Timóteo AT, Moura Branco L, Filipe F, Galrinho A, Rio P, Portugal G, Oliveira S, Ferreira RC. Cardiotoxicity in breast cancer treatment: What about left ventricular diastolic function and left atrial function? Echocardiography. 2019;36(10):1806–13.
Soufer A, Peters DC, Henry ML, Baldassarre LA. Advanced imaging of the left atrium with cardiac magnetic resonance: a review of current and emerging methods and clinical applications. Curr Radiol Rep. 2018;6(11):44.
Wylie JV Jr, Peters DC, Essebag V, Manning WJ, Josephson ME, Hauser TH. Left atrial function and scar after catheter ablation of atrial fibrillation. Heart Rhythm. 2008;5(5):656–62.
Peng P, Lekadir K, Gooya A, Shao L, Petersen SE, Frangi AF. A review of heart chamber segmentation for structural and functional analysis using cardiac magnetic resonance imaging. Magn Resonan Mater Phys Biol Med. 2016;29(2):155–95.
Petitjean C, Zuluaga MA, Bai W, Dacher J-N, Grosgeorge D, Caudron J, Ruan S, Ayed IB, Cardoso MJ, Chen H-C, et al. Right ventricle segmentation from cardiac mri: a collation study. Med Image Anal. 2015;19(1):187–202.
Tobon-Gomez C, Geers AJ, Peters J, Weese J, Pinto K, Karim R, Ammar M, Daoudi A, Margeta J, Sandoval Z, et al. Benchmark for algorithms segmenting the left atrium from 3d ct and mri datasets. IEEE Trans Med Imaging. 2015;34(7):1460–73.
Habibi M, Chahal H, Opdahl A, Gjesdal O, Helle-Valle TM, Heckbert SR, McClelland R, Wu C, Shea S, Hundley G, Bluemke DA, Lima JAC. Association of CMR-measured la function with heart failure development: results from the mesa study. JACC Cardiovasc Imaging. 2014;7(6):570–9.
Zhang X, Noga M, Martin DG, Punithakumar K. Fully automated left atrium segmentation from anatomical cine long-axis mri sequences using deep convolutional neural network with unscented Kalman filter. Med Image Anal. 2021;68:101916.
Bai W, Sinclair M, Tarroni G, Oktay O, Rajchl M, Vaillant G, Lee AM, Aung N, Lukaschuk E, Sanghvi MM, Zemrak F, Fung K, Paiva JM, Carapella V, Kim YJ, Suzuki H, Kainz B, Matthews PM, Petersen SE, Piechnik SK, Neubauer S, Glocker B, Rueckert D. Automated cardiovascular magnetic resonance image analysis with fully convolutional networks. J Cardiovasc Magn Resonan. 2018;20(1):65.
Seemann F, Pahlm U, Steding-Ehrenborg K, Ostenfeld E, Erlinge D, Dubois-Rande J-L, Jensen SE, Atar D, Arheden H, Carlsson M, Heiberg E. Time-resolved tracking of the atrioventricular plane displacement in cardiovascular magnetic resonance (CMR) images. BMC Med Imaging. 2017;17(1):19.
Gonzales RA, Onofrey JA, Lamy J, Seemann F, Heiberg E, Peters DC. Time-resolved tracking of the atrioventricular plane displacement in long-axis cine images with residual neural networks. In: International society for magnetic resonance in medicine, vol 29 (2020). p 2233.
Engblom H, Tufvesson J, Jablonowski R, Carlsson M, Aletras AH, Hoffmann P, Jacquier A, Kober F, Metzler B, Erlinge D, Atar D, Arheden H, Heiberg E. A new automatic algorithm for quantification of myocardial infarction imaged by late gadolinium enhancement cardiovascular magnetic resonance: experimental validation and comparison to expert delineations in multi-center, multi-vendor patient data. J Cardiovasc Magn Resonan. 2016;18(1):27.
Canny J. A computational approach to edge detection. IEEE Trans Pattern Anal Mach Intell. 1986;6:679–98.
Kass M, Witkin A, Terzopoulos D. Snakesa: active contour models. Int J Comput Vis. 1988;1(4):321–31.
Brandt E, Wigström L, Wranne B. Segmentation of echocardiographic image sequences using spatio-temporal information. In: International conference on medical image computing and computer-assisted intervention, Springer; 1999. p. 410–9.
Petersen SE, Aung N, Sanghvi MM, Zemrak F, Fung K, Paiva JM, Francis JM, Khanji MY, Lukaschuk E, Lee AM, Carapella V, Kim YJ, Leeson P, Piechnik SK, Neubauer S. Reference ranges for cardiac structure and function using cardiovascular magnetic resonance (CMR) in caucasians from the UK biobank population cohort. J Cardiovasc Magn Resonan. 2017;19(1):18.
Heiberg E, Sjögren J, Ugander M, Carlsson M, Engblom H, Arheden H. Design and validation of segment-freely available software for cardiovascular image analysis. BMC Med Imaging. 2010;10(1):1.
Zou KH, Warfield SK, Bharatha A, Tempany CM, Kaus MR, Haker SJ, Wells WM III, Jolesz FA, Kikinis R. Statistical validation of image segmentation quality based on a spatial overlap index1: scientific reports. Acad Radiol. 2004;11(2):178–89.
Huttenlocher DP, Rucklidge WJ, Klanderman GA. Comparing images using the Hausdorff distance under translation. In: Proceedings 1992 IEEE computer society conference on computer vision and pattern recognition; 1992. p. 654–6.
Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. 1986;327(8476):307–10.
Hudsmith L, Petersen S, Francis J, Robson M, Neubauer S. Normal human left and right ventricular and left atrial dimensions using steady state free precession magnetic resonance imaging. J Cardiovasc Magn Resonan. 2005;7(5):775–82.
Kowallick JT, Morton G, Lamata P, Jogiya R, Kutty S, Hasenfuß G, Lotz J, Nagel E, Chiribiri A, Schuster A. Quantification of atrial dynamics using cardiovascular magnetic resonance: inter-study reproducibility. J Cardiovasc Magn Resonan. 2015;17(1):36.
Zareian M, Ciuffo L, Habibi M, Opdahl A, Chamera EH, Wu CO, Bluemke DA, Lima JAC, Venkatesh BA. Left atrial structure and functional quantitation using cardiovascular magnetic resonance and multimodality tissue tracking: validation and reproducibility assessment. J Cardiovasc Magn Resonan. 2015;17(1):52.
Zhu L, Gao Y, Yezzi A, MacLeod R, Cates J, Tannenbaum A. Automatic segmentation of the left atrium from mri images using salient feature and contour evolution. In: 2012 Annual international conference of the IEEE engineering in medicine and biology society, IEEE; 2012. p. 3211–4.
Mortazi A, Karim R, Rhode K, Burt J, Bagci U. Cardiacnet: segmentation of left atrium and proximal pulmonary veins from mri using multi-view CNN. In: International conference on medical image computing and computer-assisted intervention, Springer; 2017. pp 377–85.
Shrivakshan G, Chandrasekar C. A comparison of various edge detection techniques used in image processing. Int J Comput Sci Issues. 2012;9(5):269.
John M, Rahn N. Automatic left atrium segmentation by cutting the blood pool at narrowings. In: International conference on medical image computing and computer-assisted intervention, Springer; 2005. p. 798–805.
Zhu L, Gao Y, Yezzi A, Tannenbaum A. Automatic segmentation of the left atrium from mr images via variational region growing with a moments-based shape prior. IEEE Trans Image Process. 2013;22(12):5111–22.
Acknowledgements
Not applicable.
Funding
This study was funded by the National Heart, Lung, and Blood Institute of the National Institute of Health, 5R01HL144706-02 and 1R01HL144706-01.
Author information
Authors and Affiliations
Contributions
All authors have edited the manuscript and approved the final version. RAG developed and implemented the algorithm, analyzed the results, performed time-resolved manual segmentations, the inter-observer analysis, drafted the manuscript, and helped conceive the study. FS and JL assisted in the design of the algorithm. PMA contributed by performing a second-observer analysis on the time-resolved segmentations. EH and VM contributed to the design of the algorithm and conceived the study. DCP contributed to the design of the study, the algorithm design, revised manual segmentations, and conceived the study. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All methods were carried out following relevant guidelines and regulations. The Yale Medical School IRB approved the study and waived the informed consent, as an exempt chart review study.
Consent for publication
Not applicable.
Competing interests
EH is the founder of Medviso AB, Lund, Sweden, which sells a commercial version of the software Segment. The other authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Gonzales, R.A., Seemann, F., Lamy, J. et al. Automated left atrial time-resolved segmentation in MRI long-axis cine images using active contours. BMC Med Imaging 21, 101 (2021). https://doi.org/10.1186/s12880-021-00630-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12880-021-00630-3