Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP)
<p>Image and schematic of the components of the SmartPill<sup>®</sup>; reprinted from <a href="https://doi.org/10.1016/S1369-7021(09)70272-X" target="_blank">https://doi.org/10.1016/S1369-7021(09)70272-X</a> (accessed on 2 December 2023) [<a href="#B21-sensors-24-00054" class="html-bibr">21</a>].</p> "> Figure 2
<p>Schema of the experimental setup.</p> "> Figure 3
<p>Intra-gastric and intra-peritoneal pressure as functions of pressure steps—patterns of the different signals.</p> "> Figure 4
<p>Correlation between intra-gastric pressure measured by the SmartPill™ and the wired sensor.</p> "> Figure 5
<p>Bias of measurement between intra-gastric pressure measured by the SmartPill™ and intra-gastric pressure measured by wired sensor.</p> "> Figure 6
<p>Correlation between pressure measured by the SmartPill™ and the wired intra-peritoneal sensor.</p> "> Figure 7
<p>Bias of measurement between intra-gastric pressure measured by the SmartPill™ and intra-peritoneal pressure measured by a wired sensor.</p> "> Figure 8
<p>Illustration of the SmartPill™ measurement bias. This figure represents the bias of measurement that appears of the pressure (mmHg) versus time (s) curve when placing the SmartPill™ (SP) capsule in the pig’s 1 stomach. Curve of the temperature (°C) as a function of time (s) is also represented.</p> "> Figure 9
<p>Comparison of the pressure measured by the SmartPill™ depending on chosen data export mode.</p> ">
Abstract
:1. Introduction
2. Material and Methods
2.1. Animals and Protocol Preparation
2.2. Pressure Measurements
2.2.1. Wired Sensors
2.2.2. SmartPill™
2.3. Pressure Variations
2.4. Data Processing
2.5. Statistical Analysis
3. Results
3.1. Pattern of the Different Signals
3.2. Correlation between Intra-Gastric Measurements by SmartPill™ (SP) and Wired Sensor (WIG)
3.3. Correlation between Intra-Gastric Measurements by SmartPill™ (SP) and Intra-Peritoneal Measurements (WIP)
4. Discussion
Author Contributions
Funding
Institutional Review Board Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Kirkpatrick, A.W.; Roberts, D.J.; De Waele, J.; Jaeschke, R.; Malbrain, M.L.; De Keulenaer, B.; Duchesne, J.; Bjorck, M.; Leppaniemi, A.; Ejike, J.C.; et al. Intra-abdominal hypertension and the abdominal compartment syndrome: Updated consensus definitions and clinical practice guidelines from the World Society of the Abdominal Compartment Syndrome. Intensive Care Med. 2013, 39, 1190–1206. [Google Scholar] [CrossRef] [PubMed]
- Malbrain, M.L.; Chiumello, D.; Pelosi, P.; Wilmer, A.; Brienza, N.; Malcangi, V.; Bihari, D.; Innes, R.; Cohen, J.; Singer, P.; et al. Prevalence of intra-abdominal hypertension in critically ill patients: A multicentre epidemiological study. Intensive Care Med. 2004, 30, 822–829. [Google Scholar] [CrossRef] [PubMed]
- Malbrain, M.L.N.G.; Deeren, D.; De Potter, T.J.R. Intra-abdominal hypertension in the critically ill: It is time to pay attention. Curr. Opin. Crit. Care 2005, 11, 156–171. [Google Scholar] [CrossRef] [PubMed]
- Klinge, U.; Klosterhalfen, B.; Conze, J.; Limberg, W.; Obolenski, B.; Öttinger, A.P.; Schumpelick, V. Modified mesh for hernia repair that is adapted to the physiology of the abdominal wall. Eur. J. Surg. 1998, 164, 951–960. [Google Scholar] [CrossRef] [PubMed]
- Kron, I.L.; Harman, P.K.; Nolan, S.P. The measurement of intra-abdominal pressure as a criterion for abdominal re-exploration. Ann. Surg. 1984, 199, 28–30. [Google Scholar] [CrossRef]
- Iberti, T.J.; Lieber, C.E.; Benjamin, E. Determination of intra-abdominal pressure using a transurethral bladder catheter: Clinical validation of the technique. J. Am. Soc. Anesthesiol. 1989, 70, 47–50. [Google Scholar] [CrossRef]
- Fusco, M.A.; Martin, R.S.; Chang, M.C. Estimation of Intra-abdominal Pressure by Bladder Pressure Measurement: Validity and Methodology. J. Trauma. Acute Care Surg. 2001, 50, 297–302. [Google Scholar] [CrossRef]
- Sugrue, M.; Buist, M.; Lee, A.; Sanchez Hillman, K. Intra-abdominal pressure measurement using a modified nasogastric tube: Description and validation of a new technique. Intensive Care Med. 1994, 20, 588–590. [Google Scholar] [CrossRef]
- Turnbull, D.; Webber, S.; Hamnegard, C.H.; Mills, G.H. Intra-abdominal pressure measurement: Validation of intragastric pressure as a measure of intra-abdominal pressure. BJA Br. J. Anaesth. 2007, 98, 628–634. [Google Scholar] [CrossRef]
- McCarthy, T.A. Validity of rectal pressure measurements as indication of intra-abdominal pressure changes during urodynamic evaluation. Urology 1982, 20, 657–660. [Google Scholar] [CrossRef]
- Carry, P.Y.; Banssillon, V. La pression intra-abdominale. Ann. Françaises D’anesthésie Réanimation 1994, 13, 381–399. [Google Scholar] [CrossRef] [PubMed]
- Tzelepis, G.E.; Nasiff, L.; McCool, F.D.; Hammond, J. Transmission of pressure within the abdomen. J. Appl. Physiol. 1996, 81, 1111–1114. [Google Scholar] [CrossRef] [PubMed]
- Malbrain, M.L.; Cheatham, M.L.; Kirkpatrick, A.; Sugrue, M.; Parr, M.; De Waele, J.; Balogh, Z.; Leppäniemi, A.; Olvera, C.; Ivatury, R.; et al. Results from the International Conference of Experts on Intra-abdominal Hypertension Abdominal Compartment Syndrome, I. Definitions. Intensive Care Med. 2006, 32, 1722–1732. [Google Scholar] [CrossRef] [PubMed]
- Shaw, J.M.; Hamad, N.M.; Coleman, T.J.; Egger, M.J.; Hsu, Y.; Hitchcock, R.; Nygaard, I.E. Intra-abdominal pressures during activity in women using an intra-vaginal pressure transducer. J. Sports Sci. 2014, 32, 1176–1185. [Google Scholar] [CrossRef] [PubMed]
- Tayebi, S.; Gutierrez, A.; Mohout, I.; Smets, E.; Wise, R.; Stiens, J.; Malbrain, M.L. A concise overview of non-invasive intra-abdominal pressure measurement techniques: From bench to bedside. J. Clin. Monit. Comput. 2021, 35, 51–70. [Google Scholar] [CrossRef] [PubMed]
- McCaffrey, C.; Chevalerias, O.; O’Mathuna, C.; Twomey, K. Swallowable-Capsule Technology. IEEE Pervasive Comput. 2008, 7, 23–29. [Google Scholar] [CrossRef]
- Ciccone, C.; Lee, P.; Loughlin, K.; Koh, D. What is the SmartPill®? Pharm. Wellness Rev. 2019, 6, 28–34. [Google Scholar]
- Soucasse, A.; Jourdan, A.; Edin, L.; Gillion, J.-F.; Masson, C.; Bege, T. A better understanding of daily life abdominal wall mechanical solicitation: Investigation of intra-abdominal pressure variations by intragastric wireless sensor in humans. Med. Eng. Phys. 2022, 104, 103813. [Google Scholar] [CrossRef]
- Rauch, S.; Johannes, A.; Zollhöfer, B.; Muellenbach, R.M. Evaluating intra-abdominal pressures in a porcine model of acute lung injury by using a wireless motility capsule. Med. Sci. Monit. Int. Med. J. Exp. Clin. Res. 2012, 18, BR163–BR166. [Google Scholar] [CrossRef]
- Wells, D.; Wells, D.J. Animal welfare and the 3Rs in European biomedical research. Ann. N. Y. Acad. Sci. 2011, 1245, 14–16. [Google Scholar] [CrossRef]
- Rohan, F.; David, H.G. Toward a miniaturized mechanical surgeon. Mater. Today 2009, 12, 14–20. [Google Scholar]
- Bland, M.; Altman, D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 327, 307–310. [Google Scholar] [CrossRef]
- Altman, D.G.; Bland, J.M. Assessing Agreement between Methods of Measurement. Clin. Chem. 2017, 63, 1653–1654. [Google Scholar] [CrossRef] [PubMed]
- Cobb, W.S.; Burns, J.M.; Kercher, K.W.; Matthews, B.D.; James Norton, H.; Todd Heniford, B. Normal Intraabdominal Pressure in Healthy Adults. J. Surg. Res. 2005, 129, 231–235. [Google Scholar] [CrossRef]
- De Keulenaer, B.L.; De Waele, J.J.; Powell, B.; Malbrain, M.L.N.G. What is normal intra-abdominal pressure and how is it affected by positioning, body mass and positive end-expiratory pressure? Intensive Care Med. 2009, 35, 969–976. [Google Scholar] [CrossRef]
- Romain, B.; Renard, Y.; Binquet, C.; Poghosyan, T.; Moszkowicz, D.; Gillion, J.F.; Ortega-Deballon, P.; Gadiri, N.; Mesli, S.; Bouillot, J.L.; et al. Recurrence after elective incisional hernia repair is more frequent than you think: An international prospective cohort from the French Society of Surgery. Surgery 2020, 168, 125–134. [Google Scholar] [CrossRef]
- Pereira, J.A.; López-Cano, M.; Hernández-Granados, P.; Feliu, X. Initial Results of the National Registry of Incisional Hernia. Cirugía Española 2016, 94, 595–602. [Google Scholar] [CrossRef]
PIG 1 | PIG 2 | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
WIG vs. SP | WIP vs. SP | WIG vs. SP | WIP vs. SP | ||||||||||
Steps (mmHg) | ρ | Bias | Limit | ρ | Bias | Limit | ρ | Bias | Limit | ρ | Bias | Limit | |
(mmHg) | (mmHg) | (mmHg) | (mmHg) | ||||||||||
Increase | 0 | 0.94 | −28.7 | 0.6 | 0.34 | −30.6 | 3.6 | 0.42 | −28.4 | 4.5 | 0.28 | −31 | 5.0 |
3 | 0.67 | −28.3 | 0.5 | 0.31 | −29.7 | 1.5 | 0.31 | −29.3 | 1.3 | 0.23 | −32.4 | 3.0 | |
6 | 0.83 | −28.2 | 0.5 | 0.46 | −29.9 | 2.4 | 0.61 | −28.7 | 0.8 | 0.31 | −31.5 | 1.9 | |
9 | 0.94 | −28.1 | 0.4 | 0.39 | −30.6 | 2.2 | 0.79 | −28.7 | 0.8 | 0.58 | −31.2 | 1.2 | |
12 | 0.93 | −27.7 | 0.4 | 0.70 | −29.9 | 1.3 | 0.90 | −28.7 | 0.4 | 0.3 | −31.6 | 2.6 | |
15 | 0.93 | −27.6 | 0.4 | 0.58 | −31.2 | 2.5 | 0.87 | −28.8 | 0.4 | 0.87 | −30.3 | 0.4 | |
Decrease | 12 | 0.96 | −27.6 | 0.4 | 0.64 | −30.3 | 2.3 | 0.94 | −29.4 | 0.4 | 0.93 | −30.5 | 0.5 |
9 | 0.96 | −27.8 | 0.4 | 0.55 | −28.7 | 1.4 | 0.94 | −29.6 | 0.4 | 0.81 | −30.7 | 1.3 | |
6 | 0.92 | −27.8 | 0.4 | 0.69 | −28.6 | 0.9 | 0.92 | −29.6 | 0.5 | 0.89 | −30.2 | 0.7 | |
3 | 0.92 | −27.8 | 0.3 | 0.57 | −28.8 | 0.7 | 0.89 | −29.8 | 0.4 | 0.86 | −31 | 1.6 | |
0 | 0.92 | −27.9 | 0.3 | 0.13 | −28.1 | 1.4 | 0.37 | −29.6 | 1.7 | 0.22 | −30.3 | 2.2 | |
Mean | 0.90 | −28.0 | 0.4 | 0.49 | −29.7 | 1.8 | 0.72 | −29.2 | 1.1 | 0.57 | −31.0 | 1.8 | |
SD | 0.08 | 0.3 | 0.1 | 0.18 | 1.0 | 0.9 | 0.25 | 0.5 | 1.2 | 0.30 | 0.7 | 1.3 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Soucasse, A.; Jourdan, A.; Edin, L.; Meunier, E.; Bege, T.; Masson, C. Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP). Sensors 2024, 24, 54. https://doi.org/10.3390/s24010054
Soucasse A, Jourdan A, Edin L, Meunier E, Bege T, Masson C. Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP). Sensors. 2024; 24(1):54. https://doi.org/10.3390/s24010054
Chicago/Turabian StyleSoucasse, Andréa, Arthur Jourdan, Lauriane Edin, Elise Meunier, Thierry Bege, and Catherine Masson. 2024. "Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP)" Sensors 24, no. 1: 54. https://doi.org/10.3390/s24010054
APA StyleSoucasse, A., Jourdan, A., Edin, L., Meunier, E., Bege, T., & Masson, C. (2024). Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP). Sensors, 24(1), 54. https://doi.org/10.3390/s24010054